Cargando…

Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody

Facilities and techniques to characterize heat flux sensors are under development at the National Institute of Standards and Technology. As a part of this effort, a large aperture high-temperature blackbody was commissioned recently. The graphite tube blackbody, heated electrically, has a cavity dia...

Descripción completa

Detalles Bibliográficos
Autores principales: Murthy, A. V., Tsai, B. K., Saunders, R. D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2001
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4862819/
https://www.ncbi.nlm.nih.gov/pubmed/27500049
http://dx.doi.org/10.6028/jres.106.039
_version_ 1782431399159005184
author Murthy, A. V.
Tsai, B. K.
Saunders, R. D.
author_facet Murthy, A. V.
Tsai, B. K.
Saunders, R. D.
author_sort Murthy, A. V.
collection PubMed
description Facilities and techniques to characterize heat flux sensors are under development at the National Institute of Standards and Technology. As a part of this effort, a large aperture high-temperature blackbody was commissioned recently. The graphite tube blackbody, heated electrically, has a cavity diameter of 51 mm and can operate up to a maximum temperature of 2773 K. A closed-loop cooling system using a water-to-water heat exchanger cools electrodes and the outer reflecting shield. This paper describes the newly developed blackbody facility and the validation tests conducted using a reference standard Schmidt-Boelter heat flux sensor. The transfer calibration results obtained on the Schmidt-Boelter sensor agreed with the previous data within the experimental uncertainty limits.
format Online
Article
Text
id pubmed-4862819
institution National Center for Biotechnology Information
language English
publishDate 2001
publisher [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
record_format MEDLINE/PubMed
spelling pubmed-48628192016-08-05 Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody Murthy, A. V. Tsai, B. K. Saunders, R. D. J Res Natl Inst Stand Technol Article Facilities and techniques to characterize heat flux sensors are under development at the National Institute of Standards and Technology. As a part of this effort, a large aperture high-temperature blackbody was commissioned recently. The graphite tube blackbody, heated electrically, has a cavity diameter of 51 mm and can operate up to a maximum temperature of 2773 K. A closed-loop cooling system using a water-to-water heat exchanger cools electrodes and the outer reflecting shield. This paper describes the newly developed blackbody facility and the validation tests conducted using a reference standard Schmidt-Boelter heat flux sensor. The transfer calibration results obtained on the Schmidt-Boelter sensor agreed with the previous data within the experimental uncertainty limits. [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2001 2001-10-01 /pmc/articles/PMC4862819/ /pubmed/27500049 http://dx.doi.org/10.6028/jres.106.039 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Institute of Standards and Technology is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright.
spellingShingle Article
Murthy, A. V.
Tsai, B. K.
Saunders, R. D.
Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody
title Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody
title_full Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody
title_fullStr Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody
title_full_unstemmed Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody
title_short Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody
title_sort transfer calibration validation tests on a heat flux sensor in the 51 mm high-temperature blackbody
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4862819/
https://www.ncbi.nlm.nih.gov/pubmed/27500049
http://dx.doi.org/10.6028/jres.106.039
work_keys_str_mv AT murthyav transfercalibrationvalidationtestsonaheatfluxsensorinthe51mmhightemperatureblackbody
AT tsaibk transfercalibrationvalidationtestsonaheatfluxsensorinthe51mmhightemperatureblackbody
AT saundersrd transfercalibrationvalidationtestsonaheatfluxsensorinthe51mmhightemperatureblackbody