Accuracy of delivered airway pressure and work of breathing estimation during proportional assist ventilation: a bench study

BACKGROUND: Proportional assist ventilation+ (PAV+) delivers airway pressure (P(aw)) in proportion to patient effort (P(mus)) by using the equation of motion of the respiratory system. PAV+ calculates automatically respiratory mechanics (elastance and resistance); the work of breathing (WOB) is esti...

Descripción completa

Detalles Bibliográficos
Autores principales: Beloncle, Francois, Akoumianaki, Evangelia, Rittayamai, Nuttapol, Lyazidi, Aissam, Brochard, Laurent
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Paris 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4830790/
https://www.ncbi.nlm.nih.gov/pubmed/27076185
http://dx.doi.org/10.1186/s13613-016-0131-y
_version_ 1782426948290478080
author Beloncle, Francois
Akoumianaki, Evangelia
Rittayamai, Nuttapol
Lyazidi, Aissam
Brochard, Laurent
author_facet Beloncle, Francois
Akoumianaki, Evangelia
Rittayamai, Nuttapol
Lyazidi, Aissam
Brochard, Laurent
author_sort Beloncle, Francois
collection PubMed
description BACKGROUND: Proportional assist ventilation+ (PAV+) delivers airway pressure (P(aw)) in proportion to patient effort (P(mus)) by using the equation of motion of the respiratory system. PAV+ calculates automatically respiratory mechanics (elastance and resistance); the work of breathing (WOB) is estimated by the ventilator. The accuracy of P(mus) estimation and hence accuracy of the delivered P(aw) and WOB calculation have not been assessed. This study aimed at assessing the accuracy of delivered P(aw) and calculated WOB by PAV+ and examining the factors influencing this accuracy. METHODS: Using an active lung model with different respiratory mechanics, we compared (1) the actual delivered P(aw) by the ventilator to the theoretical P(aw) as defined by the equation of motion and (2) the WOB value displayed by the ventilator to the WOB measured from a Campbell diagram. RESULTS: Irrespective of respiratory mechanics and gain, the ventilator provided a P(aw) approximately 25 % lower than expected. This underassistance was greatest at the beginning of the inspiration. Intrinsic PEEP (PEEPi), associated with an increase in trigger delay, was a major factor affecting PAV+ accuracy. The absolute value of total WOB displayed by the ventilator was underestimated, but the changes in WOB were accurately detected by the ventilator. CONCLUSION: The assistance provided by PAV+ well follows P(mus) but with a constant underassistance. This is associated with an underestimation by the ventilator of the WOB. PEEPi can be a major factor contributing to PAV+ inaccuracy. Clinical recommendations should include using a high trigger sensitivity and a careful PEEP titration. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s13613-016-0131-y) contains supplementary material, which is available to authorized users.
format Online
Article
Text
id pubmed-4830790
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Springer Paris
record_format MEDLINE/PubMed
spelling pubmed-48307902016-04-26 Accuracy of delivered airway pressure and work of breathing estimation during proportional assist ventilation: a bench study Beloncle, Francois Akoumianaki, Evangelia Rittayamai, Nuttapol Lyazidi, Aissam Brochard, Laurent Ann Intensive Care Research BACKGROUND: Proportional assist ventilation+ (PAV+) delivers airway pressure (P(aw)) in proportion to patient effort (P(mus)) by using the equation of motion of the respiratory system. PAV+ calculates automatically respiratory mechanics (elastance and resistance); the work of breathing (WOB) is estimated by the ventilator. The accuracy of P(mus) estimation and hence accuracy of the delivered P(aw) and WOB calculation have not been assessed. This study aimed at assessing the accuracy of delivered P(aw) and calculated WOB by PAV+ and examining the factors influencing this accuracy. METHODS: Using an active lung model with different respiratory mechanics, we compared (1) the actual delivered P(aw) by the ventilator to the theoretical P(aw) as defined by the equation of motion and (2) the WOB value displayed by the ventilator to the WOB measured from a Campbell diagram. RESULTS: Irrespective of respiratory mechanics and gain, the ventilator provided a P(aw) approximately 25 % lower than expected. This underassistance was greatest at the beginning of the inspiration. Intrinsic PEEP (PEEPi), associated with an increase in trigger delay, was a major factor affecting PAV+ accuracy. The absolute value of total WOB displayed by the ventilator was underestimated, but the changes in WOB were accurately detected by the ventilator. CONCLUSION: The assistance provided by PAV+ well follows P(mus) but with a constant underassistance. This is associated with an underestimation by the ventilator of the WOB. PEEPi can be a major factor contributing to PAV+ inaccuracy. Clinical recommendations should include using a high trigger sensitivity and a careful PEEP titration. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s13613-016-0131-y) contains supplementary material, which is available to authorized users. Springer Paris 2016-04-14 /pmc/articles/PMC4830790/ /pubmed/27076185 http://dx.doi.org/10.1186/s13613-016-0131-y Text en © Beloncle et al. 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Research
Beloncle, Francois
Akoumianaki, Evangelia
Rittayamai, Nuttapol
Lyazidi, Aissam
Brochard, Laurent
Accuracy of delivered airway pressure and work of breathing estimation during proportional assist ventilation: a bench study
title Accuracy of delivered airway pressure and work of breathing estimation during proportional assist ventilation: a bench study
title_full Accuracy of delivered airway pressure and work of breathing estimation during proportional assist ventilation: a bench study
title_fullStr Accuracy of delivered airway pressure and work of breathing estimation during proportional assist ventilation: a bench study
title_full_unstemmed Accuracy of delivered airway pressure and work of breathing estimation during proportional assist ventilation: a bench study
title_short Accuracy of delivered airway pressure and work of breathing estimation during proportional assist ventilation: a bench study
title_sort accuracy of delivered airway pressure and work of breathing estimation during proportional assist ventilation: a bench study
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4830790/
https://www.ncbi.nlm.nih.gov/pubmed/27076185
http://dx.doi.org/10.1186/s13613-016-0131-y
work_keys_str_mv AT belonclefrancois accuracyofdeliveredairwaypressureandworkofbreathingestimationduringproportionalassistventilationabenchstudy
AT akoumianakievangelia accuracyofdeliveredairwaypressureandworkofbreathingestimationduringproportionalassistventilationabenchstudy
AT rittayamainuttapol accuracyofdeliveredairwaypressureandworkofbreathingestimationduringproportionalassistventilationabenchstudy
AT lyazidiaissam accuracyofdeliveredairwaypressureandworkofbreathingestimationduringproportionalassistventilationabenchstudy
AT brochardlaurent accuracyofdeliveredairwaypressureandworkofbreathingestimationduringproportionalassistventilationabenchstudy