Cargando…
Accuracy of delivered airway pressure and work of breathing estimation during proportional assist ventilation: a bench study
BACKGROUND: Proportional assist ventilation+ (PAV+) delivers airway pressure (P(aw)) in proportion to patient effort (P(mus)) by using the equation of motion of the respiratory system. PAV+ calculates automatically respiratory mechanics (elastance and resistance); the work of breathing (WOB) is esti...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Paris
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4830790/ https://www.ncbi.nlm.nih.gov/pubmed/27076185 http://dx.doi.org/10.1186/s13613-016-0131-y |
Sumario: | BACKGROUND: Proportional assist ventilation+ (PAV+) delivers airway pressure (P(aw)) in proportion to patient effort (P(mus)) by using the equation of motion of the respiratory system. PAV+ calculates automatically respiratory mechanics (elastance and resistance); the work of breathing (WOB) is estimated by the ventilator. The accuracy of P(mus) estimation and hence accuracy of the delivered P(aw) and WOB calculation have not been assessed. This study aimed at assessing the accuracy of delivered P(aw) and calculated WOB by PAV+ and examining the factors influencing this accuracy. METHODS: Using an active lung model with different respiratory mechanics, we compared (1) the actual delivered P(aw) by the ventilator to the theoretical P(aw) as defined by the equation of motion and (2) the WOB value displayed by the ventilator to the WOB measured from a Campbell diagram. RESULTS: Irrespective of respiratory mechanics and gain, the ventilator provided a P(aw) approximately 25 % lower than expected. This underassistance was greatest at the beginning of the inspiration. Intrinsic PEEP (PEEPi), associated with an increase in trigger delay, was a major factor affecting PAV+ accuracy. The absolute value of total WOB displayed by the ventilator was underestimated, but the changes in WOB were accurately detected by the ventilator. CONCLUSION: The assistance provided by PAV+ well follows P(mus) but with a constant underassistance. This is associated with an underestimation by the ventilator of the WOB. PEEPi can be a major factor contributing to PAV+ inaccuracy. Clinical recommendations should include using a high trigger sensitivity and a careful PEEP titration. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s13613-016-0131-y) contains supplementary material, which is available to authorized users. |
---|