A Peculiar Case of a Floating Angio-Seal
PURPOSE: To show a complication of the use of an Angio-Seal™ closure device. CASE: We present a patient with a systolic murmur in his femoral artery after PCI. The murmur was caused by a dislocated Angio-Seal™, a vascular closure device. This was diagnosed by Doppler Ultrasound. The device was surgi...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Bentham Open
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4797688/ https://www.ncbi.nlm.nih.gov/pubmed/27053966 http://dx.doi.org/10.2174/1874192401610010044 |
_version_ | 1782422003677921280 |
---|---|
author | Dettmers, R.C. Bosman, W.M.P.F. van den Broek, M.A.J. Veger, H.T.C. Hedeman Joosten, P.Ph.A. Borger van der Burg, B.L.S. |
author_facet | Dettmers, R.C. Bosman, W.M.P.F. van den Broek, M.A.J. Veger, H.T.C. Hedeman Joosten, P.Ph.A. Borger van der Burg, B.L.S. |
author_sort | Dettmers, R.C. |
collection | PubMed |
description | PURPOSE: To show a complication of the use of an Angio-Seal™ closure device. CASE: We present a patient with a systolic murmur in his femoral artery after PCI. The murmur was caused by a dislocated Angio-Seal™, a vascular closure device. This was diagnosed by Doppler Ultrasound. The device was surgically removed. CONCLUSION: Vascular complications, such as lower limb ischemia, requiring surgical intervention tend to be higher after use of a vascular closure device. We advise routine physical examination of the puncture site after percutaneous closure with a vascular closure device, such as an Angio-Seal™. The removal of the device can be performed via an open or endoscopic approach, based on available experience. |
format | Online Article Text |
id | pubmed-4797688 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Bentham Open |
record_format | MEDLINE/PubMed |
spelling | pubmed-47976882016-04-06 A Peculiar Case of a Floating Angio-Seal Dettmers, R.C. Bosman, W.M.P.F. van den Broek, M.A.J. Veger, H.T.C. Hedeman Joosten, P.Ph.A. Borger van der Burg, B.L.S. Open Cardiovasc Med J Article PURPOSE: To show a complication of the use of an Angio-Seal™ closure device. CASE: We present a patient with a systolic murmur in his femoral artery after PCI. The murmur was caused by a dislocated Angio-Seal™, a vascular closure device. This was diagnosed by Doppler Ultrasound. The device was surgically removed. CONCLUSION: Vascular complications, such as lower limb ischemia, requiring surgical intervention tend to be higher after use of a vascular closure device. We advise routine physical examination of the puncture site after percutaneous closure with a vascular closure device, such as an Angio-Seal™. The removal of the device can be performed via an open or endoscopic approach, based on available experience. Bentham Open 2016-03-09 /pmc/articles/PMC4797688/ /pubmed/27053966 http://dx.doi.org/10.2174/1874192401610010044 Text en © Dettmers et al.; Licensee Bentham Open. https://creativecommons.org/licenses/by/4.0/legalcode This is an open access article licensed under the terms of the Creative Commons Attribution-Non-Commercial 4.0 International Public License (CC BY-NC 4.0) (https://creativecommons.org/licenses/by-nc/4.0/legalcode), which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited. |
spellingShingle | Article Dettmers, R.C. Bosman, W.M.P.F. van den Broek, M.A.J. Veger, H.T.C. Hedeman Joosten, P.Ph.A. Borger van der Burg, B.L.S. A Peculiar Case of a Floating Angio-Seal |
title | A Peculiar Case of a Floating Angio-Seal |
title_full | A Peculiar Case of a Floating Angio-Seal |
title_fullStr | A Peculiar Case of a Floating Angio-Seal |
title_full_unstemmed | A Peculiar Case of a Floating Angio-Seal |
title_short | A Peculiar Case of a Floating Angio-Seal |
title_sort | peculiar case of a floating angio-seal |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4797688/ https://www.ncbi.nlm.nih.gov/pubmed/27053966 http://dx.doi.org/10.2174/1874192401610010044 |
work_keys_str_mv | AT dettmersrc apeculiarcaseofafloatingangioseal AT bosmanwmpf apeculiarcaseofafloatingangioseal AT vandenbroekmaj apeculiarcaseofafloatingangioseal AT vegerhtc apeculiarcaseofafloatingangioseal AT hedemanjoostenppha apeculiarcaseofafloatingangioseal AT borgervanderburgbls apeculiarcaseofafloatingangioseal AT dettmersrc peculiarcaseofafloatingangioseal AT bosmanwmpf peculiarcaseofafloatingangioseal AT vandenbroekmaj peculiarcaseofafloatingangioseal AT vegerhtc peculiarcaseofafloatingangioseal AT hedemanjoostenppha peculiarcaseofafloatingangioseal AT borgervanderburgbls peculiarcaseofafloatingangioseal |