Cargando…
Comparison of computed tomography dose index in polymethyl methacrylate and nylon dosimetry phantoms
The use of computed tomography (CT) scanning has been growing steadily. Therefore, CT dose measurement is becoming increasingly important for patient protection and optimization. A phantom is an important tool for dose measurement. This paper focuses on the evaluation of a CT dosimetry phantom made...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4795417/ https://www.ncbi.nlm.nih.gov/pubmed/27051170 http://dx.doi.org/10.4103/0971-6203.177287 |
Sumario: | The use of computed tomography (CT) scanning has been growing steadily. Therefore, CT dose measurement is becoming increasingly important for patient protection and optimization. A phantom is an important tool for dose measurement. This paper focuses on the evaluation of a CT dosimetry phantom made from nylon, instead of the standard polymethyl methacrylate (PMMA), which is not readily available or is too expensive in some countries. Comparison between phantoms made from the two materials is made in terms of measurements of the CT dose indices (CTDI). These were measured for four different beam widths and kVp settings at the center and periphery in head and body phantoms made from both materials and weighted CTDIs (CTDI(w)) were calculated. CT numbers along the z-axis of the phantom were also measured at the center and four peripheral positions of each scanned slice to check phantom homogeneity. Results showed that values for the CTDIw measured in the nylon phantoms were slightly higher than those from the PMMA while CT numbers for nylon were lower than those of PMMA. This is because the mass attenuation coefficient of the nylon is higher. Nylon could be used as a substitute material for CT dosimetry phantom to enable measurements and adjustment factors are given which could be used to estimate PMMA values for making comparisons with displayed values. |
---|