Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings
The advent of modern technologies in radiotherapy poses an increased challenge in the determination of dosimetric parameters of small fields that exhibit a high degree of uncertainty. Percent depth dose and beam profiles were acquired using different detectors in two different orientations. The para...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4795411/ https://www.ncbi.nlm.nih.gov/pubmed/27051165 http://dx.doi.org/10.4103/0971-6203.177284 |
_version_ | 1782421606662930432 |
---|---|
author | Godson, Henry Finlay Ravikumar, M. Sathiyan, S. Ganesh, K. M. Ponmalar, Y. Retna Varatharaj, C. |
author_facet | Godson, Henry Finlay Ravikumar, M. Sathiyan, S. Ganesh, K. M. Ponmalar, Y. Retna Varatharaj, C. |
author_sort | Godson, Henry Finlay |
collection | PubMed |
description | The advent of modern technologies in radiotherapy poses an increased challenge in the determination of dosimetric parameters of small fields that exhibit a high degree of uncertainty. Percent depth dose and beam profiles were acquired using different detectors in two different orientations. The parameters such as relative surface dose (D(S)), depth of dose maximum (D(max)), percentage dose at 10 cm (D(10)), penumbral width, flatness, and symmetry were evaluated with different detectors. The dosimetric data were acquired for fields defined by jaws alone, multileaf collimator (MLC) alone, and by MLC while the jaws were positioned at 0, 0.25, 0.5, and 1.0 cm away from MLC leaf-end using a Varian linear accelerator with 6 MV photon beam. The accuracy in the measurement of dosimetric parameters with various detectors for three different field definitions was evaluated. The relative D(S)(38.1%) with photon field diode in parallel orientation was higher than electron field diode (EFD) (27.9%) values for 1 cm ×1 cm field. An overestimation of 5.7% and 8.6% in D(10) depth were observed for 1 cm ×1 cm field with RK ion chamber in parallel and perpendicular orientation, respectively, for the fields defined by MLC while jaw positioned at the edge of the field when compared to EFD values in parallel orientation. For this field definition, the in-plane penumbral widths obtained with ion chamber in parallel and perpendicular orientation were 3.9 mm, 5.6 mm for 1 cm ×1 cm field, respectively. Among all detectors used in the study, the unshielded diodes were found to be an appropriate choice of detector for the measurement of beam parameters in small fields. |
format | Online Article Text |
id | pubmed-4795411 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-47954112016-04-05 Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings Godson, Henry Finlay Ravikumar, M. Sathiyan, S. Ganesh, K. M. Ponmalar, Y. Retna Varatharaj, C. J Med Phys Original Article The advent of modern technologies in radiotherapy poses an increased challenge in the determination of dosimetric parameters of small fields that exhibit a high degree of uncertainty. Percent depth dose and beam profiles were acquired using different detectors in two different orientations. The parameters such as relative surface dose (D(S)), depth of dose maximum (D(max)), percentage dose at 10 cm (D(10)), penumbral width, flatness, and symmetry were evaluated with different detectors. The dosimetric data were acquired for fields defined by jaws alone, multileaf collimator (MLC) alone, and by MLC while the jaws were positioned at 0, 0.25, 0.5, and 1.0 cm away from MLC leaf-end using a Varian linear accelerator with 6 MV photon beam. The accuracy in the measurement of dosimetric parameters with various detectors for three different field definitions was evaluated. The relative D(S)(38.1%) with photon field diode in parallel orientation was higher than electron field diode (EFD) (27.9%) values for 1 cm ×1 cm field. An overestimation of 5.7% and 8.6% in D(10) depth were observed for 1 cm ×1 cm field with RK ion chamber in parallel and perpendicular orientation, respectively, for the fields defined by MLC while jaw positioned at the edge of the field when compared to EFD values in parallel orientation. For this field definition, the in-plane penumbral widths obtained with ion chamber in parallel and perpendicular orientation were 3.9 mm, 5.6 mm for 1 cm ×1 cm field, respectively. Among all detectors used in the study, the unshielded diodes were found to be an appropriate choice of detector for the measurement of beam parameters in small fields. Medknow Publications & Media Pvt Ltd 2016 /pmc/articles/PMC4795411/ /pubmed/27051165 http://dx.doi.org/10.4103/0971-6203.177284 Text en Copyright: © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Godson, Henry Finlay Ravikumar, M. Sathiyan, S. Ganesh, K. M. Ponmalar, Y. Retna Varatharaj, C. Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings |
title | Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings |
title_full | Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings |
title_fullStr | Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings |
title_full_unstemmed | Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings |
title_short | Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings |
title_sort | analysis of small field percent depth dose and profiles: comparison of measurements with various detectors and effects of detector orientation with different jaw settings |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4795411/ https://www.ncbi.nlm.nih.gov/pubmed/27051165 http://dx.doi.org/10.4103/0971-6203.177284 |
work_keys_str_mv | AT godsonhenryfinlay analysisofsmallfieldpercentdepthdoseandprofilescomparisonofmeasurementswithvariousdetectorsandeffectsofdetectororientationwithdifferentjawsettings AT ravikumarm analysisofsmallfieldpercentdepthdoseandprofilescomparisonofmeasurementswithvariousdetectorsandeffectsofdetectororientationwithdifferentjawsettings AT sathiyans analysisofsmallfieldpercentdepthdoseandprofilescomparisonofmeasurementswithvariousdetectorsandeffectsofdetectororientationwithdifferentjawsettings AT ganeshkm analysisofsmallfieldpercentdepthdoseandprofilescomparisonofmeasurementswithvariousdetectorsandeffectsofdetectororientationwithdifferentjawsettings AT ponmalaryretna analysisofsmallfieldpercentdepthdoseandprofilescomparisonofmeasurementswithvariousdetectorsandeffectsofdetectororientationwithdifferentjawsettings AT varatharajc analysisofsmallfieldpercentdepthdoseandprofilescomparisonofmeasurementswithvariousdetectorsandeffectsofdetectororientationwithdifferentjawsettings |