Reliability of a Manual Procedure for Marking the EZ Endpoint Location in Patients with Retinitis Pigmentosa

PURPOSE: We developed and evaluated a training procedure for marking the endpoints of the ellipsoid zone (EZ), also known as the inner segment/outer segment (IS/OS) border, on frequency domain optical coherence tomography (fdOCT) scans from patients with retinitis pigmentosa (RP). METHODS: A manual...

Descripción completa

Detalles Bibliográficos
Autores principales: Ramachandran, Rithambara, Cai, Cindy X., Lee, Dongwon, Epstein, Benjamin C., Locke, Kirsten G., Birch, David G., Hood, Donald C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Association for Research in Vision and Ophthalmology 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4874452/
https://www.ncbi.nlm.nih.gov/pubmed/27226930
http://dx.doi.org/10.1167/tvst.5.3.6
Descripción
Sumario:PURPOSE: We developed and evaluated a training procedure for marking the endpoints of the ellipsoid zone (EZ), also known as the inner segment/outer segment (IS/OS) border, on frequency domain optical coherence tomography (fdOCT) scans from patients with retinitis pigmentosa (RP). METHODS: A manual for marking EZ endpoints was developed and used to train 2 inexperienced graders. After training, an experienced grader and the 2 trained graders marked the endpoints on fdOCT horizontal line scans through the macula from 45 patients with RP. They marked the endpoints on these same scans again 1 month later. RESULTS: Intragrader agreement was excellent. The intraclass correlation coefficient (ICC) was 0.99, the average difference of endpoint locations (19.6 μm) was close to 0 μm, and the 95% limits were between −284 and 323 μm, approximately ±1.1°. Intergrader agreement also was excellent. The ICC values were 0.98 (time 1) and 0.97 (time 2), the average difference among graders was close to zero, and the 95% limits of these differences was less than 350 μm, approximately 1.2°, for both test times. CONCLUSIONS: While automated algorithms are becoming increasingly accurate, EZ endpoints still have to be verified manually and corrected when necessary. With training, the inter- and intragrader agreement of manually marked endpoints is excellent. TRANSLATIONAL RELEVANCE: For clinical studies, the EZ endpoints can be marked by hand if a training procedure, including a manual, is used. The endpoint confidence intervals, well under ±2.0°, are considerably smaller than the 6° spacing for the typically used static visual field.