A Comprehensive Evaluation of NIPAM Polymer Gel Dosimeters on Three Orthogonal Planes and Temporal Stability Analysis
Polymer gel dosimeters have been proven useful for dose evaluation in radiotherapy treatments. Previous studies have demonstrated that using a polymer gel dosimeter requires a 24 h reaction time to stabilize and further evaluate the measured dose distribution in two-dimensional dosimetry. In this st...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4871428/ https://www.ncbi.nlm.nih.gov/pubmed/27192217 http://dx.doi.org/10.1371/journal.pone.0155797 |
_version_ | 1782432589101924352 |
---|---|
author | Cheng, Kai-Yuan Hsieh, Ling-Ling Shih, Cheng-Ting |
author_facet | Cheng, Kai-Yuan Hsieh, Ling-Ling Shih, Cheng-Ting |
author_sort | Cheng, Kai-Yuan |
collection | PubMed |
description | Polymer gel dosimeters have been proven useful for dose evaluation in radiotherapy treatments. Previous studies have demonstrated that using a polymer gel dosimeter requires a 24 h reaction time to stabilize and further evaluate the measured dose distribution in two-dimensional dosimetry. In this study, the short-term stability within 24 h and feasibility of N-isopropylacrylamide (NIPAM) polymer gel dosimeters for use in three-dimensional dosimetry were evaluated using magnetic resonance imaging (MRI). NIPAM gels were used to measure the dose volume in a clinical case of intensity-modulated radiation therapy (IMRT). For dose readouts, MR images of irradiated NIPAM gel phantoms were acquired at 2, 5, 12, and 24 h after dose delivery. The mean standard errors of dose conversion from using dose calibration curves (DRC) were calculated. The measured dose volumes at the four time points were compared with those calculated using a treatment planning system (TPS). The mean standard errors of the dose conversion from using the DRCs were lower than 1 Gy. Mean pass rates of 2, 5, 12, and 24 h axial dose maps calculated using gamma evaluation with 3% dose difference and 3 mm distance-to-agreement criteria were 83.5% ± 0.9%, 85.9% ± 0.6%, 98.7% ± 0.3%, and 98.5% ± 0.9%, respectively. Compared with the dose volume histogram of the TPS, the absolute mean relative volume differences of the 2, 5, 12, and 24 h measured dose volumes were lower than 1% for the irradiated region with an absorbed dose higher than 2.8 Gy. It was concluded that a 12 h reaction time was sufficient to acquire accurate dose volume using the NIPAM gels with MR readouts. |
format | Online Article Text |
id | pubmed-4871428 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-48714282016-05-31 A Comprehensive Evaluation of NIPAM Polymer Gel Dosimeters on Three Orthogonal Planes and Temporal Stability Analysis Cheng, Kai-Yuan Hsieh, Ling-Ling Shih, Cheng-Ting PLoS One Research Article Polymer gel dosimeters have been proven useful for dose evaluation in radiotherapy treatments. Previous studies have demonstrated that using a polymer gel dosimeter requires a 24 h reaction time to stabilize and further evaluate the measured dose distribution in two-dimensional dosimetry. In this study, the short-term stability within 24 h and feasibility of N-isopropylacrylamide (NIPAM) polymer gel dosimeters for use in three-dimensional dosimetry were evaluated using magnetic resonance imaging (MRI). NIPAM gels were used to measure the dose volume in a clinical case of intensity-modulated radiation therapy (IMRT). For dose readouts, MR images of irradiated NIPAM gel phantoms were acquired at 2, 5, 12, and 24 h after dose delivery. The mean standard errors of dose conversion from using dose calibration curves (DRC) were calculated. The measured dose volumes at the four time points were compared with those calculated using a treatment planning system (TPS). The mean standard errors of the dose conversion from using the DRCs were lower than 1 Gy. Mean pass rates of 2, 5, 12, and 24 h axial dose maps calculated using gamma evaluation with 3% dose difference and 3 mm distance-to-agreement criteria were 83.5% ± 0.9%, 85.9% ± 0.6%, 98.7% ± 0.3%, and 98.5% ± 0.9%, respectively. Compared with the dose volume histogram of the TPS, the absolute mean relative volume differences of the 2, 5, 12, and 24 h measured dose volumes were lower than 1% for the irradiated region with an absorbed dose higher than 2.8 Gy. It was concluded that a 12 h reaction time was sufficient to acquire accurate dose volume using the NIPAM gels with MR readouts. Public Library of Science 2016-05-18 /pmc/articles/PMC4871428/ /pubmed/27192217 http://dx.doi.org/10.1371/journal.pone.0155797 Text en © 2016 Cheng et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Cheng, Kai-Yuan Hsieh, Ling-Ling Shih, Cheng-Ting A Comprehensive Evaluation of NIPAM Polymer Gel Dosimeters on Three Orthogonal Planes and Temporal Stability Analysis |
title | A Comprehensive Evaluation of NIPAM Polymer Gel Dosimeters on Three Orthogonal Planes and Temporal Stability Analysis |
title_full | A Comprehensive Evaluation of NIPAM Polymer Gel Dosimeters on Three Orthogonal Planes and Temporal Stability Analysis |
title_fullStr | A Comprehensive Evaluation of NIPAM Polymer Gel Dosimeters on Three Orthogonal Planes and Temporal Stability Analysis |
title_full_unstemmed | A Comprehensive Evaluation of NIPAM Polymer Gel Dosimeters on Three Orthogonal Planes and Temporal Stability Analysis |
title_short | A Comprehensive Evaluation of NIPAM Polymer Gel Dosimeters on Three Orthogonal Planes and Temporal Stability Analysis |
title_sort | comprehensive evaluation of nipam polymer gel dosimeters on three orthogonal planes and temporal stability analysis |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4871428/ https://www.ncbi.nlm.nih.gov/pubmed/27192217 http://dx.doi.org/10.1371/journal.pone.0155797 |
work_keys_str_mv | AT chengkaiyuan acomprehensiveevaluationofnipampolymergeldosimetersonthreeorthogonalplanesandtemporalstabilityanalysis AT hsiehlingling acomprehensiveevaluationofnipampolymergeldosimetersonthreeorthogonalplanesandtemporalstabilityanalysis AT shihchengting acomprehensiveevaluationofnipampolymergeldosimetersonthreeorthogonalplanesandtemporalstabilityanalysis AT chengkaiyuan comprehensiveevaluationofnipampolymergeldosimetersonthreeorthogonalplanesandtemporalstabilityanalysis AT hsiehlingling comprehensiveevaluationofnipampolymergeldosimetersonthreeorthogonalplanesandtemporalstabilityanalysis AT shihchengting comprehensiveevaluationofnipampolymergeldosimetersonthreeorthogonalplanesandtemporalstabilityanalysis |