Simulated-use validation of a sponge ATP method for determining the adequacy of manual cleaning of endoscope channels

BACKGROUND: The objective of this study was to validate the relative light unit (RLU) cut-off of adequate cleaning of flexible colonoscopes for an ATP (adenosine tri-phosphate) test kit that used a sponge channel collection method. METHODS: This was a simulated-use study. The instrument channel segm...