Cargando…

Electromagnetic Interference in a Private Swimming Pool: Case report

Although current lead design and filtering capabilities have greatly improved, Electromagnetic Interference (EMI) from environmental sources has been increasingly reported in patients with Cardiac Implantable Electronic Device (CIED) [1]. Few cases of inappropriate intracardiac Cardioverter Defibril...

Descripción completa

Detalles Bibliográficos
Autores principales: Iskandar, Sandia, Lavu, Madhav, Atoui, Moustapha, Lakkireddy, Dhanunjaya
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4867963/
https://www.ncbi.nlm.nih.gov/pubmed/27479205
http://dx.doi.org/10.1016/j.ipej.2016.02.008
Descripción
Sumario:Although current lead design and filtering capabilities have greatly improved, Electromagnetic Interference (EMI) from environmental sources has been increasingly reported in patients with Cardiac Implantable Electronic Device (CIED) [1]. Few cases of inappropriate intracardiac Cardioverter Defibrillator (ICD) associated with swimming pool has been described [2]. Here we present a case of 64 year old male who presented with an interesting EMI signal that was subsequently identified to be related to AC current leak in his swimming pool.