Automatic Real-Time Embedded QRS Complex Detection for a Novel Patch-Type Electrocardiogram Recorder

Cardiovascular diseases are projected to remain the single leading cause of death globally. Timely diagnosis and treatment of these diseases are crucial to prevent death and dangerous complications. One of the important tools in early diagnosis of arrhythmias is analysis of electrocardiograms (ECGs)...

Descripción completa

Detalles Bibliográficos
Formato: Online Artículo Texto
Lenguaje:English
Publicado: IEEE 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4848097/
https://www.ncbi.nlm.nih.gov/pubmed/27170891
http://dx.doi.org/10.1109/JTEHM.2015.2421901
_version_ 1782429300148928512
collection PubMed
description Cardiovascular diseases are projected to remain the single leading cause of death globally. Timely diagnosis and treatment of these diseases are crucial to prevent death and dangerous complications. One of the important tools in early diagnosis of arrhythmias is analysis of electrocardiograms (ECGs) obtained from ambulatory long-term recordings. The design of novel patch-type ECG recorders has increased the accessibility of these long-term recordings. In many applications, it is furthermore an advantage for these devices that the recorded ECGs can be analyzed automatically in real time. The purpose of this study was therefore to design a novel algorithm for automatic heart beat detection, and embed the algorithm in the CE marked ePatch heart monitor. The algorithm is based on a novel cascade of computationally efficient filters, optimized adaptive thresholding, and a refined search back mechanism. The design and optimization of the algorithm was performed on two different databases: The MIT-BIH arrhythmia database ([Formula: see text] %, [Formula: see text]) and a private ePatch training database ([Formula: see text] %, [Formula: see text] %). The offline validation was conducted on the European ST-T database ([Formula: see text] %, [Formula: see text] %). Finally, a double-blinded validation of the embedded algorithm was conducted on a private ePatch validation database ([Formula: see text] %, [Formula: see text] %). The algorithm was thus validated with high clinical performance on more than 300 ECG records from 189 different subjects with a high number of different abnormal beat morphologies. This demonstrates the strengths of the algorithm, and the potential for this embedded algorithm to improve the possibilities of early diagnosis and treatment of cardiovascular diseases.
format Online
Article
Text
id pubmed-4848097
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher IEEE
record_format MEDLINE/PubMed
spelling pubmed-48480972016-05-11 Automatic Real-Time Embedded QRS Complex Detection for a Novel Patch-Type Electrocardiogram Recorder IEEE J Transl Eng Health Med Article Cardiovascular diseases are projected to remain the single leading cause of death globally. Timely diagnosis and treatment of these diseases are crucial to prevent death and dangerous complications. One of the important tools in early diagnosis of arrhythmias is analysis of electrocardiograms (ECGs) obtained from ambulatory long-term recordings. The design of novel patch-type ECG recorders has increased the accessibility of these long-term recordings. In many applications, it is furthermore an advantage for these devices that the recorded ECGs can be analyzed automatically in real time. The purpose of this study was therefore to design a novel algorithm for automatic heart beat detection, and embed the algorithm in the CE marked ePatch heart monitor. The algorithm is based on a novel cascade of computationally efficient filters, optimized adaptive thresholding, and a refined search back mechanism. The design and optimization of the algorithm was performed on two different databases: The MIT-BIH arrhythmia database ([Formula: see text] %, [Formula: see text]) and a private ePatch training database ([Formula: see text] %, [Formula: see text] %). The offline validation was conducted on the European ST-T database ([Formula: see text] %, [Formula: see text] %). Finally, a double-blinded validation of the embedded algorithm was conducted on a private ePatch validation database ([Formula: see text] %, [Formula: see text] %). The algorithm was thus validated with high clinical performance on more than 300 ECG records from 189 different subjects with a high number of different abnormal beat morphologies. This demonstrates the strengths of the algorithm, and the potential for this embedded algorithm to improve the possibilities of early diagnosis and treatment of cardiovascular diseases. IEEE 2015-04-10 /pmc/articles/PMC4848097/ /pubmed/27170891 http://dx.doi.org/10.1109/JTEHM.2015.2421901 Text en 2168-2372 © 2015 IEEE. Translations and content mining are permitted for academic research only. Personal use is also permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
spellingShingle Article
Automatic Real-Time Embedded QRS Complex Detection for a Novel Patch-Type Electrocardiogram Recorder
title Automatic Real-Time Embedded QRS Complex Detection for a Novel Patch-Type Electrocardiogram Recorder
title_full Automatic Real-Time Embedded QRS Complex Detection for a Novel Patch-Type Electrocardiogram Recorder
title_fullStr Automatic Real-Time Embedded QRS Complex Detection for a Novel Patch-Type Electrocardiogram Recorder
title_full_unstemmed Automatic Real-Time Embedded QRS Complex Detection for a Novel Patch-Type Electrocardiogram Recorder
title_short Automatic Real-Time Embedded QRS Complex Detection for a Novel Patch-Type Electrocardiogram Recorder
title_sort automatic real-time embedded qrs complex detection for a novel patch-type electrocardiogram recorder
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4848097/
https://www.ncbi.nlm.nih.gov/pubmed/27170891
http://dx.doi.org/10.1109/JTEHM.2015.2421901
work_keys_str_mv AT automaticrealtimeembeddedqrscomplexdetectionforanovelpatchtypeelectrocardiogramrecorder
AT automaticrealtimeembeddedqrscomplexdetectionforanovelpatchtypeelectrocardiogramrecorder
AT automaticrealtimeembeddedqrscomplexdetectionforanovelpatchtypeelectrocardiogramrecorder
AT automaticrealtimeembeddedqrscomplexdetectionforanovelpatchtypeelectrocardiogramrecorder
AT automaticrealtimeembeddedqrscomplexdetectionforanovelpatchtypeelectrocardiogramrecorder
AT automaticrealtimeembeddedqrscomplexdetectionforanovelpatchtypeelectrocardiogramrecorder
AT automaticrealtimeembeddedqrscomplexdetectionforanovelpatchtypeelectrocardiogramrecorder
AT automaticrealtimeembeddedqrscomplexdetectionforanovelpatchtypeelectrocardiogramrecorder
AT automaticrealtimeembeddedqrscomplexdetectionforanovelpatchtypeelectrocardiogramrecorder
AT automaticrealtimeembeddedqrscomplexdetectionforanovelpatchtypeelectrocardiogramrecorder