Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus

Every year, bar-headed geese (Anser indicus) perform some of the most remarkable trans-Himalayan migrations, and researchers are increasingly interested in understanding the physiology underlying their high-altitude flight performance. A major challenge is generating reliable measurements of blood p...

Descripción completa

Detalles Bibliográficos
Autores principales: Harter, T. S., Reichert, M., Brauner, C. J., Milsom, W. K.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4778437/
https://www.ncbi.nlm.nih.gov/pubmed/27293706
http://dx.doi.org/10.1093/conphys/cov021
_version_ 1782419462160384000
author Harter, T. S.
Reichert, M.
Brauner, C. J.
Milsom, W. K.
author_facet Harter, T. S.
Reichert, M.
Brauner, C. J.
Milsom, W. K.
author_sort Harter, T. S.
collection PubMed
description Every year, bar-headed geese (Anser indicus) perform some of the most remarkable trans-Himalayan migrations, and researchers are increasingly interested in understanding the physiology underlying their high-altitude flight performance. A major challenge is generating reliable measurements of blood parameters on wild birds in the field, where established analytical techniques are often not available. Therefore, we validated two commonly used portable clinical analysers (PCAs), the i-STAT and the HemoCue systems, for the analysis of blood parameters in bar-headed geese. The pH, partial pressures of O(2) and CO(2) (PO(2) and PCO(2)), haemoglobin O(2) saturation (sO(2)), haematocrit (Hct) and haemoglobin concentration [Hb] were simultaneously measured with the two PCA systems (i-STAT for all parameters; HemoCue for [Hb]) and with conventional laboratory techniques over a physiological range of PO(2), PCO(2) and Hct. Our results indicate that the i-STAT system can generate reliable values on bar-headed goose whole blood pH, PO(2), PCO(2) and Hct, but we recommend correcting the obtained values using the linear equations determined here for higher accuracy. The i-STAT is probably not able to produce meaningful measurements of sO(2) and [Hb] over a range of physiologically relevant environmental conditions. However, we can recommend the use of the HemoCue to measure [Hb] in the bar-headed goose, if results are corrected. We emphasize that the equations that we provide to correct PCA results are applicable only to bar-headed goose whole blood under the conditions that we tested. We encourage researchers to validate i-STAT or HemoCue results thoroughly for their specific study conditions and species in order to yield accurate results.
format Online
Article
Text
id pubmed-4778437
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-47784372016-06-10 Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus Harter, T. S. Reichert, M. Brauner, C. J. Milsom, W. K. Conserv Physiol Research Articles Every year, bar-headed geese (Anser indicus) perform some of the most remarkable trans-Himalayan migrations, and researchers are increasingly interested in understanding the physiology underlying their high-altitude flight performance. A major challenge is generating reliable measurements of blood parameters on wild birds in the field, where established analytical techniques are often not available. Therefore, we validated two commonly used portable clinical analysers (PCAs), the i-STAT and the HemoCue systems, for the analysis of blood parameters in bar-headed geese. The pH, partial pressures of O(2) and CO(2) (PO(2) and PCO(2)), haemoglobin O(2) saturation (sO(2)), haematocrit (Hct) and haemoglobin concentration [Hb] were simultaneously measured with the two PCA systems (i-STAT for all parameters; HemoCue for [Hb]) and with conventional laboratory techniques over a physiological range of PO(2), PCO(2) and Hct. Our results indicate that the i-STAT system can generate reliable values on bar-headed goose whole blood pH, PO(2), PCO(2) and Hct, but we recommend correcting the obtained values using the linear equations determined here for higher accuracy. The i-STAT is probably not able to produce meaningful measurements of sO(2) and [Hb] over a range of physiologically relevant environmental conditions. However, we can recommend the use of the HemoCue to measure [Hb] in the bar-headed goose, if results are corrected. We emphasize that the equations that we provide to correct PCA results are applicable only to bar-headed goose whole blood under the conditions that we tested. We encourage researchers to validate i-STAT or HemoCue results thoroughly for their specific study conditions and species in order to yield accurate results. Oxford University Press 2015-05-19 /pmc/articles/PMC4778437/ /pubmed/27293706 http://dx.doi.org/10.1093/conphys/cov021 Text en © The Author 2015. Published by Oxford University Press and the Society for Experimental Biology. http://creativecommons.org/licenses/by/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Articles
Harter, T. S.
Reichert, M.
Brauner, C. J.
Milsom, W. K.
Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus
title Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus
title_full Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus
title_fullStr Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus
title_full_unstemmed Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus
title_short Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus
title_sort validation of the i-stat and hemocue systems for the analysis of blood parameters in the bar-headed goose, anser indicus
topic Research Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4778437/
https://www.ncbi.nlm.nih.gov/pubmed/27293706
http://dx.doi.org/10.1093/conphys/cov021
work_keys_str_mv AT harterts validationoftheistatandhemocuesystemsfortheanalysisofbloodparametersinthebarheadedgooseanserindicus
AT reichertm validationoftheistatandhemocuesystemsfortheanalysisofbloodparametersinthebarheadedgooseanserindicus
AT braunercj validationoftheistatandhemocuesystemsfortheanalysisofbloodparametersinthebarheadedgooseanserindicus
AT milsomwk validationoftheistatandhemocuesystemsfortheanalysisofbloodparametersinthebarheadedgooseanserindicus