Cargando…
Standard Reference Materials (SRMs) for the Calibration and Validation of Analytical Methods for PCBs (as Aroclor Mixtures)
Six Standard Reference Materials (SRMs®) have been prepared by the National Institute of Standards and Technology (NIST) for the determination of PCBs as different Aroclor mixtures in methanol. Six additional SRMs of the same Aroclors in transformer oil have also been prepared. Specifically, solutio...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
2004
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4853117/ https://www.ncbi.nlm.nih.gov/pubmed/27366608 http://dx.doi.org/10.6028/jres.109.016 |
_version_ | 1782430039516643328 |
---|---|
author | Poster, Dianne L. Schantz, Michele M. Leigh, Stefan D. Wise, Stephen A. |
author_facet | Poster, Dianne L. Schantz, Michele M. Leigh, Stefan D. Wise, Stephen A. |
author_sort | Poster, Dianne L. |
collection | PubMed |
description | Six Standard Reference Materials (SRMs®) have been prepared by the National Institute of Standards and Technology (NIST) for the determination of PCBs as different Aroclor mixtures in methanol. Six additional SRMs of the same Aroclors in transformer oil have also been prepared. Specifically, solutions of Aroclors 1016, 1232, 1242, 1254, and 1260 have been gravimetrically prepared (individually) in methanol and transformer oil, mixed, and transferred to amber glass ampoules in approximately 1.2 mL aliquots. Gas chromatography with electron capture detection (GC-ECD) has been used to verify the gravimetric data for each solution and transformer oil SRM. Liquid chromatography was used for the isolation of the Aroclors from the transformer oil SRMs prior to GC-ECD analysis. Separate calibration solutions and oils were prepared with Aroclor levels similar to those in each methanol solution and transformer oil SRM and were processed alongside the samples. The GC-ECD response of each Aroclor was monitored relative to internal standards that were added to the complex mixtures for quantification. The gravimetric concentrations of Aroclors 1242 and 1254 in methanol were also examined by the same method of analysis (GC-ECD) using several different sources of Aroclors and two different capillary GC columns: a 5 % phenyl methylpolysiloxane phase and a relatively non-polar phase. The preparation of the materials, the gas chromatographic results, and the certified concentration values for each Aroclor SRM are described in this paper. |
format | Online Article Text |
id | pubmed-4853117 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2004 |
publisher | [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology |
record_format | MEDLINE/PubMed |
spelling | pubmed-48531172016-06-30 Standard Reference Materials (SRMs) for the Calibration and Validation of Analytical Methods for PCBs (as Aroclor Mixtures) Poster, Dianne L. Schantz, Michele M. Leigh, Stefan D. Wise, Stephen A. J Res Natl Inst Stand Technol Article Six Standard Reference Materials (SRMs®) have been prepared by the National Institute of Standards and Technology (NIST) for the determination of PCBs as different Aroclor mixtures in methanol. Six additional SRMs of the same Aroclors in transformer oil have also been prepared. Specifically, solutions of Aroclors 1016, 1232, 1242, 1254, and 1260 have been gravimetrically prepared (individually) in methanol and transformer oil, mixed, and transferred to amber glass ampoules in approximately 1.2 mL aliquots. Gas chromatography with electron capture detection (GC-ECD) has been used to verify the gravimetric data for each solution and transformer oil SRM. Liquid chromatography was used for the isolation of the Aroclors from the transformer oil SRMs prior to GC-ECD analysis. Separate calibration solutions and oils were prepared with Aroclor levels similar to those in each methanol solution and transformer oil SRM and were processed alongside the samples. The GC-ECD response of each Aroclor was monitored relative to internal standards that were added to the complex mixtures for quantification. The gravimetric concentrations of Aroclors 1242 and 1254 in methanol were also examined by the same method of analysis (GC-ECD) using several different sources of Aroclors and two different capillary GC columns: a 5 % phenyl methylpolysiloxane phase and a relatively non-polar phase. The preparation of the materials, the gas chromatographic results, and the certified concentration values for each Aroclor SRM are described in this paper. [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2004 2004-04-01 /pmc/articles/PMC4853117/ /pubmed/27366608 http://dx.doi.org/10.6028/jres.109.016 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Institute of Standards and Technology is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright. |
spellingShingle | Article Poster, Dianne L. Schantz, Michele M. Leigh, Stefan D. Wise, Stephen A. Standard Reference Materials (SRMs) for the Calibration and Validation of Analytical Methods for PCBs (as Aroclor Mixtures) |
title | Standard Reference Materials (SRMs) for the Calibration and Validation of Analytical Methods for PCBs (as Aroclor Mixtures) |
title_full | Standard Reference Materials (SRMs) for the Calibration and Validation of Analytical Methods for PCBs (as Aroclor Mixtures) |
title_fullStr | Standard Reference Materials (SRMs) for the Calibration and Validation of Analytical Methods for PCBs (as Aroclor Mixtures) |
title_full_unstemmed | Standard Reference Materials (SRMs) for the Calibration and Validation of Analytical Methods for PCBs (as Aroclor Mixtures) |
title_short | Standard Reference Materials (SRMs) for the Calibration and Validation of Analytical Methods for PCBs (as Aroclor Mixtures) |
title_sort | standard reference materials (srms) for the calibration and validation of analytical methods for pcbs (as aroclor mixtures) |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4853117/ https://www.ncbi.nlm.nih.gov/pubmed/27366608 http://dx.doi.org/10.6028/jres.109.016 |
work_keys_str_mv | AT posterdiannel standardreferencematerialssrmsforthecalibrationandvalidationofanalyticalmethodsforpcbsasaroclormixtures AT schantzmichelem standardreferencematerialssrmsforthecalibrationandvalidationofanalyticalmethodsforpcbsasaroclormixtures AT leighstefand standardreferencematerialssrmsforthecalibrationandvalidationofanalyticalmethodsforpcbsasaroclormixtures AT wisestephena standardreferencematerialssrmsforthecalibrationandvalidationofanalyticalmethodsforpcbsasaroclormixtures |