Effect of Flow Rate on In Vitro Aerodynamic Performance of NEXThaler(®) in Comparison with Diskus(®) and Turbohaler(®) Dry Powder Inhalers

Background: European and United States Pharmacopoeia compendial procedures for assessing the in vitro emitted dose and aerodynamic size distribution of a dry powder inhaler require that 4.0 L of air at a pressure drop of 4 kPa be drawn through the inhaler. However, the product performance should be...

Descripción completa

Detalles Bibliográficos
Autores principales: Buttini, Francesca, Brambilla, Gaetano, Copelli, Diego, Sisti, Viviana, Balducci, Anna Giulia, Bettini, Ruggero, Pasquali, Irene
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Mary Ann Liebert, Inc. 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4841907/
https://www.ncbi.nlm.nih.gov/pubmed/26355743
http://dx.doi.org/10.1089/jamp.2015.1220
_version_ 1782428446884888576
author Buttini, Francesca
Brambilla, Gaetano
Copelli, Diego
Sisti, Viviana
Balducci, Anna Giulia
Bettini, Ruggero
Pasquali, Irene
author_facet Buttini, Francesca
Brambilla, Gaetano
Copelli, Diego
Sisti, Viviana
Balducci, Anna Giulia
Bettini, Ruggero
Pasquali, Irene
author_sort Buttini, Francesca
collection PubMed
description Background: European and United States Pharmacopoeia compendial procedures for assessing the in vitro emitted dose and aerodynamic size distribution of a dry powder inhaler require that 4.0 L of air at a pressure drop of 4 kPa be drawn through the inhaler. However, the product performance should be investigated using conditions more representative of what is achievable by the patient population. This work compares the delivered dose and the drug deposition profile at different flow rates (30, 40, 60, and 90 L/min) of Foster NEXThaler(®) (beclomethasone dipropionate/formoterol fumarate), Seretide(®) Diskus(®) (fluticasone propionate/salmeterol xinafoate), and Symbicort(®) Turbohaler(®) (budesonide/formoterol fumarate). Methods: The delivered dose uniformity was tested using a dose unit sampling apparatus (DUSA) at inhalation volumes either 2.0 or 4.0 L and flow rates 30, 40, 60, or 90 L/min. The aerodynamic assessment was carried out using a Next Generation Impactor by discharging each inhaler at 30, 40, 60, or 90 L/min for a time sufficient to obtain an air volume of 4 L. Results: Foster(®) NEXThaler(®) and Seretide(®) Diskus(®) showed a consistent dose delivery for both the drugs included in the formulation, independently of the applied flow rate. Contrary, Symbicort(®) Turbohaler(®) showed a high decrease of the emitted dose for both budesonide and formoterol fumarate when the device was operated at airflow rate lower that 60 L/min. The aerosolizing performance of NEXThaler(®) and Diskus(®) was unaffected by the flow rate applied. Turbohaler(®) proved to be the inhaler most sensitive to changes in flow rate in terms of fine particle fraction (FPF) for both components. Among the combinations tested, Foster NEXThaler(®) was the only one capable to deliver around 50% of extra-fine particles relative to delivered dose. Conclusions: NEXThaler(®) and Diskus(®) were substantially unaffected by flow rate through the inhaler in terms of both delivered dose and fine particle mass.
format Online
Article
Text
id pubmed-4841907
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Mary Ann Liebert, Inc.
record_format MEDLINE/PubMed
spelling pubmed-48419072016-05-06 Effect of Flow Rate on In Vitro Aerodynamic Performance of NEXThaler(®) in Comparison with Diskus(®) and Turbohaler(®) Dry Powder Inhalers Buttini, Francesca Brambilla, Gaetano Copelli, Diego Sisti, Viviana Balducci, Anna Giulia Bettini, Ruggero Pasquali, Irene J Aerosol Med Pulm Drug Deliv Original Research Background: European and United States Pharmacopoeia compendial procedures for assessing the in vitro emitted dose and aerodynamic size distribution of a dry powder inhaler require that 4.0 L of air at a pressure drop of 4 kPa be drawn through the inhaler. However, the product performance should be investigated using conditions more representative of what is achievable by the patient population. This work compares the delivered dose and the drug deposition profile at different flow rates (30, 40, 60, and 90 L/min) of Foster NEXThaler(®) (beclomethasone dipropionate/formoterol fumarate), Seretide(®) Diskus(®) (fluticasone propionate/salmeterol xinafoate), and Symbicort(®) Turbohaler(®) (budesonide/formoterol fumarate). Methods: The delivered dose uniformity was tested using a dose unit sampling apparatus (DUSA) at inhalation volumes either 2.0 or 4.0 L and flow rates 30, 40, 60, or 90 L/min. The aerodynamic assessment was carried out using a Next Generation Impactor by discharging each inhaler at 30, 40, 60, or 90 L/min for a time sufficient to obtain an air volume of 4 L. Results: Foster(®) NEXThaler(®) and Seretide(®) Diskus(®) showed a consistent dose delivery for both the drugs included in the formulation, independently of the applied flow rate. Contrary, Symbicort(®) Turbohaler(®) showed a high decrease of the emitted dose for both budesonide and formoterol fumarate when the device was operated at airflow rate lower that 60 L/min. The aerosolizing performance of NEXThaler(®) and Diskus(®) was unaffected by the flow rate applied. Turbohaler(®) proved to be the inhaler most sensitive to changes in flow rate in terms of fine particle fraction (FPF) for both components. Among the combinations tested, Foster NEXThaler(®) was the only one capable to deliver around 50% of extra-fine particles relative to delivered dose. Conclusions: NEXThaler(®) and Diskus(®) were substantially unaffected by flow rate through the inhaler in terms of both delivered dose and fine particle mass. Mary Ann Liebert, Inc. 2016-04-01 /pmc/articles/PMC4841907/ /pubmed/26355743 http://dx.doi.org/10.1089/jamp.2015.1220 Text en © The Author(s) 2015; Published by Mary Ann Liebert, Inc. This Open Access article is distributed under the terms of the Creative Commons License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited.
spellingShingle Original Research
Buttini, Francesca
Brambilla, Gaetano
Copelli, Diego
Sisti, Viviana
Balducci, Anna Giulia
Bettini, Ruggero
Pasquali, Irene
Effect of Flow Rate on In Vitro Aerodynamic Performance of NEXThaler(®) in Comparison with Diskus(®) and Turbohaler(®) Dry Powder Inhalers
title Effect of Flow Rate on In Vitro Aerodynamic Performance of NEXThaler(®) in Comparison with Diskus(®) and Turbohaler(®) Dry Powder Inhalers
title_full Effect of Flow Rate on In Vitro Aerodynamic Performance of NEXThaler(®) in Comparison with Diskus(®) and Turbohaler(®) Dry Powder Inhalers
title_fullStr Effect of Flow Rate on In Vitro Aerodynamic Performance of NEXThaler(®) in Comparison with Diskus(®) and Turbohaler(®) Dry Powder Inhalers
title_full_unstemmed Effect of Flow Rate on In Vitro Aerodynamic Performance of NEXThaler(®) in Comparison with Diskus(®) and Turbohaler(®) Dry Powder Inhalers
title_short Effect of Flow Rate on In Vitro Aerodynamic Performance of NEXThaler(®) in Comparison with Diskus(®) and Turbohaler(®) Dry Powder Inhalers
title_sort effect of flow rate on in vitro aerodynamic performance of nexthaler(®) in comparison with diskus(®) and turbohaler(®) dry powder inhalers
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4841907/
https://www.ncbi.nlm.nih.gov/pubmed/26355743
http://dx.doi.org/10.1089/jamp.2015.1220
work_keys_str_mv AT buttinifrancesca effectofflowrateoninvitroaerodynamicperformanceofnexthalerincomparisonwithdiskusandturbohalerdrypowderinhalers
AT brambillagaetano effectofflowrateoninvitroaerodynamicperformanceofnexthalerincomparisonwithdiskusandturbohalerdrypowderinhalers
AT copellidiego effectofflowrateoninvitroaerodynamicperformanceofnexthalerincomparisonwithdiskusandturbohalerdrypowderinhalers
AT sistiviviana effectofflowrateoninvitroaerodynamicperformanceofnexthalerincomparisonwithdiskusandturbohalerdrypowderinhalers
AT balducciannagiulia effectofflowrateoninvitroaerodynamicperformanceofnexthalerincomparisonwithdiskusandturbohalerdrypowderinhalers
AT bettiniruggero effectofflowrateoninvitroaerodynamicperformanceofnexthalerincomparisonwithdiskusandturbohalerdrypowderinhalers
AT pasqualiirene effectofflowrateoninvitroaerodynamicperformanceofnexthalerincomparisonwithdiskusandturbohalerdrypowderinhalers