Cargando…
Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody
Facilities and techniques to characterize heat flux sensors are under development at the National Institute of Standards and Technology. As a part of this effort, a large aperture high-temperature blackbody was commissioned recently. The graphite tube blackbody, heated electrically, has a cavity dia...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
2001
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4862819/ https://www.ncbi.nlm.nih.gov/pubmed/27500049 http://dx.doi.org/10.6028/jres.106.039 |
_version_ | 1782431399159005184 |
---|---|
author | Murthy, A. V. Tsai, B. K. Saunders, R. D. |
author_facet | Murthy, A. V. Tsai, B. K. Saunders, R. D. |
author_sort | Murthy, A. V. |
collection | PubMed |
description | Facilities and techniques to characterize heat flux sensors are under development at the National Institute of Standards and Technology. As a part of this effort, a large aperture high-temperature blackbody was commissioned recently. The graphite tube blackbody, heated electrically, has a cavity diameter of 51 mm and can operate up to a maximum temperature of 2773 K. A closed-loop cooling system using a water-to-water heat exchanger cools electrodes and the outer reflecting shield. This paper describes the newly developed blackbody facility and the validation tests conducted using a reference standard Schmidt-Boelter heat flux sensor. The transfer calibration results obtained on the Schmidt-Boelter sensor agreed with the previous data within the experimental uncertainty limits. |
format | Online Article Text |
id | pubmed-4862819 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2001 |
publisher | [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology |
record_format | MEDLINE/PubMed |
spelling | pubmed-48628192016-08-05 Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody Murthy, A. V. Tsai, B. K. Saunders, R. D. J Res Natl Inst Stand Technol Article Facilities and techniques to characterize heat flux sensors are under development at the National Institute of Standards and Technology. As a part of this effort, a large aperture high-temperature blackbody was commissioned recently. The graphite tube blackbody, heated electrically, has a cavity diameter of 51 mm and can operate up to a maximum temperature of 2773 K. A closed-loop cooling system using a water-to-water heat exchanger cools electrodes and the outer reflecting shield. This paper describes the newly developed blackbody facility and the validation tests conducted using a reference standard Schmidt-Boelter heat flux sensor. The transfer calibration results obtained on the Schmidt-Boelter sensor agreed with the previous data within the experimental uncertainty limits. [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2001 2001-10-01 /pmc/articles/PMC4862819/ /pubmed/27500049 http://dx.doi.org/10.6028/jres.106.039 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Institute of Standards and Technology is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright. |
spellingShingle | Article Murthy, A. V. Tsai, B. K. Saunders, R. D. Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody |
title | Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody |
title_full | Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody |
title_fullStr | Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody |
title_full_unstemmed | Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody |
title_short | Transfer Calibration Validation Tests on a Heat Flux Sensor in the 51 mm High-Temperature Blackbody |
title_sort | transfer calibration validation tests on a heat flux sensor in the 51 mm high-temperature blackbody |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4862819/ https://www.ncbi.nlm.nih.gov/pubmed/27500049 http://dx.doi.org/10.6028/jres.106.039 |
work_keys_str_mv | AT murthyav transfercalibrationvalidationtestsonaheatfluxsensorinthe51mmhightemperatureblackbody AT tsaibk transfercalibrationvalidationtestsonaheatfluxsensorinthe51mmhightemperatureblackbody AT saundersrd transfercalibrationvalidationtestsonaheatfluxsensorinthe51mmhightemperatureblackbody |