Cargando…
Mechanical Failure of Revision Knee Prosthesis at both Femoral and Tibial Modular Metaphyseal Stem Junctions
INTRODUCTION: This is a report of a mechanical failure of an S-ROM revision total knee prosthesis. The prosthesis was used as a revision implant following deep peri-prosthetic infection in a morbidly obese male. The prosthesis failed on both the femoral and tibial sides at the modular metaphyseal st...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Indian Orthopaedic Research Group
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4845409/ https://www.ncbi.nlm.nih.gov/pubmed/27299124 http://dx.doi.org/10.13107/jocr.2250-0685.373 |
_version_ | 1782428938914496512 |
---|---|
author | Woodgate, Ian G Rooney, John Mulford, Johnathan S Gillies, R Mark |
author_facet | Woodgate, Ian G Rooney, John Mulford, Johnathan S Gillies, R Mark |
author_sort | Woodgate, Ian G |
collection | PubMed |
description | INTRODUCTION: This is a report of a mechanical failure of an S-ROM revision total knee prosthesis. The prosthesis was used as a revision implant following deep peri-prosthetic infection in a morbidly obese male. The prosthesis failed on both the femoral and tibial sides at the modular metaphyseal stem junctions and required a further revision using the same type of implant after infection was excluded. CASE PRESENTATION: A 57 year old male had previously undergone a left total knee arthroplasty in 1999 for osteoarthritis. He acquired a late deep peri-prosthetic infection with a multi-resistant Staphylococcus epidermidis. The organism was sensitive to vancomycin and rifampicin. A two stage revision was undertaken after clinical signs of infection had resolved and blood parameters had normalized. Intra-operative gram stain was negative for micro-organisms and frozen section of deep tissue was less than five polymorphs per high power field. A cemented S-ROM prosthesis was implanted using a coronal tibial osteotomy and a lateral release for exposure. After three years of the second stage of revision, the patient again presented to the orthopaedic department after reportedly falling on a wet floor six weeks ago. CONCLUSION: Radiographically, there was a broken tibial wire, osteolysis and pedestal formation around both the femoral and distal tibial stem extensions. The prosthesis was bent at the proximal tibial sleeve and stem junction. The prosthesis was considered loose with mechanical failure. At implant removal, it was noted that the femoral and tibial components at the modular metaphyseal sleeve-stem junction were fractured. Surgeons should be cautious in the use of these implants in morbidly obese patients where the stresses generated maybe above the yield stress of the material and the frictional forces that may overcome the modular taper junction’s locking mechanism. |
format | Online Article Text |
id | pubmed-4845409 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Indian Orthopaedic Research Group |
record_format | MEDLINE/PubMed |
spelling | pubmed-48454092016-06-13 Mechanical Failure of Revision Knee Prosthesis at both Femoral and Tibial Modular Metaphyseal Stem Junctions Woodgate, Ian G Rooney, John Mulford, Johnathan S Gillies, R Mark J Orthop Case Reports Case Report INTRODUCTION: This is a report of a mechanical failure of an S-ROM revision total knee prosthesis. The prosthesis was used as a revision implant following deep peri-prosthetic infection in a morbidly obese male. The prosthesis failed on both the femoral and tibial sides at the modular metaphyseal stem junctions and required a further revision using the same type of implant after infection was excluded. CASE PRESENTATION: A 57 year old male had previously undergone a left total knee arthroplasty in 1999 for osteoarthritis. He acquired a late deep peri-prosthetic infection with a multi-resistant Staphylococcus epidermidis. The organism was sensitive to vancomycin and rifampicin. A two stage revision was undertaken after clinical signs of infection had resolved and blood parameters had normalized. Intra-operative gram stain was negative for micro-organisms and frozen section of deep tissue was less than five polymorphs per high power field. A cemented S-ROM prosthesis was implanted using a coronal tibial osteotomy and a lateral release for exposure. After three years of the second stage of revision, the patient again presented to the orthopaedic department after reportedly falling on a wet floor six weeks ago. CONCLUSION: Radiographically, there was a broken tibial wire, osteolysis and pedestal formation around both the femoral and distal tibial stem extensions. The prosthesis was bent at the proximal tibial sleeve and stem junction. The prosthesis was considered loose with mechanical failure. At implant removal, it was noted that the femoral and tibial components at the modular metaphyseal sleeve-stem junction were fractured. Surgeons should be cautious in the use of these implants in morbidly obese patients where the stresses generated maybe above the yield stress of the material and the frictional forces that may overcome the modular taper junction’s locking mechanism. Indian Orthopaedic Research Group 2016 /pmc/articles/PMC4845409/ /pubmed/27299124 http://dx.doi.org/10.13107/jocr.2250-0685.373 Text en Copyright: © Indian Orthopaedic Research Group http://creativecommons.org/licenses/by-nc-sa/3.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc-sa/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Case Report Woodgate, Ian G Rooney, John Mulford, Johnathan S Gillies, R Mark Mechanical Failure of Revision Knee Prosthesis at both Femoral and Tibial Modular Metaphyseal Stem Junctions |
title | Mechanical Failure of Revision Knee Prosthesis at both Femoral and Tibial Modular Metaphyseal Stem Junctions |
title_full | Mechanical Failure of Revision Knee Prosthesis at both Femoral and Tibial Modular Metaphyseal Stem Junctions |
title_fullStr | Mechanical Failure of Revision Knee Prosthesis at both Femoral and Tibial Modular Metaphyseal Stem Junctions |
title_full_unstemmed | Mechanical Failure of Revision Knee Prosthesis at both Femoral and Tibial Modular Metaphyseal Stem Junctions |
title_short | Mechanical Failure of Revision Knee Prosthesis at both Femoral and Tibial Modular Metaphyseal Stem Junctions |
title_sort | mechanical failure of revision knee prosthesis at both femoral and tibial modular metaphyseal stem junctions |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4845409/ https://www.ncbi.nlm.nih.gov/pubmed/27299124 http://dx.doi.org/10.13107/jocr.2250-0685.373 |
work_keys_str_mv | AT woodgateiang mechanicalfailureofrevisionkneeprosthesisatbothfemoralandtibialmodularmetaphysealstemjunctions AT rooneyjohn mechanicalfailureofrevisionkneeprosthesisatbothfemoralandtibialmodularmetaphysealstemjunctions AT mulfordjohnathans mechanicalfailureofrevisionkneeprosthesisatbothfemoralandtibialmodularmetaphysealstemjunctions AT gilliesrmark mechanicalfailureofrevisionkneeprosthesisatbothfemoralandtibialmodularmetaphysealstemjunctions |