Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies
Background: Insulin pump users experience periods of unexplained hyperglycemia. In some cases these may be due to insulin flow interruptions termed “silent occlusions,” which occur without activating the pump alarm and may require set replacement. Materials and Methods: In-line pressure profiles of...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Mary Ann Liebert, Inc.
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4790215/ https://www.ncbi.nlm.nih.gov/pubmed/26701357 http://dx.doi.org/10.1089/dia.2015.0342 |
_version_ | 1782420960862797824 |
---|---|
author | Gibney, Michael Xue, Zhenyi Swinney, Monica Bialonczyk, Damian Hirsch, Laurence |
author_facet | Gibney, Michael Xue, Zhenyi Swinney, Monica Bialonczyk, Damian Hirsch, Laurence |
author_sort | Gibney, Michael |
collection | PubMed |
description | Background: Insulin pump users experience periods of unexplained hyperglycemia. In some cases these may be due to insulin flow interruptions termed “silent occlusions,” which occur without activating the pump alarm and may require set replacement. Materials and Methods: In-line pressure profiles of a novel infusion set with a 6-mm, 28-gauge polymer, dual-ported catheter (BD FlowSmart™; Becton Dickinson and Co., Franklin Lakes, NJ) were compared with those of an existing infusion set (Quick-set(®); Medtronic MiniMed, Northridge, CA) in two separate studies involving insulin diluent infusions over 2.5–4.5-h periods in healthy adults without diabetes. Study 1, a pilot study (n = 25), compared the occurrence of flow interruption events (silent occlusions and/or occlusion alarms) between the two infusion sets and between manual or device-assisted insertion methods. Study 2 (n = 60) was designed to show ≥50% reduction in flow interruption events with the BD set after manual insertions. (Silent occlusions were defined by a continuous pressure rise for ≥30 min.) Results: In Study 1, significantly fewer silent occlusions were seen with BD FlowSmart versus Quick-set infusion sets for both manual (three of 22 [13.6%] vs. 12 of 24 [50%]; P = 0.012) and mechanical (two of 24 [8.3%] vs. nine of 25 [36%]; P = 0.037) insertions, yielding risk reductions of 73% (95% confidence interval [CI], 25–91%) and 77% (95% CI, 17–94%), respectively. In Study 2, flow interruption events occurred in three of 117 (2.6%) and 12 of 118 (10.2%) BD FlowSmart and Quick-set infusion sets, respectively, yielding a 75% risk reduction (95% CI, 20–92%; P = 0.030). Percentage of time with flow interruption was significantly lower with BD sets in both studies (P < 0.02). Leakage (>0.5 IU or 5 μL) occurred infrequently and did not differ between sets. Conclusions: A novel side-ported insulin infusion set demonstrated significant reductions in flow interruptions, including silent occlusions, versus a leading marketed set, which may improve insulin delivery. |
format | Online Article Text |
id | pubmed-4790215 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Mary Ann Liebert, Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-47902152016-03-24 Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies Gibney, Michael Xue, Zhenyi Swinney, Monica Bialonczyk, Damian Hirsch, Laurence Diabetes Technol Ther Original Articles Background: Insulin pump users experience periods of unexplained hyperglycemia. In some cases these may be due to insulin flow interruptions termed “silent occlusions,” which occur without activating the pump alarm and may require set replacement. Materials and Methods: In-line pressure profiles of a novel infusion set with a 6-mm, 28-gauge polymer, dual-ported catheter (BD FlowSmart™; Becton Dickinson and Co., Franklin Lakes, NJ) were compared with those of an existing infusion set (Quick-set(®); Medtronic MiniMed, Northridge, CA) in two separate studies involving insulin diluent infusions over 2.5–4.5-h periods in healthy adults without diabetes. Study 1, a pilot study (n = 25), compared the occurrence of flow interruption events (silent occlusions and/or occlusion alarms) between the two infusion sets and between manual or device-assisted insertion methods. Study 2 (n = 60) was designed to show ≥50% reduction in flow interruption events with the BD set after manual insertions. (Silent occlusions were defined by a continuous pressure rise for ≥30 min.) Results: In Study 1, significantly fewer silent occlusions were seen with BD FlowSmart versus Quick-set infusion sets for both manual (three of 22 [13.6%] vs. 12 of 24 [50%]; P = 0.012) and mechanical (two of 24 [8.3%] vs. nine of 25 [36%]; P = 0.037) insertions, yielding risk reductions of 73% (95% confidence interval [CI], 25–91%) and 77% (95% CI, 17–94%), respectively. In Study 2, flow interruption events occurred in three of 117 (2.6%) and 12 of 118 (10.2%) BD FlowSmart and Quick-set infusion sets, respectively, yielding a 75% risk reduction (95% CI, 20–92%; P = 0.030). Percentage of time with flow interruption was significantly lower with BD sets in both studies (P < 0.02). Leakage (>0.5 IU or 5 μL) occurred infrequently and did not differ between sets. Conclusions: A novel side-ported insulin infusion set demonstrated significant reductions in flow interruptions, including silent occlusions, versus a leading marketed set, which may improve insulin delivery. Mary Ann Liebert, Inc. 2016-03-01 /pmc/articles/PMC4790215/ /pubmed/26701357 http://dx.doi.org/10.1089/dia.2015.0342 Text en © The Author(s) 2016; Published by Mary Ann Liebert, Inc. This Open Access article is distributed under the terms of the Creative Commons Attribution Noncommercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited. |
spellingShingle | Original Articles Gibney, Michael Xue, Zhenyi Swinney, Monica Bialonczyk, Damian Hirsch, Laurence Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies |
title | Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies |
title_full | Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies |
title_fullStr | Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies |
title_full_unstemmed | Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies |
title_short | Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies |
title_sort | reduced silent occlusions with a novel catheter infusion set (bd flowsmart): results from two open-label comparative studies |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4790215/ https://www.ncbi.nlm.nih.gov/pubmed/26701357 http://dx.doi.org/10.1089/dia.2015.0342 |
work_keys_str_mv | AT gibneymichael reducedsilentocclusionswithanovelcatheterinfusionsetbdflowsmartresultsfromtwoopenlabelcomparativestudies AT xuezhenyi reducedsilentocclusionswithanovelcatheterinfusionsetbdflowsmartresultsfromtwoopenlabelcomparativestudies AT swinneymonica reducedsilentocclusionswithanovelcatheterinfusionsetbdflowsmartresultsfromtwoopenlabelcomparativestudies AT bialonczykdamian reducedsilentocclusionswithanovelcatheterinfusionsetbdflowsmartresultsfromtwoopenlabelcomparativestudies AT hirschlaurence reducedsilentocclusionswithanovelcatheterinfusionsetbdflowsmartresultsfromtwoopenlabelcomparativestudies |