Cargando…
A Summary of Heat-Flux Sensor Calibration Data
This paper presents a statistical evaluation of the responsivity data on a number of heat-flux sensors, calibrated using an electrical substitution radiometer as a transfer standard up to 5 W·cm(−2). The sensors, furnished by the customers, were of circular-foil or thermopile type. Comparison of the...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
2005
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4847575/ https://www.ncbi.nlm.nih.gov/pubmed/27308106 http://dx.doi.org/10.6028/jres.110.009 |
_version_ | 1782429240536334336 |
---|---|
author | Murthy, A. V. Fraser, G. T. DeWitt, D. P. |
author_facet | Murthy, A. V. Fraser, G. T. DeWitt, D. P. |
author_sort | Murthy, A. V. |
collection | PubMed |
description | This paper presents a statistical evaluation of the responsivity data on a number of heat-flux sensors, calibrated using an electrical substitution radiometer as a transfer standard up to 5 W·cm(−2). The sensors, furnished by the customers, were of circular-foil or thermopile type. Comparison of the NIST and the customer measured responsivity values showed that the measurements agree within 3 % for more than half the number of sensors tested, so far. Considering the variation in the customer calibration techniques and the wide measuring range of the sensors used in the calibration, the agreement is encouraging. |
format | Online Article Text |
id | pubmed-4847575 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2005 |
publisher | [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology |
record_format | MEDLINE/PubMed |
spelling | pubmed-48475752016-06-15 A Summary of Heat-Flux Sensor Calibration Data Murthy, A. V. Fraser, G. T. DeWitt, D. P. J Res Natl Inst Stand Technol Article This paper presents a statistical evaluation of the responsivity data on a number of heat-flux sensors, calibrated using an electrical substitution radiometer as a transfer standard up to 5 W·cm(−2). The sensors, furnished by the customers, were of circular-foil or thermopile type. Comparison of the NIST and the customer measured responsivity values showed that the measurements agree within 3 % for more than half the number of sensors tested, so far. Considering the variation in the customer calibration techniques and the wide measuring range of the sensors used in the calibration, the agreement is encouraging. [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2005 2005-04-01 /pmc/articles/PMC4847575/ /pubmed/27308106 http://dx.doi.org/10.6028/jres.110.009 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Institute of Standards and Technology is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright. |
spellingShingle | Article Murthy, A. V. Fraser, G. T. DeWitt, D. P. A Summary of Heat-Flux Sensor Calibration Data |
title | A Summary of Heat-Flux Sensor Calibration Data |
title_full | A Summary of Heat-Flux Sensor Calibration Data |
title_fullStr | A Summary of Heat-Flux Sensor Calibration Data |
title_full_unstemmed | A Summary of Heat-Flux Sensor Calibration Data |
title_short | A Summary of Heat-Flux Sensor Calibration Data |
title_sort | summary of heat-flux sensor calibration data |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4847575/ https://www.ncbi.nlm.nih.gov/pubmed/27308106 http://dx.doi.org/10.6028/jres.110.009 |
work_keys_str_mv | AT murthyav asummaryofheatfluxsensorcalibrationdata AT frasergt asummaryofheatfluxsensorcalibrationdata AT dewittdp asummaryofheatfluxsensorcalibrationdata AT murthyav summaryofheatfluxsensorcalibrationdata AT frasergt summaryofheatfluxsensorcalibrationdata AT dewittdp summaryofheatfluxsensorcalibrationdata |