Quantification of cephalomedullary nail fit in the femur using 3D computer modelling: a comparison between 1.0 and 1.5m bow designs

BACKGROUND: The radius of curvature (ROC) misfit of cephalomedullary nails during anterograde nailing can lead to complications such as distal anterior cortical encroachment. This study quantified the anatomical fit of a new nail with 1.0-m ROC (TFN-ADVANCED(™) Proximal Femoral Nailing System [TFNA]...

Descripción completa

Detalles Bibliográficos
Autores principales: Schmutz, Beat, Amarathunga, Jayani, Kmiec, Stanley, Yarlagadda, Prasad, Schuetz, Michael
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4848874/
https://www.ncbi.nlm.nih.gov/pubmed/27122018
http://dx.doi.org/10.1186/s13018-016-0389-7
_version_ 1782429437524967424
author Schmutz, Beat
Amarathunga, Jayani
Kmiec, Stanley
Yarlagadda, Prasad
Schuetz, Michael
author_facet Schmutz, Beat
Amarathunga, Jayani
Kmiec, Stanley
Yarlagadda, Prasad
Schuetz, Michael
author_sort Schmutz, Beat
collection PubMed
description BACKGROUND: The radius of curvature (ROC) misfit of cephalomedullary nails during anterograde nailing can lead to complications such as distal anterior cortical encroachment. This study quantified the anatomical fit of a new nail with 1.0-m ROC (TFN-ADVANCED(™) Proximal Femoral Nailing System [TFNA]) compared with a nail with 1.5-m ROC (Gamma3 Long Nail R1.5 [Gamma3]). METHODS: We generated 63 three-dimensional models (48 female, 45 right femur) representing the cortical surfaces of the femora (31 Caucasian, 28 Japanese, and 4 Thai). The mean age of the specimens was 77 years (±8.1), and the mean height was 158.5 cm (±9.6). Utilizing a customized software tool, nail fit was determined from the total surface area of nail protrusion from the inner cortex surface and maximum distance of nail protrusion in the axial plane; the position of the distal nail tip within the canal was also determined. RESULTS: Overall, TFNA had both a significantly smaller mean total surface area of nail protrusion (915.8 vs. 1181.6 mm(2); P < 0.05) and a mean maximum distance of nail protrusion in the axial plane (1.9 vs. 2.1 mm; P = 0.007) when compared with Gamma3. The mean total surface area of nail protrusion was significantly smaller with TFNA versus Gamma3 in both the Caucasian (P = 0.0009) and Asian (Japanese and Thai) samples (P = 0.000002); the mean maximum distance of TFNA protrusion was significantly smaller in Asians (P = 0.04), but not in Caucasians (P = 0.08). Most tip positions for both nail types were anterior, but TFNA had a higher number of center positions than Gamma3 (13 vs. 7) and a shift from the far anterior cortex to the center of the medullary canal (overall and in Caucasians). In Asians, the most prominent position was far anterior for both nails. CONCLUSIONS: The 1.0-m ROC TFNA nail resulted in better fit than the 1.5-m ROC Gamma3 nail. Clinical trials and case studies should be conducted in the future to verify if these findings would also result in clinical improvements.
format Online
Article
Text
id pubmed-4848874
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-48488742016-04-29 Quantification of cephalomedullary nail fit in the femur using 3D computer modelling: a comparison between 1.0 and 1.5m bow designs Schmutz, Beat Amarathunga, Jayani Kmiec, Stanley Yarlagadda, Prasad Schuetz, Michael J Orthop Surg Res Research Article BACKGROUND: The radius of curvature (ROC) misfit of cephalomedullary nails during anterograde nailing can lead to complications such as distal anterior cortical encroachment. This study quantified the anatomical fit of a new nail with 1.0-m ROC (TFN-ADVANCED(™) Proximal Femoral Nailing System [TFNA]) compared with a nail with 1.5-m ROC (Gamma3 Long Nail R1.5 [Gamma3]). METHODS: We generated 63 three-dimensional models (48 female, 45 right femur) representing the cortical surfaces of the femora (31 Caucasian, 28 Japanese, and 4 Thai). The mean age of the specimens was 77 years (±8.1), and the mean height was 158.5 cm (±9.6). Utilizing a customized software tool, nail fit was determined from the total surface area of nail protrusion from the inner cortex surface and maximum distance of nail protrusion in the axial plane; the position of the distal nail tip within the canal was also determined. RESULTS: Overall, TFNA had both a significantly smaller mean total surface area of nail protrusion (915.8 vs. 1181.6 mm(2); P < 0.05) and a mean maximum distance of nail protrusion in the axial plane (1.9 vs. 2.1 mm; P = 0.007) when compared with Gamma3. The mean total surface area of nail protrusion was significantly smaller with TFNA versus Gamma3 in both the Caucasian (P = 0.0009) and Asian (Japanese and Thai) samples (P = 0.000002); the mean maximum distance of TFNA protrusion was significantly smaller in Asians (P = 0.04), but not in Caucasians (P = 0.08). Most tip positions for both nail types were anterior, but TFNA had a higher number of center positions than Gamma3 (13 vs. 7) and a shift from the far anterior cortex to the center of the medullary canal (overall and in Caucasians). In Asians, the most prominent position was far anterior for both nails. CONCLUSIONS: The 1.0-m ROC TFNA nail resulted in better fit than the 1.5-m ROC Gamma3 nail. Clinical trials and case studies should be conducted in the future to verify if these findings would also result in clinical improvements. BioMed Central 2016-04-27 /pmc/articles/PMC4848874/ /pubmed/27122018 http://dx.doi.org/10.1186/s13018-016-0389-7 Text en © Schmutz et al. 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Schmutz, Beat
Amarathunga, Jayani
Kmiec, Stanley
Yarlagadda, Prasad
Schuetz, Michael
Quantification of cephalomedullary nail fit in the femur using 3D computer modelling: a comparison between 1.0 and 1.5m bow designs
title Quantification of cephalomedullary nail fit in the femur using 3D computer modelling: a comparison between 1.0 and 1.5m bow designs
title_full Quantification of cephalomedullary nail fit in the femur using 3D computer modelling: a comparison between 1.0 and 1.5m bow designs
title_fullStr Quantification of cephalomedullary nail fit in the femur using 3D computer modelling: a comparison between 1.0 and 1.5m bow designs
title_full_unstemmed Quantification of cephalomedullary nail fit in the femur using 3D computer modelling: a comparison between 1.0 and 1.5m bow designs
title_short Quantification of cephalomedullary nail fit in the femur using 3D computer modelling: a comparison between 1.0 and 1.5m bow designs
title_sort quantification of cephalomedullary nail fit in the femur using 3d computer modelling: a comparison between 1.0 and 1.5m bow designs
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4848874/
https://www.ncbi.nlm.nih.gov/pubmed/27122018
http://dx.doi.org/10.1186/s13018-016-0389-7
work_keys_str_mv AT schmutzbeat quantificationofcephalomedullarynailfitinthefemurusing3dcomputermodellingacomparisonbetween10and15mbowdesigns
AT amarathungajayani quantificationofcephalomedullarynailfitinthefemurusing3dcomputermodellingacomparisonbetween10and15mbowdesigns
AT kmiecstanley quantificationofcephalomedullarynailfitinthefemurusing3dcomputermodellingacomparisonbetween10and15mbowdesigns
AT yarlagaddaprasad quantificationofcephalomedullarynailfitinthefemurusing3dcomputermodellingacomparisonbetween10and15mbowdesigns
AT schuetzmichael quantificationofcephalomedullarynailfitinthefemurusing3dcomputermodellingacomparisonbetween10and15mbowdesigns