Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT)

BACKGROUND: This study is to report 1) the sensitivity of intensity modulated radiation therapy (IMRT) QA method for clinical volumetric modulated arc therapy (VMAT) plans with multi-leaf collimator (MLC) leaf errors that will not trigger MLC interlock during beam delivery; 2) the effect of non-beam...

Descripción completa

Detalles Bibliográficos
Autores principales: Xu, Zhengzheng, Wang, Iris Z., Kumaraswamy, Lalith K., Podgorsak, Matthew B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: De Gruyter 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4825334/
https://www.ncbi.nlm.nih.gov/pubmed/27069458
http://dx.doi.org/10.1515/raon-2016-0008
_version_ 1782426199293689856
author Xu, Zhengzheng
Wang, Iris Z.
Kumaraswamy, Lalith K.
Podgorsak, Matthew B.
author_facet Xu, Zhengzheng
Wang, Iris Z.
Kumaraswamy, Lalith K.
Podgorsak, Matthew B.
author_sort Xu, Zhengzheng
collection PubMed
description BACKGROUND: This study is to report 1) the sensitivity of intensity modulated radiation therapy (IMRT) QA method for clinical volumetric modulated arc therapy (VMAT) plans with multi-leaf collimator (MLC) leaf errors that will not trigger MLC interlock during beam delivery; 2) the effect of non-beam-hold MLC leaf errors on the quality of VMAT plan dose delivery. MATERIALS AND METHODS. Eleven VMAT plans were selected and modified using an in-house developed software. For each control point of a VMAT arc, MLC leaves with the highest speed (1.87-1.95 cm/s) were set to move at the maximal allowable speed (2.3 cm/s), which resulted in a leaf position difference of less than 2 mm. The modified plans were considered as ‘standard’ plans, and the original plans were treated as the ‘slowing MLC’ plans for simulating ‘standard’ plans with leaves moving at relatively lower speed. The measurement of each ‘slowing MLC’ plan using MapCHECK®2 was compared with calculated planar dose of the ‘standard’ plan with respect to absolute dose Van Dyk distance-to-agreement (DTA) comparisons using 3%/3 mm and 2%/2 mm criteria. RESULTS: All ‘slowing MLC’ plans passed the 90% pass rate threshold using 3%/3 mm criteria while one brain and three anal VMAT cases were below 90% with 2%/2 mm criteria. For ten out of eleven cases, DVH comparisons between ‘standard’ and ‘slowing MLC’ plans demonstrated minimal dosimetric changes in targets and organs-at-risk. CONCLUSIONS: For highly modulated VMAT plans, pass rate threshold (90%) using 3%/3mm criteria is not sensitive in detecting MLC leaf errors that will not trigger the MLC leaf interlock. However, the consequential effects of non-beam hold MLC errors on target and OAR doses are negligible, which supports the reliability of current patient-specific IMRT quality assurance (QA) method for VMAT plans.
format Online
Article
Text
id pubmed-4825334
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher De Gruyter
record_format MEDLINE/PubMed
spelling pubmed-48253342016-04-11 Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT) Xu, Zhengzheng Wang, Iris Z. Kumaraswamy, Lalith K. Podgorsak, Matthew B. Radiol Oncol Research Article BACKGROUND: This study is to report 1) the sensitivity of intensity modulated radiation therapy (IMRT) QA method for clinical volumetric modulated arc therapy (VMAT) plans with multi-leaf collimator (MLC) leaf errors that will not trigger MLC interlock during beam delivery; 2) the effect of non-beam-hold MLC leaf errors on the quality of VMAT plan dose delivery. MATERIALS AND METHODS. Eleven VMAT plans were selected and modified using an in-house developed software. For each control point of a VMAT arc, MLC leaves with the highest speed (1.87-1.95 cm/s) were set to move at the maximal allowable speed (2.3 cm/s), which resulted in a leaf position difference of less than 2 mm. The modified plans were considered as ‘standard’ plans, and the original plans were treated as the ‘slowing MLC’ plans for simulating ‘standard’ plans with leaves moving at relatively lower speed. The measurement of each ‘slowing MLC’ plan using MapCHECK®2 was compared with calculated planar dose of the ‘standard’ plan with respect to absolute dose Van Dyk distance-to-agreement (DTA) comparisons using 3%/3 mm and 2%/2 mm criteria. RESULTS: All ‘slowing MLC’ plans passed the 90% pass rate threshold using 3%/3 mm criteria while one brain and three anal VMAT cases were below 90% with 2%/2 mm criteria. For ten out of eleven cases, DVH comparisons between ‘standard’ and ‘slowing MLC’ plans demonstrated minimal dosimetric changes in targets and organs-at-risk. CONCLUSIONS: For highly modulated VMAT plans, pass rate threshold (90%) using 3%/3mm criteria is not sensitive in detecting MLC leaf errors that will not trigger the MLC leaf interlock. However, the consequential effects of non-beam hold MLC errors on target and OAR doses are negligible, which supports the reliability of current patient-specific IMRT quality assurance (QA) method for VMAT plans. De Gruyter 2016-02-16 /pmc/articles/PMC4825334/ /pubmed/27069458 http://dx.doi.org/10.1515/raon-2016-0008 Text en © 2016 Radiol Oncol
spellingShingle Research Article
Xu, Zhengzheng
Wang, Iris Z.
Kumaraswamy, Lalith K.
Podgorsak, Matthew B.
Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT)
title Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT)
title_full Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT)
title_fullStr Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT)
title_full_unstemmed Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT)
title_short Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT)
title_sort evaluation of dosimetric effect caused by slowing with multi-leaf collimator (mlc) leaves for volumetric modulated arc therapy (vmat)
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4825334/
https://www.ncbi.nlm.nih.gov/pubmed/27069458
http://dx.doi.org/10.1515/raon-2016-0008
work_keys_str_mv AT xuzhengzheng evaluationofdosimetriceffectcausedbyslowingwithmultileafcollimatormlcleavesforvolumetricmodulatedarctherapyvmat
AT wangirisz evaluationofdosimetriceffectcausedbyslowingwithmultileafcollimatormlcleavesforvolumetricmodulatedarctherapyvmat
AT kumaraswamylalithk evaluationofdosimetriceffectcausedbyslowingwithmultileafcollimatormlcleavesforvolumetricmodulatedarctherapyvmat
AT podgorsakmatthewb evaluationofdosimetriceffectcausedbyslowingwithmultileafcollimatormlcleavesforvolumetricmodulatedarctherapyvmat