Cargando…

Calibration of PET/CT scanners for multicenter studies on differentiated thyroid cancer with (124)I

BACKGROUND: Studies on imaging of differentiated thyroid cancer (DTC) using (124)I often require a multicenter approach, as the prevalence of DTC is low. Calibration of participating scanners is required to obtain comparable quantification. As determination of a well-defined range of recovery coeffi...

Descripción completa

Detalles Bibliográficos
Autores principales: Kist, Jakob W., van der Vlies, Manfred, Hoekstra, Otto S., Greuter, Henri N. J. M., de Keizer, Bart, Stokkel, Marcel P. M., V. Vogel, Wouter, Huisman, Marc C., van Lingen, Arthur
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4848279/
https://www.ncbi.nlm.nih.gov/pubmed/27118538
http://dx.doi.org/10.1186/s13550-016-0191-x
_version_ 1782429308340404224
author Kist, Jakob W.
van der Vlies, Manfred
Hoekstra, Otto S.
Greuter, Henri N. J. M.
de Keizer, Bart
Stokkel, Marcel P. M.
V. Vogel, Wouter
Huisman, Marc C.
van Lingen, Arthur
author_facet Kist, Jakob W.
van der Vlies, Manfred
Hoekstra, Otto S.
Greuter, Henri N. J. M.
de Keizer, Bart
Stokkel, Marcel P. M.
V. Vogel, Wouter
Huisman, Marc C.
van Lingen, Arthur
author_sort Kist, Jakob W.
collection PubMed
description BACKGROUND: Studies on imaging of differentiated thyroid cancer (DTC) using (124)I often require a multicenter approach, as the prevalence of DTC is low. Calibration of participating scanners is required to obtain comparable quantification. As determination of a well-defined range of recovery coefficients is complicated for various reasons, a simpler approach based on the assumption that the iodine uptake is highly focal with a background that significantly lacks radioactivity might be more efficient. For each scanner, a linear conversion between known and observed activity can be derived, allowing quantification that can be traced to a common source for all scanners within one study-protocol. The aim of this paper is to outline a procedure using this approach in order to set up a multicenter calibration of PET/CT scanners for (124)I. METHODS: A cylindrical polyethylene phantom contained six 2-ml vials with reference activities of ~2, 10, 20, 100, 400, and 2000 kBq, produced by dilution from a known activity. The phantom was scanned twice on PET/CT scanners of participating centers within 1 week. For each scanner, the best proportional and linear fit between measured and known activities were derived and based on statistical analyses of the results of all scanners; it was determined which fit should be applied. In addition, a Bland-Altman analysis was done on calibrated activities with respect to reference activities to asses the relative precision of the scanners. RESULTS: Nine Philips (vendor A) and nine Siemens (vendor B) PET/CT scanners were calibrated in a time period of 3 days before and after the reference time. No significant differences were detected between the two subsequent scans on any scanner. Six fitted intercepts of vendor A were significantly different from zero, so the linear model was used. Intercepts ranged from −8 to 26 kBq and slopes ranged from 0.80 to 0.98. Bland-Altman analysis of calibrated and reference activities showed that the relative error of calibrated activities was smaller than that of uncalibrated activities. CONCLUSIONS: A simplified multicenter calibration procedure for PET/CT scans that show highly focal uptake and negligible background is feasible and results in more precise quantification. Our procedure can be used in multicenter (124)I PET scans focusing on (recurrent) DTC.
format Online
Article
Text
id pubmed-4848279
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-48482792016-05-16 Calibration of PET/CT scanners for multicenter studies on differentiated thyroid cancer with (124)I Kist, Jakob W. van der Vlies, Manfred Hoekstra, Otto S. Greuter, Henri N. J. M. de Keizer, Bart Stokkel, Marcel P. M. V. Vogel, Wouter Huisman, Marc C. van Lingen, Arthur EJNMMI Res Original Research BACKGROUND: Studies on imaging of differentiated thyroid cancer (DTC) using (124)I often require a multicenter approach, as the prevalence of DTC is low. Calibration of participating scanners is required to obtain comparable quantification. As determination of a well-defined range of recovery coefficients is complicated for various reasons, a simpler approach based on the assumption that the iodine uptake is highly focal with a background that significantly lacks radioactivity might be more efficient. For each scanner, a linear conversion between known and observed activity can be derived, allowing quantification that can be traced to a common source for all scanners within one study-protocol. The aim of this paper is to outline a procedure using this approach in order to set up a multicenter calibration of PET/CT scanners for (124)I. METHODS: A cylindrical polyethylene phantom contained six 2-ml vials with reference activities of ~2, 10, 20, 100, 400, and 2000 kBq, produced by dilution from a known activity. The phantom was scanned twice on PET/CT scanners of participating centers within 1 week. For each scanner, the best proportional and linear fit between measured and known activities were derived and based on statistical analyses of the results of all scanners; it was determined which fit should be applied. In addition, a Bland-Altman analysis was done on calibrated activities with respect to reference activities to asses the relative precision of the scanners. RESULTS: Nine Philips (vendor A) and nine Siemens (vendor B) PET/CT scanners were calibrated in a time period of 3 days before and after the reference time. No significant differences were detected between the two subsequent scans on any scanner. Six fitted intercepts of vendor A were significantly different from zero, so the linear model was used. Intercepts ranged from −8 to 26 kBq and slopes ranged from 0.80 to 0.98. Bland-Altman analysis of calibrated and reference activities showed that the relative error of calibrated activities was smaller than that of uncalibrated activities. CONCLUSIONS: A simplified multicenter calibration procedure for PET/CT scans that show highly focal uptake and negligible background is feasible and results in more precise quantification. Our procedure can be used in multicenter (124)I PET scans focusing on (recurrent) DTC. Springer Berlin Heidelberg 2016-04-27 /pmc/articles/PMC4848279/ /pubmed/27118538 http://dx.doi.org/10.1186/s13550-016-0191-x Text en © Kist et al. 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Research
Kist, Jakob W.
van der Vlies, Manfred
Hoekstra, Otto S.
Greuter, Henri N. J. M.
de Keizer, Bart
Stokkel, Marcel P. M.
V. Vogel, Wouter
Huisman, Marc C.
van Lingen, Arthur
Calibration of PET/CT scanners for multicenter studies on differentiated thyroid cancer with (124)I
title Calibration of PET/CT scanners for multicenter studies on differentiated thyroid cancer with (124)I
title_full Calibration of PET/CT scanners for multicenter studies on differentiated thyroid cancer with (124)I
title_fullStr Calibration of PET/CT scanners for multicenter studies on differentiated thyroid cancer with (124)I
title_full_unstemmed Calibration of PET/CT scanners for multicenter studies on differentiated thyroid cancer with (124)I
title_short Calibration of PET/CT scanners for multicenter studies on differentiated thyroid cancer with (124)I
title_sort calibration of pet/ct scanners for multicenter studies on differentiated thyroid cancer with (124)i
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4848279/
https://www.ncbi.nlm.nih.gov/pubmed/27118538
http://dx.doi.org/10.1186/s13550-016-0191-x
work_keys_str_mv AT kistjakobw calibrationofpetctscannersformulticenterstudiesondifferentiatedthyroidcancerwith124i
AT vandervliesmanfred calibrationofpetctscannersformulticenterstudiesondifferentiatedthyroidcancerwith124i
AT hoekstraottos calibrationofpetctscannersformulticenterstudiesondifferentiatedthyroidcancerwith124i
AT greuterhenrinjm calibrationofpetctscannersformulticenterstudiesondifferentiatedthyroidcancerwith124i
AT dekeizerbart calibrationofpetctscannersformulticenterstudiesondifferentiatedthyroidcancerwith124i
AT stokkelmarcelpm calibrationofpetctscannersformulticenterstudiesondifferentiatedthyroidcancerwith124i
AT vvogelwouter calibrationofpetctscannersformulticenterstudiesondifferentiatedthyroidcancerwith124i
AT huismanmarcc calibrationofpetctscannersformulticenterstudiesondifferentiatedthyroidcancerwith124i
AT vanlingenarthur calibrationofpetctscannersformulticenterstudiesondifferentiatedthyroidcancerwith124i