Quantification of perflutren microsphere contrast destruction during transit through an ex vivo extracorporeal membrane oxygenation circuit

BACKGROUND: Echocardiography is a key investigation in the management of patients on extracorporeal membrane oxygenation (ECMO). However, echocardiographic images are often non-diagnostic in this patient population. Contrast-enhanced echocardiography may overcome many of these limitations but contra...

Descripción completa

Detalles Bibliográficos
Autores principales: Platts, David G., McDonald, Charles, Shekar, Kiran, Burstow, Darryl J., Mullany, Daniel, Ziegenfuss, Marc, Diab, Sara, Fraser, John F.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4788667/
https://www.ncbi.nlm.nih.gov/pubmed/26969640
http://dx.doi.org/10.1186/s40635-016-0079-0
_version_ 1782420748316442624
author Platts, David G.
McDonald, Charles
Shekar, Kiran
Burstow, Darryl J.
Mullany, Daniel
Ziegenfuss, Marc
Diab, Sara
Fraser, John F.
author_facet Platts, David G.
McDonald, Charles
Shekar, Kiran
Burstow, Darryl J.
Mullany, Daniel
Ziegenfuss, Marc
Diab, Sara
Fraser, John F.
author_sort Platts, David G.
collection PubMed
description BACKGROUND: Echocardiography is a key investigation in the management of patients on extracorporeal membrane oxygenation (ECMO). However, echocardiographic images are often non-diagnostic in this patient population. Contrast-enhanced echocardiography may overcome many of these limitations but contrast microspheres are hydrodynamically labile structures prone to destruction from shear forces and turbulent flow, which may exist within an ECMO circuit. This study sought to evaluate microsphere destruction (utilising signal intensity as a marker of contrast concentration) during transit through an ECMO circuit. METHODS: Activated Definity® contrast was diluted to 50 ml with normal saline and infused into a crystalloid primed ex vivo ECMO with a Quadrox oxygenator at 150 ml/h. Imaging was performed on pre- and post-pump head/oxygenator sections of the circuit using a Philips iE33 scanner and S5-1 transducer. Five-millimetre regions of interest were placed in the centre of the ultrasound field. Average signal intensity (decibels) was calculated at speeds of 1000, 2000, 3000 and 4000 rpm and then repeated with an infusion rate of 300 ml/h. The oxygenator was then spliced out of the circuit and the measures repeated. RESULTS: There was a significant reduction in contrast concentration during passage through the ECMO circuit at all speeds (with higher pump head speeds resulting in greater microsphere destruction). In a circuit with an oxygenator, relative decrease in signal intensity was 21.4 versus 5.2 % without an oxygenator. There was significant destruction of contrast microspheres during passage through the ECMO circuit at all pump head speeds. An oxygenator contributed to microsphere destruction at a significantly greater level than the pump head alone. There was no significant difference in mean signal intensity reduction in the circuit between an infusion of 150 or 300 ml/h (3.5 ± 3.2 versus 3.6 ± 2.5 dB, respectively, p = 0.79). CONCLUSIONS: Flow of contrast through an ECMO circuit results in significant destruction of microspheres. Circuits with an oxygenator result in significantly greater levels of contrast destruction than by the pump head alone. Clinicians should be cognisant of the relationship between ECMO circuit configurations, pump head speed and contrast destruction when performing a contrast-enhanced echocardiogram in patients supported with ECMO. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s40635-016-0079-0) contains supplementary material, which is available to authorized users.
format Online
Article
Text
id pubmed-4788667
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-47886672016-04-09 Quantification of perflutren microsphere contrast destruction during transit through an ex vivo extracorporeal membrane oxygenation circuit Platts, David G. McDonald, Charles Shekar, Kiran Burstow, Darryl J. Mullany, Daniel Ziegenfuss, Marc Diab, Sara Fraser, John F. Intensive Care Med Exp Research BACKGROUND: Echocardiography is a key investigation in the management of patients on extracorporeal membrane oxygenation (ECMO). However, echocardiographic images are often non-diagnostic in this patient population. Contrast-enhanced echocardiography may overcome many of these limitations but contrast microspheres are hydrodynamically labile structures prone to destruction from shear forces and turbulent flow, which may exist within an ECMO circuit. This study sought to evaluate microsphere destruction (utilising signal intensity as a marker of contrast concentration) during transit through an ECMO circuit. METHODS: Activated Definity® contrast was diluted to 50 ml with normal saline and infused into a crystalloid primed ex vivo ECMO with a Quadrox oxygenator at 150 ml/h. Imaging was performed on pre- and post-pump head/oxygenator sections of the circuit using a Philips iE33 scanner and S5-1 transducer. Five-millimetre regions of interest were placed in the centre of the ultrasound field. Average signal intensity (decibels) was calculated at speeds of 1000, 2000, 3000 and 4000 rpm and then repeated with an infusion rate of 300 ml/h. The oxygenator was then spliced out of the circuit and the measures repeated. RESULTS: There was a significant reduction in contrast concentration during passage through the ECMO circuit at all speeds (with higher pump head speeds resulting in greater microsphere destruction). In a circuit with an oxygenator, relative decrease in signal intensity was 21.4 versus 5.2 % without an oxygenator. There was significant destruction of contrast microspheres during passage through the ECMO circuit at all pump head speeds. An oxygenator contributed to microsphere destruction at a significantly greater level than the pump head alone. There was no significant difference in mean signal intensity reduction in the circuit between an infusion of 150 or 300 ml/h (3.5 ± 3.2 versus 3.6 ± 2.5 dB, respectively, p = 0.79). CONCLUSIONS: Flow of contrast through an ECMO circuit results in significant destruction of microspheres. Circuits with an oxygenator result in significantly greater levels of contrast destruction than by the pump head alone. Clinicians should be cognisant of the relationship between ECMO circuit configurations, pump head speed and contrast destruction when performing a contrast-enhanced echocardiogram in patients supported with ECMO. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s40635-016-0079-0) contains supplementary material, which is available to authorized users. Springer International Publishing 2016-03-11 /pmc/articles/PMC4788667/ /pubmed/26969640 http://dx.doi.org/10.1186/s40635-016-0079-0 Text en © Platts et al. 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Research
Platts, David G.
McDonald, Charles
Shekar, Kiran
Burstow, Darryl J.
Mullany, Daniel
Ziegenfuss, Marc
Diab, Sara
Fraser, John F.
Quantification of perflutren microsphere contrast destruction during transit through an ex vivo extracorporeal membrane oxygenation circuit
title Quantification of perflutren microsphere contrast destruction during transit through an ex vivo extracorporeal membrane oxygenation circuit
title_full Quantification of perflutren microsphere contrast destruction during transit through an ex vivo extracorporeal membrane oxygenation circuit
title_fullStr Quantification of perflutren microsphere contrast destruction during transit through an ex vivo extracorporeal membrane oxygenation circuit
title_full_unstemmed Quantification of perflutren microsphere contrast destruction during transit through an ex vivo extracorporeal membrane oxygenation circuit
title_short Quantification of perflutren microsphere contrast destruction during transit through an ex vivo extracorporeal membrane oxygenation circuit
title_sort quantification of perflutren microsphere contrast destruction during transit through an ex vivo extracorporeal membrane oxygenation circuit
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4788667/
https://www.ncbi.nlm.nih.gov/pubmed/26969640
http://dx.doi.org/10.1186/s40635-016-0079-0
work_keys_str_mv AT plattsdavidg quantificationofperflutrenmicrospherecontrastdestructionduringtransitthroughanexvivoextracorporealmembraneoxygenationcircuit
AT mcdonaldcharles quantificationofperflutrenmicrospherecontrastdestructionduringtransitthroughanexvivoextracorporealmembraneoxygenationcircuit
AT shekarkiran quantificationofperflutrenmicrospherecontrastdestructionduringtransitthroughanexvivoextracorporealmembraneoxygenationcircuit
AT burstowdarrylj quantificationofperflutrenmicrospherecontrastdestructionduringtransitthroughanexvivoextracorporealmembraneoxygenationcircuit
AT mullanydaniel quantificationofperflutrenmicrospherecontrastdestructionduringtransitthroughanexvivoextracorporealmembraneoxygenationcircuit
AT ziegenfussmarc quantificationofperflutrenmicrospherecontrastdestructionduringtransitthroughanexvivoextracorporealmembraneoxygenationcircuit
AT diabsara quantificationofperflutrenmicrospherecontrastdestructionduringtransitthroughanexvivoextracorporealmembraneoxygenationcircuit
AT fraserjohnf quantificationofperflutrenmicrospherecontrastdestructionduringtransitthroughanexvivoextracorporealmembraneoxygenationcircuit