Assessing heating distribution by therapeutic ultrasound on bone phantoms and in vitro human samples using infrared thermography
BACKGROUND: Bioheat models have been proposed to predict heat distribution in multilayered biological tissues after therapeutic ultrasound (TUS) stimulation. However, evidence on its therapeutic benefit is still controversial for many clinical conditions. The aim of this study was to evaluate and to...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4820859/ https://www.ncbi.nlm.nih.gov/pubmed/27051520 http://dx.doi.org/10.1186/s40349-016-0058-7 |
Sumario: | BACKGROUND: Bioheat models have been proposed to predict heat distribution in multilayered biological tissues after therapeutic ultrasound (TUS) stimulation. However, evidence on its therapeutic benefit is still controversial for many clinical conditions. The aim of this study was to evaluate and to compare the TUS heating distribution on commercially available bone phantoms and in vitro femur and tibia human samples, at 1 MHz and several ultrasonic pulse regimens, by means of a thermographic image processing technique. METHODS: An infrared camera was used to capture an image after each 5-min 1-MHz TUS stimulation on bone phantoms, as well as in vitro femur and tibia samples (N = 10). An intensity-based processing algorithm was applied to estimate temperature distribution. Sections of five femurs in the coronal plane were also used for the evaluation of heat distribution inside the medullar canal. RESULTS: Temperature increased up to 8.2 and 9.8 °C for the femur and tibia, respectively. Moreover, the temperature increased up to 10.8 °C inside the medullar canal. Although temperature distributions inside the region of interest (ROI) were significantly different (p < 0.001), the average and standard deviation values for bone phantoms were more similar to the femur than to the tibia samples. Pulsed regimens caused lower increments in temperature than continuous sonication, as expected. CONCLUSIONS: Commercially available bone phantoms could be used in research focusing on thermal effects of ultrasound. Small differences in mean and standard deviation temperatures were observed between bone samples and phantoms. Temperature can reach more than 10 °C inside the medullar canal on a fixed probe position which may lead to severe cellular damage. |
---|