A Primary Dead-Weight Tester for Pressures (0.05–1.0) MPa

Recent advances in technology on two fronts, 1) the fabrication of large-diameter pistons and cylinders with good geometry, and 2) the ability to measure the dimensions of these components with high accuracy, have allowed dead-weight testers at the National Institute of Standards and Technology (NIS...

Descripción completa

Detalles Bibliográficos
Autores principales: Jain, Kamlesh, Bowers, Walt, Schmidt, James W.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2003
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4844511/
https://www.ncbi.nlm.nih.gov/pubmed/27413600
http://dx.doi.org/10.6028/jres.108.013
_version_ 1782428786861539328
author Jain, Kamlesh
Bowers, Walt
Schmidt, James W.
author_facet Jain, Kamlesh
Bowers, Walt
Schmidt, James W.
author_sort Jain, Kamlesh
collection PubMed
description Recent advances in technology on two fronts, 1) the fabrication of large-diameter pistons and cylinders with good geometry, and 2) the ability to measure the dimensions of these components with high accuracy, have allowed dead-weight testers at the National Institute of Standards and Technology (NIST) to generate pressures that approach total relative uncertainties previously obtained only by manometers. This paper describes a 35 mm diameter piston/cylinder assembly (known within NIST as PG-39) that serves as a pressure standard in which both the piston and the cylinder have been accurately dimensioned by Physikalisch Technische Bundesanstalt (PTB). Both artifacts (piston and cylinder) appeared to be round within ±30 nm and straight within ±100 nm over a substantial fraction of their heights. Based on the diameters at 20 °C provided by PTB (±15 nm) and on the good geometry of the artifact, the relative uncertainties for the effective area were estimated to be about 2.2 × 10(−6) (1σ).
format Online
Article
Text
id pubmed-4844511
institution National Center for Biotechnology Information
language English
publishDate 2003
publisher [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
record_format MEDLINE/PubMed
spelling pubmed-48445112016-07-13 A Primary Dead-Weight Tester for Pressures (0.05–1.0) MPa Jain, Kamlesh Bowers, Walt Schmidt, James W. J Res Natl Inst Stand Technol Article Recent advances in technology on two fronts, 1) the fabrication of large-diameter pistons and cylinders with good geometry, and 2) the ability to measure the dimensions of these components with high accuracy, have allowed dead-weight testers at the National Institute of Standards and Technology (NIST) to generate pressures that approach total relative uncertainties previously obtained only by manometers. This paper describes a 35 mm diameter piston/cylinder assembly (known within NIST as PG-39) that serves as a pressure standard in which both the piston and the cylinder have been accurately dimensioned by Physikalisch Technische Bundesanstalt (PTB). Both artifacts (piston and cylinder) appeared to be round within ±30 nm and straight within ±100 nm over a substantial fraction of their heights. Based on the diameters at 20 °C provided by PTB (±15 nm) and on the good geometry of the artifact, the relative uncertainties for the effective area were estimated to be about 2.2 × 10(−6) (1σ). [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2003 2003-04-01 /pmc/articles/PMC4844511/ /pubmed/27413600 http://dx.doi.org/10.6028/jres.108.013 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Institute of Standards and Technology is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright.
spellingShingle Article
Jain, Kamlesh
Bowers, Walt
Schmidt, James W.
A Primary Dead-Weight Tester for Pressures (0.05–1.0) MPa
title A Primary Dead-Weight Tester for Pressures (0.05–1.0) MPa
title_full A Primary Dead-Weight Tester for Pressures (0.05–1.0) MPa
title_fullStr A Primary Dead-Weight Tester for Pressures (0.05–1.0) MPa
title_full_unstemmed A Primary Dead-Weight Tester for Pressures (0.05–1.0) MPa
title_short A Primary Dead-Weight Tester for Pressures (0.05–1.0) MPa
title_sort primary dead-weight tester for pressures (0.05–1.0) mpa
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4844511/
https://www.ncbi.nlm.nih.gov/pubmed/27413600
http://dx.doi.org/10.6028/jres.108.013
work_keys_str_mv AT jainkamlesh aprimarydeadweighttesterforpressures00510mpa
AT bowerswalt aprimarydeadweighttesterforpressures00510mpa
AT schmidtjamesw aprimarydeadweighttesterforpressures00510mpa
AT jainkamlesh primarydeadweighttesterforpressures00510mpa
AT bowerswalt primarydeadweighttesterforpressures00510mpa
AT schmidtjamesw primarydeadweighttesterforpressures00510mpa