How Accurate is Your Sclerostin Measurement? Comparison Between Three Commercially Available Sclerostin ELISA Kits

Sclerostin, bone formation antagonist is in the spotlight as a potential biomarker for diseases presenting with associated bone disorders such as chronic kidney disease (CDK-MBD). Accurate measurement of sclerostin is therefore important. Several immunoassays are available to measure sclerostin in s...

Descripción completa

Detalles Bibliográficos
Autores principales: Piec, Isabelle, Washbourne, Christopher, Tang, Jonathan, Fisher, Emily, Greeves, Julie, Jackson, Sarah, Fraser, William D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer US 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4860200/
https://www.ncbi.nlm.nih.gov/pubmed/26749312
http://dx.doi.org/10.1007/s00223-015-0105-3
_version_ 1782431029650259968
author Piec, Isabelle
Washbourne, Christopher
Tang, Jonathan
Fisher, Emily
Greeves, Julie
Jackson, Sarah
Fraser, William D.
author_facet Piec, Isabelle
Washbourne, Christopher
Tang, Jonathan
Fisher, Emily
Greeves, Julie
Jackson, Sarah
Fraser, William D.
author_sort Piec, Isabelle
collection PubMed
description Sclerostin, bone formation antagonist is in the spotlight as a potential biomarker for diseases presenting with associated bone disorders such as chronic kidney disease (CDK-MBD). Accurate measurement of sclerostin is therefore important. Several immunoassays are available to measure sclerostin in serum and plasma. We compared the performance of three commercial ELISA kits. We measured sclerostin concentrations in serum and EDTA plasma obtained from healthy young (18–26 years) human subjects using kits from Biomedica, TECOmedical and from R&D Systems. The circulating sclerostin concentrations were systematically higher when measured with the Biomedica assay (serum: 35.5 ± 1.1 pmol/L; EDTA: 39.4 ± 2.0 pmol/L; mean ± SD) as compared with TECOmedical (serum: 21.8 ± 0.7 pmol/L; EDTA: 27.2 ± 1.3 pmol/L) and R&D Systems (serum: 7.6 ± 0.3 pmol/L; EDTA: 30.9 ± 1.5 pmol/L). We found a good correlation between the assay for EDTA plasma (r > 0.6; p < 0.001) while in serum, only measurements obtained using TECOmedical and R&D Systems assays correlated significantly (r = 0.78; p < 0.001). There was no correlation between matrices results when using the Biomedica kit (r = 0.20). The variability in values generated from Biomedica, R&D Systems and TECOmedical assays raises questions regarding the accuracy and specificity of the assays. Direct comparison of studies using different kits is not possible and great care should be given to measurement of sclerostin, with traceability of reagents. Standardization with appropriate material is required before different sclerostin assays can be introduced in clinical practice.
format Online
Article
Text
id pubmed-4860200
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Springer US
record_format MEDLINE/PubMed
spelling pubmed-48602002016-05-21 How Accurate is Your Sclerostin Measurement? Comparison Between Three Commercially Available Sclerostin ELISA Kits Piec, Isabelle Washbourne, Christopher Tang, Jonathan Fisher, Emily Greeves, Julie Jackson, Sarah Fraser, William D. Calcif Tissue Int Original Research Sclerostin, bone formation antagonist is in the spotlight as a potential biomarker for diseases presenting with associated bone disorders such as chronic kidney disease (CDK-MBD). Accurate measurement of sclerostin is therefore important. Several immunoassays are available to measure sclerostin in serum and plasma. We compared the performance of three commercial ELISA kits. We measured sclerostin concentrations in serum and EDTA plasma obtained from healthy young (18–26 years) human subjects using kits from Biomedica, TECOmedical and from R&D Systems. The circulating sclerostin concentrations were systematically higher when measured with the Biomedica assay (serum: 35.5 ± 1.1 pmol/L; EDTA: 39.4 ± 2.0 pmol/L; mean ± SD) as compared with TECOmedical (serum: 21.8 ± 0.7 pmol/L; EDTA: 27.2 ± 1.3 pmol/L) and R&D Systems (serum: 7.6 ± 0.3 pmol/L; EDTA: 30.9 ± 1.5 pmol/L). We found a good correlation between the assay for EDTA plasma (r > 0.6; p < 0.001) while in serum, only measurements obtained using TECOmedical and R&D Systems assays correlated significantly (r = 0.78; p < 0.001). There was no correlation between matrices results when using the Biomedica kit (r = 0.20). The variability in values generated from Biomedica, R&D Systems and TECOmedical assays raises questions regarding the accuracy and specificity of the assays. Direct comparison of studies using different kits is not possible and great care should be given to measurement of sclerostin, with traceability of reagents. Standardization with appropriate material is required before different sclerostin assays can be introduced in clinical practice. Springer US 2016-01-09 2016 /pmc/articles/PMC4860200/ /pubmed/26749312 http://dx.doi.org/10.1007/s00223-015-0105-3 Text en © The Author(s) 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Research
Piec, Isabelle
Washbourne, Christopher
Tang, Jonathan
Fisher, Emily
Greeves, Julie
Jackson, Sarah
Fraser, William D.
How Accurate is Your Sclerostin Measurement? Comparison Between Three Commercially Available Sclerostin ELISA Kits
title How Accurate is Your Sclerostin Measurement? Comparison Between Three Commercially Available Sclerostin ELISA Kits
title_full How Accurate is Your Sclerostin Measurement? Comparison Between Three Commercially Available Sclerostin ELISA Kits
title_fullStr How Accurate is Your Sclerostin Measurement? Comparison Between Three Commercially Available Sclerostin ELISA Kits
title_full_unstemmed How Accurate is Your Sclerostin Measurement? Comparison Between Three Commercially Available Sclerostin ELISA Kits
title_short How Accurate is Your Sclerostin Measurement? Comparison Between Three Commercially Available Sclerostin ELISA Kits
title_sort how accurate is your sclerostin measurement? comparison between three commercially available sclerostin elisa kits
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4860200/
https://www.ncbi.nlm.nih.gov/pubmed/26749312
http://dx.doi.org/10.1007/s00223-015-0105-3
work_keys_str_mv AT piecisabelle howaccurateisyoursclerostinmeasurementcomparisonbetweenthreecommerciallyavailablesclerostinelisakits
AT washbournechristopher howaccurateisyoursclerostinmeasurementcomparisonbetweenthreecommerciallyavailablesclerostinelisakits
AT tangjonathan howaccurateisyoursclerostinmeasurementcomparisonbetweenthreecommerciallyavailablesclerostinelisakits
AT fisheremily howaccurateisyoursclerostinmeasurementcomparisonbetweenthreecommerciallyavailablesclerostinelisakits
AT greevesjulie howaccurateisyoursclerostinmeasurementcomparisonbetweenthreecommerciallyavailablesclerostinelisakits
AT jacksonsarah howaccurateisyoursclerostinmeasurementcomparisonbetweenthreecommerciallyavailablesclerostinelisakits
AT fraserwilliamd howaccurateisyoursclerostinmeasurementcomparisonbetweenthreecommerciallyavailablesclerostinelisakits