Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry

The airflow perturbation device (APD) has been developed as a portable, easy to use, and a rapid response instrument for measuring respiratory resistance in humans. However, the APD has limited data validating it against the established techniques. This study used a mechanical system to simulate the...

Descripción completa

Detalles Bibliográficos
Autores principales: Pan, J., Saltos, A., Smith, D., Johnson, A., Vossoughi, J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4782682/
https://www.ncbi.nlm.nih.gov/pubmed/27006908
http://dx.doi.org/10.1155/2013/165782
_version_ 1782420003461529600
author Pan, J.
Saltos, A.
Smith, D.
Johnson, A.
Vossoughi, J.
author_facet Pan, J.
Saltos, A.
Smith, D.
Johnson, A.
Vossoughi, J.
author_sort Pan, J.
collection PubMed
description The airflow perturbation device (APD) has been developed as a portable, easy to use, and a rapid response instrument for measuring respiratory resistance in humans. However, the APD has limited data validating it against the established techniques. This study used a mechanical system to simulate the normal range of human breathing to validate the APD with the clinically accepted impulse oscillometry (IOS) technique. The validation system consisted of a sinusoidal flow generator with ten standardized resistance configurations that were shown to represent a total range of resistances from 0.12 to 0.95 kPa·L(−1) ·s (1.2–9.7 cm H(2)O·L(−1) ·s). Impulse oscillometry measurements and APD measurements of the mechanical system were recorded and compared at a constant airflow of 0.15 L·s(−1). Both the IOS and APD measurments were accurate in assessing nominal resistance. In addition, a strong linear relationship was observed between APD measurements and IOS measurements (R (2) = 0.999). A second series of measurements was made on ten human volunteers with external resistors added in their respiratory flow paths. Once calibrated with the mechanical system, the APD gave respiratory resistance measurements within 5% of IOS measurements. Because of their comparability to IOS measurements, APD measurements are shown to be valid representations of respiratory resistance.
format Online
Article
Text
id pubmed-4782682
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-47826822016-03-22 Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry Pan, J. Saltos, A. Smith, D. Johnson, A. Vossoughi, J. J Med Eng Research Article The airflow perturbation device (APD) has been developed as a portable, easy to use, and a rapid response instrument for measuring respiratory resistance in humans. However, the APD has limited data validating it against the established techniques. This study used a mechanical system to simulate the normal range of human breathing to validate the APD with the clinically accepted impulse oscillometry (IOS) technique. The validation system consisted of a sinusoidal flow generator with ten standardized resistance configurations that were shown to represent a total range of resistances from 0.12 to 0.95 kPa·L(−1) ·s (1.2–9.7 cm H(2)O·L(−1) ·s). Impulse oscillometry measurements and APD measurements of the mechanical system were recorded and compared at a constant airflow of 0.15 L·s(−1). Both the IOS and APD measurments were accurate in assessing nominal resistance. In addition, a strong linear relationship was observed between APD measurements and IOS measurements (R (2) = 0.999). A second series of measurements was made on ten human volunteers with external resistors added in their respiratory flow paths. Once calibrated with the mechanical system, the APD gave respiratory resistance measurements within 5% of IOS measurements. Because of their comparability to IOS measurements, APD measurements are shown to be valid representations of respiratory resistance. Hindawi Publishing Corporation 2013 2013-04-04 /pmc/articles/PMC4782682/ /pubmed/27006908 http://dx.doi.org/10.1155/2013/165782 Text en Copyright © 2013 J. Pan et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Pan, J.
Saltos, A.
Smith, D.
Johnson, A.
Vossoughi, J.
Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry
title Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry
title_full Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry
title_fullStr Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry
title_full_unstemmed Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry
title_short Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry
title_sort comparison of respiratory resistance measurements made with an airflow perturbation device with those from impulse oscillometry
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4782682/
https://www.ncbi.nlm.nih.gov/pubmed/27006908
http://dx.doi.org/10.1155/2013/165782
work_keys_str_mv AT panj comparisonofrespiratoryresistancemeasurementsmadewithanairflowperturbationdevicewiththosefromimpulseoscillometry
AT saltosa comparisonofrespiratoryresistancemeasurementsmadewithanairflowperturbationdevicewiththosefromimpulseoscillometry
AT smithd comparisonofrespiratoryresistancemeasurementsmadewithanairflowperturbationdevicewiththosefromimpulseoscillometry
AT johnsona comparisonofrespiratoryresistancemeasurementsmadewithanairflowperturbationdevicewiththosefromimpulseoscillometry
AT vossoughij comparisonofrespiratoryresistancemeasurementsmadewithanairflowperturbationdevicewiththosefromimpulseoscillometry