Optical-Fiber Power Meter Comparison Between NIST and PTB
We describe the results of a comparison of reference standards between the National Institute of Standards and Technology (NIST-USA) and Physikalisch-Technische Bundesanstalt (PTB-Germany) at nominal wavelengths of 1300 nm and 1550 nm using an optical-fiber cable. Both laboratories used thermal dete...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
2003
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4847584/ https://www.ncbi.nlm.nih.gov/pubmed/27413617 http://dx.doi.org/10.6028/jres.108.033 |
Sumario: | We describe the results of a comparison of reference standards between the National Institute of Standards and Technology (NIST-USA) and Physikalisch-Technische Bundesanstalt (PTB-Germany) at nominal wavelengths of 1300 nm and 1550 nm using an optical-fiber cable. Both laboratories used thermal detectors as reference standards. A novel temperature-controlled, optical-trap detector was used as a transfer standard to compare two reference standards. Measurement results showed differences of less than 1.5 × 10(−3), which is within the combined uncertainty for both laboratories. |
---|