Sources of Error in UV Radiation Measurements
Increasing commercial, scientific, and technical applications involving ultraviolet (UV) radiation have led to the demand for improved understanding of the performance of instrumentation used to measure this radiation. There has been an effort by manufacturers of UV measuring devices (meters) to pro...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
2001
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4862826/ https://www.ncbi.nlm.nih.gov/pubmed/27500041 http://dx.doi.org/10.6028/jres.106.030 |
_version_ | 1782431400798978048 |
---|---|
author | Larason, Thomas C. Cromer, Christopher L. |
author_facet | Larason, Thomas C. Cromer, Christopher L. |
author_sort | Larason, Thomas C. |
collection | PubMed |
description | Increasing commercial, scientific, and technical applications involving ultraviolet (UV) radiation have led to the demand for improved understanding of the performance of instrumentation used to measure this radiation. There has been an effort by manufacturers of UV measuring devices (meters) to produce simple, optically filtered sensor systems to accomplish the varied measurement needs. We address common sources of measurement errors using these meters. The uncertainty in the calibration of the instrument depends on the response of the UV meter to the spectrum of the sources used and its similarity to the spectrum of the quantity to be measured. In addition, large errors can occur due to out-of-band, non-linear, and non-ideal geometric or spatial response of the UV meters. Finally, in many applications, how well the response of the UV meter approximates the presumed action spectrum needs to be understood for optimal use of the meters. |
format | Online Article Text |
id | pubmed-4862826 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2001 |
publisher | [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology |
record_format | MEDLINE/PubMed |
spelling | pubmed-48628262016-08-05 Sources of Error in UV Radiation Measurements Larason, Thomas C. Cromer, Christopher L. J Res Natl Inst Stand Technol Article Increasing commercial, scientific, and technical applications involving ultraviolet (UV) radiation have led to the demand for improved understanding of the performance of instrumentation used to measure this radiation. There has been an effort by manufacturers of UV measuring devices (meters) to produce simple, optically filtered sensor systems to accomplish the varied measurement needs. We address common sources of measurement errors using these meters. The uncertainty in the calibration of the instrument depends on the response of the UV meter to the spectrum of the sources used and its similarity to the spectrum of the quantity to be measured. In addition, large errors can occur due to out-of-band, non-linear, and non-ideal geometric or spatial response of the UV meters. Finally, in many applications, how well the response of the UV meter approximates the presumed action spectrum needs to be understood for optimal use of the meters. [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2001 2001-08-01 /pmc/articles/PMC4862826/ /pubmed/27500041 http://dx.doi.org/10.6028/jres.106.030 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Institute of Standards and Technology is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright. |
spellingShingle | Article Larason, Thomas C. Cromer, Christopher L. Sources of Error in UV Radiation Measurements |
title | Sources of Error in UV Radiation Measurements |
title_full | Sources of Error in UV Radiation Measurements |
title_fullStr | Sources of Error in UV Radiation Measurements |
title_full_unstemmed | Sources of Error in UV Radiation Measurements |
title_short | Sources of Error in UV Radiation Measurements |
title_sort | sources of error in uv radiation measurements |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4862826/ https://www.ncbi.nlm.nih.gov/pubmed/27500041 http://dx.doi.org/10.6028/jres.106.030 |
work_keys_str_mv | AT larasonthomasc sourcesoferrorinuvradiationmeasurements AT cromerchristopherl sourcesoferrorinuvradiationmeasurements |