Cargando…
Optical-Fiber Power Meter Comparison Between NIST and PTB
We describe the results of a comparison of reference standards between the National Institute of Standards and Technology (NIST-USA) and Physikalisch-Technische Bundesanstalt (PTB-Germany) at nominal wavelengths of 1300 nm and 1550 nm using an optical-fiber cable. Both laboratories used thermal dete...
Autores principales: | Vayshenker, I., Haars, H., Li, X., Lehman, J. H., Livigni, D. J. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
2003
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4847584/ https://www.ncbi.nlm.nih.gov/pubmed/27413617 http://dx.doi.org/10.6028/jres.108.033 |