Standardized Loads Acting in Hip Implants
With the increasing success of hip joint replacements, the average age of patients has decreased, patients have become more active and their expectations of the implant durability have risen. Thus, pre-clinical endurance tests on hip implants require defining realistic in vivo loads from younger and...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4873223/ https://www.ncbi.nlm.nih.gov/pubmed/27195789 http://dx.doi.org/10.1371/journal.pone.0155612 |
_version_ | 1782432866231123968 |
---|---|
author | Bergmann, Georg Bender, Alwina Dymke, Jörn Duda, Georg Damm, Philipp |
author_facet | Bergmann, Georg Bender, Alwina Dymke, Jörn Duda, Georg Damm, Philipp |
author_sort | Bergmann, Georg |
collection | PubMed |
description | With the increasing success of hip joint replacements, the average age of patients has decreased, patients have become more active and their expectations of the implant durability have risen. Thus, pre-clinical endurance tests on hip implants require defining realistic in vivo loads from younger and more active patients. These loads require simplifications to be applicable for simulator tests and numerical analyses. Here, the contact forces in the joint were measured with instrumented hip implants in ten subjects during nine of the most physically demanding and frequent activities of daily living. Typical levels and directions of average and high joint loads were extracted from the intra- and inter-individually widely varying individual data. These data can also be used to analyse bone remodelling at the implant-bone interface, evaluate tissue straining in finite element studies or validate analytical loading predictions, among other uses. The current ISO standards for endurance tests of implant stems and necks are based on historic analytical data from the 1970s. Comparisons of these test forces with in vivo loads unveiled that their unidirectional orientations deviate from the time-dependent in vivo directions during walking and most other activities. The ISO force for testing the stem is substantially too low while the ISO force for the neck better matches typical in vivo magnitudes. Because the magnitudes and orientations of peak forces substantially vary among the activities, load scenarios that reflect a collection of time-dependent high forces should be applied rather than using unidirectional forces. Based on data from ten patients, proposals for the most demanding activities, the time courses of the contact forces and the required cycle numbers for testing are given here. Friction moments in the joint were measured in addition to the contact forces. The moment data were also standardized and can be applied to wear tests of the implant. It was shown that friction only very slightly influences the stresses in the implant neck and shaft. |
format | Online Article Text |
id | pubmed-4873223 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-48732232016-06-09 Standardized Loads Acting in Hip Implants Bergmann, Georg Bender, Alwina Dymke, Jörn Duda, Georg Damm, Philipp PLoS One Research Article With the increasing success of hip joint replacements, the average age of patients has decreased, patients have become more active and their expectations of the implant durability have risen. Thus, pre-clinical endurance tests on hip implants require defining realistic in vivo loads from younger and more active patients. These loads require simplifications to be applicable for simulator tests and numerical analyses. Here, the contact forces in the joint were measured with instrumented hip implants in ten subjects during nine of the most physically demanding and frequent activities of daily living. Typical levels and directions of average and high joint loads were extracted from the intra- and inter-individually widely varying individual data. These data can also be used to analyse bone remodelling at the implant-bone interface, evaluate tissue straining in finite element studies or validate analytical loading predictions, among other uses. The current ISO standards for endurance tests of implant stems and necks are based on historic analytical data from the 1970s. Comparisons of these test forces with in vivo loads unveiled that their unidirectional orientations deviate from the time-dependent in vivo directions during walking and most other activities. The ISO force for testing the stem is substantially too low while the ISO force for the neck better matches typical in vivo magnitudes. Because the magnitudes and orientations of peak forces substantially vary among the activities, load scenarios that reflect a collection of time-dependent high forces should be applied rather than using unidirectional forces. Based on data from ten patients, proposals for the most demanding activities, the time courses of the contact forces and the required cycle numbers for testing are given here. Friction moments in the joint were measured in addition to the contact forces. The moment data were also standardized and can be applied to wear tests of the implant. It was shown that friction only very slightly influences the stresses in the implant neck and shaft. Public Library of Science 2016-05-19 /pmc/articles/PMC4873223/ /pubmed/27195789 http://dx.doi.org/10.1371/journal.pone.0155612 Text en © 2016 Bergmann et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Bergmann, Georg Bender, Alwina Dymke, Jörn Duda, Georg Damm, Philipp Standardized Loads Acting in Hip Implants |
title | Standardized Loads Acting in Hip Implants |
title_full | Standardized Loads Acting in Hip Implants |
title_fullStr | Standardized Loads Acting in Hip Implants |
title_full_unstemmed | Standardized Loads Acting in Hip Implants |
title_short | Standardized Loads Acting in Hip Implants |
title_sort | standardized loads acting in hip implants |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4873223/ https://www.ncbi.nlm.nih.gov/pubmed/27195789 http://dx.doi.org/10.1371/journal.pone.0155612 |
work_keys_str_mv | AT bergmanngeorg standardizedloadsactinginhipimplants AT benderalwina standardizedloadsactinginhipimplants AT dymkejorn standardizedloadsactinginhipimplants AT dudageorg standardizedloadsactinginhipimplants AT dammphilipp standardizedloadsactinginhipimplants |