Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish
Haemoglobin concentration ([Hb]) is measured for a wide variety of animal studies. The use of point-of-care devices, such as the HemoCue, is becoming increasingly common because of their portability, relative ease of use and low cost. In this study, we aimed to determine whether the [Hb] of blue-ton...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4804733/ https://www.ncbi.nlm.nih.gov/pubmed/27293758 http://dx.doi.org/10.1093/conphys/cow006 |
Sumario: | Haemoglobin concentration ([Hb]) is measured for a wide variety of animal studies. The use of point-of-care devices, such as the HemoCue, is becoming increasingly common because of their portability, relative ease of use and low cost. In this study, we aimed to determine whether the [Hb] of blue-tongued skink (Tiliqua nigrolutea) blood can be determined accurately using the HemoCue and whether the HemoCue overestimates the [Hb] of reptile blood in a similar manner to fish blood. Additionally, we aimed to test whether ploidy affected [Hb] determined by the HemoCue using blood from diploid and triploid Atlantic salmon (Salmo salar). The HemoCue Hb 201(+) systematically overestimated [Hb] in both blue-tongued skinks and Atlantic salmon, and there was no difference between calibration equations determined for diploid or triploid salmon. The overestimation was systematic in both species and, as such, [Hb] determined by the HemoCue can be corrected using appropriate calibration equations. |
---|