Reliability of the i‐STAT for the determination of blood electrolyte (K(+), Na(+), and CI (−)) concentrations in cattle

BACKGROUND: Rapid determination of blood electrolyte concentrations can help determine electrolyte status and delivery of effective volume of electrolyte solutions in field conditions. OBJECTIVE: To evaluate reliability of the i‐STAT, a point‐of‐care (POC) device, in measuring blood K(+), Na(+), and...

Descripción completa

Detalles Bibliográficos
Autores principales: Yildirim, E., Karapinar, T., Hayirli, A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4858094/
https://www.ncbi.nlm.nih.gov/pubmed/25619525
http://dx.doi.org/10.1111/jvim.12526
_version_ 1782430750508843008
author Yildirim, E.
Karapinar, T.
Hayirli, A.
author_facet Yildirim, E.
Karapinar, T.
Hayirli, A.
author_sort Yildirim, E.
collection PubMed
description BACKGROUND: Rapid determination of blood electrolyte concentrations can help determine electrolyte status and delivery of effective volume of electrolyte solutions in field conditions. OBJECTIVE: To evaluate reliability of the i‐STAT, a point‐of‐care (POC) device, in measuring blood K(+), Na(+), and CI (−) concentrations in cattle. ANIMALS: Ninety‐eight cattle with various diseases. METHODS: In this prospective study, blood samples collected from the jugular vein were processed for determination of K(+), Na(+), and CI (−) concentrations in blood and plasma using the i‐STAT and auto‐analyzer (Cobas C501), respectively. Blood and plasma electrolyte data were subjected to student t‐test for comparison, the concordance analysis for agreement, accuracy, and precision, the Passing‐Bablok regression and the Bland‐Altman plot for reliability, and receiver operating characteristics curves for sensitivity (Se) and specificity (Sp). RESULTS: Plasma concentrations of K(+) (4.39 versus 4.2 mmol/L; P < .0001) and CI (−) (100.30 versus 99.4 mmol/L; P < .04) were greater than their concentrations in blood. Plasma and blood Na(+) concentrations were similar (136.95 versus 136.8 mmol/L). The i‐STAT results were highly correlated with the Cobas C501 results (r = 0.970, 0.922, and 0.866 for K(+), Na(+), and CI (−), respectively). Regression equations fitting blood (Y) and plasma (X) concentration did not deviate from the identity line for K(+) (Y = −0.10 + 0.98 × X), Na(+) (Y = X), and CI (−) (Y = 3.04 + 0.96 × X). The mean bias (blood concentration ‐ plasma concentration) was −0.20 for K(+) (P = .03), −0.16 for Na(+) (P = .12), and −0.87 for CI (−) (P = .93). The i‐STAT had 76–100% Se and 87.7–100% Sp for assessing electrolyte statuses. CONCLUSIONS AND CLINICAL IMPORTANCE: The i‐STAT yielded results that were in agreement with the auto‐analyzer, with negligible biases in measurement of plasma K(+), Na(+), and CI (−) concentrations. The i‐STAT is a reliable POC device and can be used in field condition to assess electrolyte status in cattle.
format Online
Article
Text
id pubmed-4858094
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-48580942016-06-22 Reliability of the i‐STAT for the determination of blood electrolyte (K(+), Na(+), and CI (−)) concentrations in cattle Yildirim, E. Karapinar, T. Hayirli, A. J Vet Intern Med Standard Articles BACKGROUND: Rapid determination of blood electrolyte concentrations can help determine electrolyte status and delivery of effective volume of electrolyte solutions in field conditions. OBJECTIVE: To evaluate reliability of the i‐STAT, a point‐of‐care (POC) device, in measuring blood K(+), Na(+), and CI (−) concentrations in cattle. ANIMALS: Ninety‐eight cattle with various diseases. METHODS: In this prospective study, blood samples collected from the jugular vein were processed for determination of K(+), Na(+), and CI (−) concentrations in blood and plasma using the i‐STAT and auto‐analyzer (Cobas C501), respectively. Blood and plasma electrolyte data were subjected to student t‐test for comparison, the concordance analysis for agreement, accuracy, and precision, the Passing‐Bablok regression and the Bland‐Altman plot for reliability, and receiver operating characteristics curves for sensitivity (Se) and specificity (Sp). RESULTS: Plasma concentrations of K(+) (4.39 versus 4.2 mmol/L; P < .0001) and CI (−) (100.30 versus 99.4 mmol/L; P < .04) were greater than their concentrations in blood. Plasma and blood Na(+) concentrations were similar (136.95 versus 136.8 mmol/L). The i‐STAT results were highly correlated with the Cobas C501 results (r = 0.970, 0.922, and 0.866 for K(+), Na(+), and CI (−), respectively). Regression equations fitting blood (Y) and plasma (X) concentration did not deviate from the identity line for K(+) (Y = −0.10 + 0.98 × X), Na(+) (Y = X), and CI (−) (Y = 3.04 + 0.96 × X). The mean bias (blood concentration ‐ plasma concentration) was −0.20 for K(+) (P = .03), −0.16 for Na(+) (P = .12), and −0.87 for CI (−) (P = .93). The i‐STAT had 76–100% Se and 87.7–100% Sp for assessing electrolyte statuses. CONCLUSIONS AND CLINICAL IMPORTANCE: The i‐STAT yielded results that were in agreement with the auto‐analyzer, with negligible biases in measurement of plasma K(+), Na(+), and CI (−) concentrations. The i‐STAT is a reliable POC device and can be used in field condition to assess electrolyte status in cattle. John Wiley and Sons Inc. 2015-01-25 2015 /pmc/articles/PMC4858094/ /pubmed/25619525 http://dx.doi.org/10.1111/jvim.12526 Text en Copyright © 2015 by the American College of Veterinary Internal Medicine
spellingShingle Standard Articles
Yildirim, E.
Karapinar, T.
Hayirli, A.
Reliability of the i‐STAT for the determination of blood electrolyte (K(+), Na(+), and CI (−)) concentrations in cattle
title Reliability of the i‐STAT for the determination of blood electrolyte (K(+), Na(+), and CI (−)) concentrations in cattle
title_full Reliability of the i‐STAT for the determination of blood electrolyte (K(+), Na(+), and CI (−)) concentrations in cattle
title_fullStr Reliability of the i‐STAT for the determination of blood electrolyte (K(+), Na(+), and CI (−)) concentrations in cattle
title_full_unstemmed Reliability of the i‐STAT for the determination of blood electrolyte (K(+), Na(+), and CI (−)) concentrations in cattle
title_short Reliability of the i‐STAT for the determination of blood electrolyte (K(+), Na(+), and CI (−)) concentrations in cattle
title_sort reliability of the i‐stat for the determination of blood electrolyte (k(+), na(+), and ci (−)) concentrations in cattle
topic Standard Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4858094/
https://www.ncbi.nlm.nih.gov/pubmed/25619525
http://dx.doi.org/10.1111/jvim.12526
work_keys_str_mv AT yildirime reliabilityoftheistatforthedeterminationofbloodelectrolyteknaandciconcentrationsincattle
AT karapinart reliabilityoftheistatforthedeterminationofbloodelectrolyteknaandciconcentrationsincattle
AT hayirlia reliabilityoftheistatforthedeterminationofbloodelectrolyteknaandciconcentrationsincattle