Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies

Background: Insulin pump users experience periods of unexplained hyperglycemia. In some cases these may be due to insulin flow interruptions termed “silent occlusions,” which occur without activating the pump alarm and may require set replacement. Materials and Methods: In-line pressure profiles of...

Descripción completa

Detalles Bibliográficos
Autores principales: Gibney, Michael, Xue, Zhenyi, Swinney, Monica, Bialonczyk, Damian, Hirsch, Laurence
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Mary Ann Liebert, Inc. 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4790215/
https://www.ncbi.nlm.nih.gov/pubmed/26701357
http://dx.doi.org/10.1089/dia.2015.0342
_version_ 1782420960862797824
author Gibney, Michael
Xue, Zhenyi
Swinney, Monica
Bialonczyk, Damian
Hirsch, Laurence
author_facet Gibney, Michael
Xue, Zhenyi
Swinney, Monica
Bialonczyk, Damian
Hirsch, Laurence
author_sort Gibney, Michael
collection PubMed
description Background: Insulin pump users experience periods of unexplained hyperglycemia. In some cases these may be due to insulin flow interruptions termed “silent occlusions,” which occur without activating the pump alarm and may require set replacement. Materials and Methods: In-line pressure profiles of a novel infusion set with a 6-mm, 28-gauge polymer, dual-ported catheter (BD FlowSmart™; Becton Dickinson and Co., Franklin Lakes, NJ) were compared with those of an existing infusion set (Quick-set(®); Medtronic MiniMed, Northridge, CA) in two separate studies involving insulin diluent infusions over 2.5–4.5-h periods in healthy adults without diabetes. Study 1, a pilot study (n = 25), compared the occurrence of flow interruption events (silent occlusions and/or occlusion alarms) between the two infusion sets and between manual or device-assisted insertion methods. Study 2 (n = 60) was designed to show ≥50% reduction in flow interruption events with the BD set after manual insertions. (Silent occlusions were defined by a continuous pressure rise for ≥30 min.) Results: In Study 1, significantly fewer silent occlusions were seen with BD FlowSmart versus Quick-set infusion sets for both manual (three of 22 [13.6%] vs. 12 of 24 [50%]; P = 0.012) and mechanical (two of 24 [8.3%] vs. nine of 25 [36%]; P = 0.037) insertions, yielding risk reductions of 73% (95% confidence interval [CI], 25–91%) and 77% (95% CI, 17–94%), respectively. In Study 2, flow interruption events occurred in three of 117 (2.6%) and 12 of 118 (10.2%) BD FlowSmart and Quick-set infusion sets, respectively, yielding a 75% risk reduction (95% CI, 20–92%; P = 0.030). Percentage of time with flow interruption was significantly lower with BD sets in both studies (P < 0.02). Leakage (>0.5 IU or 5 μL) occurred infrequently and did not differ between sets. Conclusions: A novel side-ported insulin infusion set demonstrated significant reductions in flow interruptions, including silent occlusions, versus a leading marketed set, which may improve insulin delivery.
format Online
Article
Text
id pubmed-4790215
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Mary Ann Liebert, Inc.
record_format MEDLINE/PubMed
spelling pubmed-47902152016-03-24 Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies Gibney, Michael Xue, Zhenyi Swinney, Monica Bialonczyk, Damian Hirsch, Laurence Diabetes Technol Ther Original Articles Background: Insulin pump users experience periods of unexplained hyperglycemia. In some cases these may be due to insulin flow interruptions termed “silent occlusions,” which occur without activating the pump alarm and may require set replacement. Materials and Methods: In-line pressure profiles of a novel infusion set with a 6-mm, 28-gauge polymer, dual-ported catheter (BD FlowSmart™; Becton Dickinson and Co., Franklin Lakes, NJ) were compared with those of an existing infusion set (Quick-set(®); Medtronic MiniMed, Northridge, CA) in two separate studies involving insulin diluent infusions over 2.5–4.5-h periods in healthy adults without diabetes. Study 1, a pilot study (n = 25), compared the occurrence of flow interruption events (silent occlusions and/or occlusion alarms) between the two infusion sets and between manual or device-assisted insertion methods. Study 2 (n = 60) was designed to show ≥50% reduction in flow interruption events with the BD set after manual insertions. (Silent occlusions were defined by a continuous pressure rise for ≥30 min.) Results: In Study 1, significantly fewer silent occlusions were seen with BD FlowSmart versus Quick-set infusion sets for both manual (three of 22 [13.6%] vs. 12 of 24 [50%]; P = 0.012) and mechanical (two of 24 [8.3%] vs. nine of 25 [36%]; P = 0.037) insertions, yielding risk reductions of 73% (95% confidence interval [CI], 25–91%) and 77% (95% CI, 17–94%), respectively. In Study 2, flow interruption events occurred in three of 117 (2.6%) and 12 of 118 (10.2%) BD FlowSmart and Quick-set infusion sets, respectively, yielding a 75% risk reduction (95% CI, 20–92%; P = 0.030). Percentage of time with flow interruption was significantly lower with BD sets in both studies (P < 0.02). Leakage (>0.5 IU or 5 μL) occurred infrequently and did not differ between sets. Conclusions: A novel side-ported insulin infusion set demonstrated significant reductions in flow interruptions, including silent occlusions, versus a leading marketed set, which may improve insulin delivery. Mary Ann Liebert, Inc. 2016-03-01 /pmc/articles/PMC4790215/ /pubmed/26701357 http://dx.doi.org/10.1089/dia.2015.0342 Text en © The Author(s) 2016; Published by Mary Ann Liebert, Inc. This Open Access article is distributed under the terms of the Creative Commons Attribution Noncommercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
spellingShingle Original Articles
Gibney, Michael
Xue, Zhenyi
Swinney, Monica
Bialonczyk, Damian
Hirsch, Laurence
Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies
title Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies
title_full Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies
title_fullStr Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies
title_full_unstemmed Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies
title_short Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies
title_sort reduced silent occlusions with a novel catheter infusion set (bd flowsmart): results from two open-label comparative studies
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4790215/
https://www.ncbi.nlm.nih.gov/pubmed/26701357
http://dx.doi.org/10.1089/dia.2015.0342
work_keys_str_mv AT gibneymichael reducedsilentocclusionswithanovelcatheterinfusionsetbdflowsmartresultsfromtwoopenlabelcomparativestudies
AT xuezhenyi reducedsilentocclusionswithanovelcatheterinfusionsetbdflowsmartresultsfromtwoopenlabelcomparativestudies
AT swinneymonica reducedsilentocclusionswithanovelcatheterinfusionsetbdflowsmartresultsfromtwoopenlabelcomparativestudies
AT bialonczykdamian reducedsilentocclusionswithanovelcatheterinfusionsetbdflowsmartresultsfromtwoopenlabelcomparativestudies
AT hirschlaurence reducedsilentocclusionswithanovelcatheterinfusionsetbdflowsmartresultsfromtwoopenlabelcomparativestudies