Cargando…
Comparison of the NIST and BIPM Medium-Energy X-Ray Air-Kerma Measurements
The air-kerma standards used for the measurement of medium-energy x rays were compared at the National Institute of Standards and Technology (NIST) and at the Bureau International des Poids et Mesures (BIPM). The comparison involved a series of measurements at the BIPM and the NIST using the air-ker...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
2003
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4847580/ https://www.ncbi.nlm.nih.gov/pubmed/27413616 http://dx.doi.org/10.6028/jres.108.032 |
_version_ | 1782429241644679168 |
---|---|
author | Burns, D. T. O’Brien, M. Lamperti, P. Boutillon, M. |
author_facet | Burns, D. T. O’Brien, M. Lamperti, P. Boutillon, M. |
author_sort | Burns, D. T. |
collection | PubMed |
description | The air-kerma standards used for the measurement of medium-energy x rays were compared at the National Institute of Standards and Technology (NIST) and at the Bureau International des Poids et Mesures (BIPM). The comparison involved a series of measurements at the BIPM and the NIST using the air-kerma standards and two NIST reference-class transfer ionization standards. Reference beam qualities in the range from 60 kV to 300 kV were used. The results show the standards to be in agreement within the combined standard uncertainty of the comparison of 0.35 %. |
format | Online Article Text |
id | pubmed-4847580 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2003 |
publisher | [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology |
record_format | MEDLINE/PubMed |
spelling | pubmed-48475802016-07-13 Comparison of the NIST and BIPM Medium-Energy X-Ray Air-Kerma Measurements Burns, D. T. O’Brien, M. Lamperti, P. Boutillon, M. J Res Natl Inst Stand Technol Article The air-kerma standards used for the measurement of medium-energy x rays were compared at the National Institute of Standards and Technology (NIST) and at the Bureau International des Poids et Mesures (BIPM). The comparison involved a series of measurements at the BIPM and the NIST using the air-kerma standards and two NIST reference-class transfer ionization standards. Reference beam qualities in the range from 60 kV to 300 kV were used. The results show the standards to be in agreement within the combined standard uncertainty of the comparison of 0.35 %. [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2003 2003-10-01 /pmc/articles/PMC4847580/ /pubmed/27413616 http://dx.doi.org/10.6028/jres.108.032 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Institute of Standards and Technology is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright. |
spellingShingle | Article Burns, D. T. O’Brien, M. Lamperti, P. Boutillon, M. Comparison of the NIST and BIPM Medium-Energy X-Ray Air-Kerma Measurements |
title | Comparison of the NIST and BIPM Medium-Energy X-Ray Air-Kerma Measurements |
title_full | Comparison of the NIST and BIPM Medium-Energy X-Ray Air-Kerma Measurements |
title_fullStr | Comparison of the NIST and BIPM Medium-Energy X-Ray Air-Kerma Measurements |
title_full_unstemmed | Comparison of the NIST and BIPM Medium-Energy X-Ray Air-Kerma Measurements |
title_short | Comparison of the NIST and BIPM Medium-Energy X-Ray Air-Kerma Measurements |
title_sort | comparison of the nist and bipm medium-energy x-ray air-kerma measurements |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4847580/ https://www.ncbi.nlm.nih.gov/pubmed/27413616 http://dx.doi.org/10.6028/jres.108.032 |
work_keys_str_mv | AT burnsdt comparisonofthenistandbipmmediumenergyxrayairkermameasurements AT obrienm comparisonofthenistandbipmmediumenergyxrayairkermameasurements AT lampertip comparisonofthenistandbipmmediumenergyxrayairkermameasurements AT boutillonm comparisonofthenistandbipmmediumenergyxrayairkermameasurements |