Cargando…
Thermal-Conductivity Apparatus for Steady-State, Comparative Measurement of Ceramic Coatings
An apparatus has been developed to measure the thermal conductivity of ceramic coatings. Since the method uses an infrared microscope for temperature measurement, coatings as thin as 20 μm can, in principle, be measured using this technique. This steady-state, comparative measurement method uses the...
Autor principal: | |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
2000
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4877160/ https://www.ncbi.nlm.nih.gov/pubmed/27551628 http://dx.doi.org/10.6028/jres.105.050 |
Sumario: | An apparatus has been developed to measure the thermal conductivity of ceramic coatings. Since the method uses an infrared microscope for temperature measurement, coatings as thin as 20 μm can, in principle, be measured using this technique. This steady-state, comparative measurement method uses the known thermal conductivity of the substrate material as the reference material for heat-flow measurement. The experimental method is validated by measuring a plasma-sprayed coating that has been previously measured using an absolute, steady-state measurement method. The new measurement method has a relative standard uncertainty of about 10 %. The measurement of the plasma-sprayed coating gives 0.58 W·m(−1)·K(−l) which compares well with the 0.62 W·m(−1)·K(−l) measured using the absolute method. |
---|