Open-cell Stent Deployment across the Wide Neck of a Large Middle Cerebral Aneurysm Using the Stent Anchor Technique
We describe a case of successful open-cell stent deployment across the wide neck of a large middle cerebral artery aneurysm using the stent anchor technique. A microcatheter was looped through the aneurysm and navigated into a distal vessel across the aneurysm neck. Although the loop of the microcat...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Korean Society of Cerebrovascular Surgeons; Society of Korean Endovascular Neurosurgeons
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4842907/ https://www.ncbi.nlm.nih.gov/pubmed/27114965 http://dx.doi.org/10.7461/jcen.2016.18.1.38 |
Sumario: | We describe a case of successful open-cell stent deployment across the wide neck of a large middle cerebral artery aneurysm using the stent anchor technique. A microcatheter was looped through the aneurysm and navigated into a distal vessel across the aneurysm neck. Although the loop of the microcatheter in the aneurysm straightened as it was gently withdrawn, the microcatheter again protruded into the aneurysm by open-cell stent navigation. The stent was partially deployed in a vessel distal to the aneurysm neck, withdrawn slowly to straighten the loop of the microcatheter in the aneurysm, and completely deployed across the aneurysm neck. After successful stent deployment, stent-assisted coil embolization was performed without complications. The stent anchor technique was successfully used to deploy an open-cell stent across the aneurysm neck in this case of microcatheter protrusion into the aneurysm during stent navigation. |
---|