Validation of simple and inexpensive algometry using sphygmomanometer cuff and neuromuscular junction monitor with standardized laboratory algometer

BACKGROUND AND AIMS: The availability, ergonomics and economics prohibit the routine use of algometers in clinical practice and research by the anesthesiologists. A simple bedside technique of quantitative pain measurement would enable the routine use of algometry. We proposed to validate simple pai...

Descripción completa

Detalles Bibliográficos
Autores principales: Durga, Padmaja, Wudaru, Sreedhar Reddy, Khambam, Sunil Kumar Reddy, Chandra, Shobha Jagadish, Ramachandran, Gopinath
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4784219/
https://www.ncbi.nlm.nih.gov/pubmed/27006546
http://dx.doi.org/10.4103/0970-9185.173390
_version_ 1782420229841747968
author Durga, Padmaja
Wudaru, Sreedhar Reddy
Khambam, Sunil Kumar Reddy
Chandra, Shobha Jagadish
Ramachandran, Gopinath
author_facet Durga, Padmaja
Wudaru, Sreedhar Reddy
Khambam, Sunil Kumar Reddy
Chandra, Shobha Jagadish
Ramachandran, Gopinath
author_sort Durga, Padmaja
collection PubMed
description BACKGROUND AND AIMS: The availability, ergonomics and economics prohibit the routine use of algometers in clinical practice and research by the anesthesiologists. A simple bedside technique of quantitative pain measurement would enable the routine use of algometry. We proposed to validate simple pain provocation using sphygmomanometer cuff and the electric stimulation of neuromuscular junction monitor (TOF-guard, Organon Teknika) to measure pain against a standardized laboratory pressure algometer. MATERIAL AND METHODS: Pain detection threshold (Pdt) and pain tolerance threshold (Ptt) were measured in forty healthy volunteers of both genders, using the above three techniques. All measurements were repeated three times. The co-efficient of inter-rater reliability (or consistency) between three independent measurements obtained from each of the techniques was determined by Cronbach's co-efficient alpha (α C). The correlation between the mean Pdt and Ptt values recorded by standardized algometer and the sphygmomanometer technique and nerve stimulator technique was performed using Pearson Correlation. An r >0.5 and a two-tailed significance of <0.05 were considered as good correlation between the standardized algometer and the tested techniques. RESULTS: There was a good inter-rater reliability (α C > 0.7) for the three techniques. There was a good correlation with r >0.65 (P < 0.001) between the measurements of standardized pressure algometer and the two techniques being tested as alternatives for algometer to measure pain. CONCLUSION: The sphygmomanometer cuff technique and electrical stimulation with the peripheral nerve stimulator to measure pain threshold and tolerance provide a simple, efficient, repeatable measure of pain intensity and can be used as suitable alternatives to standard algometers.
format Online
Article
Text
id pubmed-4784219
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-47842192016-03-22 Validation of simple and inexpensive algometry using sphygmomanometer cuff and neuromuscular junction monitor with standardized laboratory algometer Durga, Padmaja Wudaru, Sreedhar Reddy Khambam, Sunil Kumar Reddy Chandra, Shobha Jagadish Ramachandran, Gopinath J Anaesthesiol Clin Pharmacol Original Article BACKGROUND AND AIMS: The availability, ergonomics and economics prohibit the routine use of algometers in clinical practice and research by the anesthesiologists. A simple bedside technique of quantitative pain measurement would enable the routine use of algometry. We proposed to validate simple pain provocation using sphygmomanometer cuff and the electric stimulation of neuromuscular junction monitor (TOF-guard, Organon Teknika) to measure pain against a standardized laboratory pressure algometer. MATERIAL AND METHODS: Pain detection threshold (Pdt) and pain tolerance threshold (Ptt) were measured in forty healthy volunteers of both genders, using the above three techniques. All measurements were repeated three times. The co-efficient of inter-rater reliability (or consistency) between three independent measurements obtained from each of the techniques was determined by Cronbach's co-efficient alpha (α C). The correlation between the mean Pdt and Ptt values recorded by standardized algometer and the sphygmomanometer technique and nerve stimulator technique was performed using Pearson Correlation. An r >0.5 and a two-tailed significance of <0.05 were considered as good correlation between the standardized algometer and the tested techniques. RESULTS: There was a good inter-rater reliability (α C > 0.7) for the three techniques. There was a good correlation with r >0.65 (P < 0.001) between the measurements of standardized pressure algometer and the two techniques being tested as alternatives for algometer to measure pain. CONCLUSION: The sphygmomanometer cuff technique and electrical stimulation with the peripheral nerve stimulator to measure pain threshold and tolerance provide a simple, efficient, repeatable measure of pain intensity and can be used as suitable alternatives to standard algometers. Medknow Publications & Media Pvt Ltd 2016 /pmc/articles/PMC4784219/ /pubmed/27006546 http://dx.doi.org/10.4103/0970-9185.173390 Text en Copyright: © Journal of Anaesthesiology Clinical Pharmacology http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution NonCommercial ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non commercially, as long as the author is credited and the new creations are licensed under the identical terms.
spellingShingle Original Article
Durga, Padmaja
Wudaru, Sreedhar Reddy
Khambam, Sunil Kumar Reddy
Chandra, Shobha Jagadish
Ramachandran, Gopinath
Validation of simple and inexpensive algometry using sphygmomanometer cuff and neuromuscular junction monitor with standardized laboratory algometer
title Validation of simple and inexpensive algometry using sphygmomanometer cuff and neuromuscular junction monitor with standardized laboratory algometer
title_full Validation of simple and inexpensive algometry using sphygmomanometer cuff and neuromuscular junction monitor with standardized laboratory algometer
title_fullStr Validation of simple and inexpensive algometry using sphygmomanometer cuff and neuromuscular junction monitor with standardized laboratory algometer
title_full_unstemmed Validation of simple and inexpensive algometry using sphygmomanometer cuff and neuromuscular junction monitor with standardized laboratory algometer
title_short Validation of simple and inexpensive algometry using sphygmomanometer cuff and neuromuscular junction monitor with standardized laboratory algometer
title_sort validation of simple and inexpensive algometry using sphygmomanometer cuff and neuromuscular junction monitor with standardized laboratory algometer
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4784219/
https://www.ncbi.nlm.nih.gov/pubmed/27006546
http://dx.doi.org/10.4103/0970-9185.173390
work_keys_str_mv AT durgapadmaja validationofsimpleandinexpensivealgometryusingsphygmomanometercuffandneuromuscularjunctionmonitorwithstandardizedlaboratoryalgometer
AT wudarusreedharreddy validationofsimpleandinexpensivealgometryusingsphygmomanometercuffandneuromuscularjunctionmonitorwithstandardizedlaboratoryalgometer
AT khambamsunilkumarreddy validationofsimpleandinexpensivealgometryusingsphygmomanometercuffandneuromuscularjunctionmonitorwithstandardizedlaboratoryalgometer
AT chandrashobhajagadish validationofsimpleandinexpensivealgometryusingsphygmomanometercuffandneuromuscularjunctionmonitorwithstandardizedlaboratoryalgometer
AT ramachandrangopinath validationofsimpleandinexpensivealgometryusingsphygmomanometercuffandneuromuscularjunctionmonitorwithstandardizedlaboratoryalgometer