Design, development and validation of a new laryngo-pharyngeal endoscopic esthesiometer and range-finder based on the assessment of air-pulse variability determinants
BACKGROUND: Laryngo-pharyngeal mechano-sensitivity (LPMS) is involved in dysphagia, sleep apnea, stroke, irritable larynx syndrome and cough hypersensitivity syndrome among other disorders. These conditions are associated with a wide range of airway reflex abnormalities. However, the current device...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4862145/ https://www.ncbi.nlm.nih.gov/pubmed/27160751 http://dx.doi.org/10.1186/s12938-016-0166-1 |
_version_ | 1782431315264536576 |
---|---|
author | Giraldo-Cadavid, Luis F. Agudelo-Otalora, Luis Mauricio Burguete, Javier Arbulu, Mario Moscoso, William Daniel Martínez, Fabio Ortiz, Andrés Felipe Diaz, Juan Pantoja, Jaime A. Rueda-Arango, Andrés Felipe Fernández, Secundino |
author_facet | Giraldo-Cadavid, Luis F. Agudelo-Otalora, Luis Mauricio Burguete, Javier Arbulu, Mario Moscoso, William Daniel Martínez, Fabio Ortiz, Andrés Felipe Diaz, Juan Pantoja, Jaime A. Rueda-Arango, Andrés Felipe Fernández, Secundino |
author_sort | Giraldo-Cadavid, Luis F. |
collection | PubMed |
description | BACKGROUND: Laryngo-pharyngeal mechano-sensitivity (LPMS) is involved in dysphagia, sleep apnea, stroke, irritable larynx syndrome and cough hypersensitivity syndrome among other disorders. These conditions are associated with a wide range of airway reflex abnormalities. However, the current device for exploring LPMS is limited because it assesses only the laryngeal adductor reflex during fiber-optic endoscopic evaluations of swallowing and requires a high degree of expertise to obtain reliable results, introducing intrinsic expert variability and subjectivity. METHODS: We designed, developed and validated a new air-pulse laryngo-pharyngeal endoscopic esthesiometer with a built-in laser range-finder (LPEER) based on the evaluation and control of air-pulse variability determinants and on intrinsic observer variability and subjectivity determinants of the distance, angle and site of stimulus impact. The LPEER was designed to be capable of delivering precise and accurate stimuli with a wide range of intensities that can explore most laryngo-pharyngeal reflexes. RESULTS: We initially explored the potential factors affecting the reliability of LPMS tests and included these factors in a multiple linear regression model. The following factors significantly affected the precision and accuracy of the test (P < 0.001): the tube conducting the air-pulses, the supply pressure of the system, the duration of the air-pulses, and the distance and angle between the end of the tube conducting the air-pulses and the site of impact. To control all of these factors, an LPEER consisting of an air-pulse generator and an endoscopic laser range-finder was designed and manufactured. We assessed the precision and accuracy of the LPEER’s stimulus and range-finder according to the coefficient of variation (CV) and by looking at the differences between the measured properties and the desired values, and we performed a pilot validation on ten human subjects. The air-pulses and range-finder exhibited good precision and accuracy (CV < 0.06), with differences between the desired and measured properties at <3 % and a range-finder measurement error of <1 mm. The tests in patients demonstrated obtainable and reproducible thresholds for the laryngeal adductor, cough and gag reflexes. CONCLUSIONS: The new LPEER was capable of delivering precise and accurate stimuli for exploring laryngo-pharyngeal reflexes. |
format | Online Article Text |
id | pubmed-4862145 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-48621452016-05-11 Design, development and validation of a new laryngo-pharyngeal endoscopic esthesiometer and range-finder based on the assessment of air-pulse variability determinants Giraldo-Cadavid, Luis F. Agudelo-Otalora, Luis Mauricio Burguete, Javier Arbulu, Mario Moscoso, William Daniel Martínez, Fabio Ortiz, Andrés Felipe Diaz, Juan Pantoja, Jaime A. Rueda-Arango, Andrés Felipe Fernández, Secundino Biomed Eng Online Research BACKGROUND: Laryngo-pharyngeal mechano-sensitivity (LPMS) is involved in dysphagia, sleep apnea, stroke, irritable larynx syndrome and cough hypersensitivity syndrome among other disorders. These conditions are associated with a wide range of airway reflex abnormalities. However, the current device for exploring LPMS is limited because it assesses only the laryngeal adductor reflex during fiber-optic endoscopic evaluations of swallowing and requires a high degree of expertise to obtain reliable results, introducing intrinsic expert variability and subjectivity. METHODS: We designed, developed and validated a new air-pulse laryngo-pharyngeal endoscopic esthesiometer with a built-in laser range-finder (LPEER) based on the evaluation and control of air-pulse variability determinants and on intrinsic observer variability and subjectivity determinants of the distance, angle and site of stimulus impact. The LPEER was designed to be capable of delivering precise and accurate stimuli with a wide range of intensities that can explore most laryngo-pharyngeal reflexes. RESULTS: We initially explored the potential factors affecting the reliability of LPMS tests and included these factors in a multiple linear regression model. The following factors significantly affected the precision and accuracy of the test (P < 0.001): the tube conducting the air-pulses, the supply pressure of the system, the duration of the air-pulses, and the distance and angle between the end of the tube conducting the air-pulses and the site of impact. To control all of these factors, an LPEER consisting of an air-pulse generator and an endoscopic laser range-finder was designed and manufactured. We assessed the precision and accuracy of the LPEER’s stimulus and range-finder according to the coefficient of variation (CV) and by looking at the differences between the measured properties and the desired values, and we performed a pilot validation on ten human subjects. The air-pulses and range-finder exhibited good precision and accuracy (CV < 0.06), with differences between the desired and measured properties at <3 % and a range-finder measurement error of <1 mm. The tests in patients demonstrated obtainable and reproducible thresholds for the laryngeal adductor, cough and gag reflexes. CONCLUSIONS: The new LPEER was capable of delivering precise and accurate stimuli for exploring laryngo-pharyngeal reflexes. BioMed Central 2016-05-10 /pmc/articles/PMC4862145/ /pubmed/27160751 http://dx.doi.org/10.1186/s12938-016-0166-1 Text en © Giraldo-Cadavid et al. 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Giraldo-Cadavid, Luis F. Agudelo-Otalora, Luis Mauricio Burguete, Javier Arbulu, Mario Moscoso, William Daniel Martínez, Fabio Ortiz, Andrés Felipe Diaz, Juan Pantoja, Jaime A. Rueda-Arango, Andrés Felipe Fernández, Secundino Design, development and validation of a new laryngo-pharyngeal endoscopic esthesiometer and range-finder based on the assessment of air-pulse variability determinants |
title | Design, development and validation of a new laryngo-pharyngeal endoscopic esthesiometer and range-finder based on the assessment of air-pulse variability determinants |
title_full | Design, development and validation of a new laryngo-pharyngeal endoscopic esthesiometer and range-finder based on the assessment of air-pulse variability determinants |
title_fullStr | Design, development and validation of a new laryngo-pharyngeal endoscopic esthesiometer and range-finder based on the assessment of air-pulse variability determinants |
title_full_unstemmed | Design, development and validation of a new laryngo-pharyngeal endoscopic esthesiometer and range-finder based on the assessment of air-pulse variability determinants |
title_short | Design, development and validation of a new laryngo-pharyngeal endoscopic esthesiometer and range-finder based on the assessment of air-pulse variability determinants |
title_sort | design, development and validation of a new laryngo-pharyngeal endoscopic esthesiometer and range-finder based on the assessment of air-pulse variability determinants |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4862145/ https://www.ncbi.nlm.nih.gov/pubmed/27160751 http://dx.doi.org/10.1186/s12938-016-0166-1 |
work_keys_str_mv | AT giraldocadavidluisf designdevelopmentandvalidationofanewlaryngopharyngealendoscopicesthesiometerandrangefinderbasedontheassessmentofairpulsevariabilitydeterminants AT agudelootaloraluismauricio designdevelopmentandvalidationofanewlaryngopharyngealendoscopicesthesiometerandrangefinderbasedontheassessmentofairpulsevariabilitydeterminants AT burguetejavier designdevelopmentandvalidationofanewlaryngopharyngealendoscopicesthesiometerandrangefinderbasedontheassessmentofairpulsevariabilitydeterminants AT arbulumario designdevelopmentandvalidationofanewlaryngopharyngealendoscopicesthesiometerandrangefinderbasedontheassessmentofairpulsevariabilitydeterminants AT moscosowilliamdaniel designdevelopmentandvalidationofanewlaryngopharyngealendoscopicesthesiometerandrangefinderbasedontheassessmentofairpulsevariabilitydeterminants AT martinezfabio designdevelopmentandvalidationofanewlaryngopharyngealendoscopicesthesiometerandrangefinderbasedontheassessmentofairpulsevariabilitydeterminants AT ortizandresfelipe designdevelopmentandvalidationofanewlaryngopharyngealendoscopicesthesiometerandrangefinderbasedontheassessmentofairpulsevariabilitydeterminants AT diazjuan designdevelopmentandvalidationofanewlaryngopharyngealendoscopicesthesiometerandrangefinderbasedontheassessmentofairpulsevariabilitydeterminants AT pantojajaimea designdevelopmentandvalidationofanewlaryngopharyngealendoscopicesthesiometerandrangefinderbasedontheassessmentofairpulsevariabilitydeterminants AT ruedaarangoandresfelipe designdevelopmentandvalidationofanewlaryngopharyngealendoscopicesthesiometerandrangefinderbasedontheassessmentofairpulsevariabilitydeterminants AT fernandezsecundino designdevelopmentandvalidationofanewlaryngopharyngealendoscopicesthesiometerandrangefinderbasedontheassessmentofairpulsevariabilitydeterminants |