X-Ray Microanalysis in the Variable Pressure (Environmental) Scanning Electron Microscope

Electron-excited x-ray microanalysis performed in the variable pressure and environmental scanning electron microscopes is subject to additional artifacts beyond those encountered in the conventional scanning electron microscope. Gas scattering leads to direct contributions to the spectrum from the...

Descripción completa

Detalles Bibliográficos
Autor principal: Newbury, Dale E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 2002
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4863848/
https://www.ncbi.nlm.nih.gov/pubmed/27446754
http://dx.doi.org/10.6028/jres.107.048
Descripción
Sumario:Electron-excited x-ray microanalysis performed in the variable pressure and environmental scanning electron microscopes is subject to additional artifacts beyond those encountered in the conventional scanning electron microscope. Gas scattering leads to direct contributions to the spectrum from the environmental gas, as well as remote generation of x rays by electrons scattered out of the focussed beam. The analyst can exert some degree of control over these artifacts, but depending on the exact situation, spurious elements can appear at the trace (< 0.01 mass fraction), minor (0.01 mass fraction to 0.1 mass fraction), or even major (> 0.1 mass fraction) levels. Dispersed particle samples give the least compromised results, while fine scale microstructures are the most severely compromised. Procedures to optimize the situation based upon specimen preparation as well as spectral processing are described.