Cargando…

Diagnostic Accuracy of Ultrasound B scan using 10 MHz linear probe in ocular trauma;results from a high burden country

OBJECTIVE: To study the diagnostic accuracy of Ultrasound B scan using 10 MHz linear probe in ocular trauma. METHODS: A total of 61 patients with 63 ocular injuries were assessed during July 2013 to January 2014. All patients were referred to the department of Radiology from Emergency Room since ade...

Descripción completa

Detalles Bibliográficos
Autores principales: Shazlee, Muhammad Kashif, Ali, Muhammad, SaadAhmed, Muhammad, Hussain, Ammad, Hameed, Kamran, Lutfi, Irfan Amjad, Khan, Muhammad Tahir
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Professional Medical Publications 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4859028/
https://www.ncbi.nlm.nih.gov/pubmed/27182245
http://dx.doi.org/10.12669/pjms.322.9241
Descripción
Sumario:OBJECTIVE: To study the diagnostic accuracy of Ultrasound B scan using 10 MHz linear probe in ocular trauma. METHODS: A total of 61 patients with 63 ocular injuries were assessed during July 2013 to January 2014. All patients were referred to the department of Radiology from Emergency Room since adequate clinical assessment of the fundus was impossible because of the presence of opaque ocular media. Based on radiological diagnosis, the patients were provided treatment (surgical or medical). Clinical diagnosis was confirmed during surgical procedures or clinical follow-up. RESULTS: A total of 63 ocular injuries were examined in 61 patients. The overall sensitivity was 91.5%, Specificity was 98.87%, Positive predictive value was 87.62 and Negative predictive value was 99%. CONCLUSION: Ultrasound B-scan is a sensitive, non invasive and rapid way of assessing intraocular damage caused by blunt or penetrating eye injuries.