Single session, intrauser repeatability of anterior chamber biometric and corneal pachy-volumetric parameters using a new Scheimpflug + Placido device
PURPOSE: To analyze single session, intrauser reliability of a Scheimpflug device for anterior chamber (AC) and corneal parameters. METHODS: In this observational study, 100 normal candidates underwent Scheimpflug analysis with Sirius 3D Rotating Scheimpflug Camera and Topography System (Costruzione...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4812002/ https://www.ncbi.nlm.nih.gov/pubmed/26338543 http://dx.doi.org/10.1016/j.optom.2015.07.001 |
Sumario: | PURPOSE: To analyze single session, intrauser reliability of a Scheimpflug device for anterior chamber (AC) and corneal parameters. METHODS: In this observational study, 100 normal candidates underwent Scheimpflug analysis with Sirius 3D Rotating Scheimpflug Camera and Topography System (Costruzione Strumenti Oftalmici, Italy). Two scans in dark room conditions were performed by the same experienced user. The candidates were asked to keep both eyes closed for 5 min before the scans. Exclusion criteria were previous ocular surgery, corneal scarring and anterior segment/posterior segment anomalies. Only the right eyes were used for the analysis. Both corneal (central, minimum, and apical thickness, volume, horizontal visible iris diameter, and apical curvature) and anterior chamber (volume, depth, angle, horizontal diameter) measurements were evaluated. RESULTS: There was no difference in the means of repeated measurements (p > 0.05, ANOVA). Intraclass correlations between the measures were high and ranged from 0.995–0.997 for corneal to 0.964–0.997 for anterior chamber (AC) parameters. The precision of repeatability measures (1.96 × Sw) was approximately 5 μ for the central and minimum corneal thickness, 8 μ for the apical corneal thickness, 0.06 mm for AC (anterior chamber) depth and less than 2° for the AC angle. CONCLUSIONS: Sirius Scheimpflug system has high repeatability for both corneal and AC parameters in normal eyes. |
---|