Cargando…

Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard

BACKGROUND: The purpose of the study is to evaluate the physical performance of a Biograph mCT Flow 64-4R PET/CT system (Siemens Healthcare, Germany) and to compare clinical image quality in step-and-shoot (SS) and continuous table motion (CTM) acquisitions. METHODS: The spatial resolution, sensitiv...

Descripción completa

Detalles Bibliográficos
Autores principales: Rausch, Ivo, Cal-González, Jacobo, Dapra, David, Gallowitsch, Hans Jürgen, Lind, Peter, Beyer, Thomas, Minear, Gregory
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4883615/
https://www.ncbi.nlm.nih.gov/pubmed/26501827
http://dx.doi.org/10.1186/s40658-015-0132-1
_version_ 1782434280571404288
author Rausch, Ivo
Cal-González, Jacobo
Dapra, David
Gallowitsch, Hans Jürgen
Lind, Peter
Beyer, Thomas
Minear, Gregory
author_facet Rausch, Ivo
Cal-González, Jacobo
Dapra, David
Gallowitsch, Hans Jürgen
Lind, Peter
Beyer, Thomas
Minear, Gregory
author_sort Rausch, Ivo
collection PubMed
description BACKGROUND: The purpose of the study is to evaluate the physical performance of a Biograph mCT Flow 64-4R PET/CT system (Siemens Healthcare, Germany) and to compare clinical image quality in step-and-shoot (SS) and continuous table motion (CTM) acquisitions. METHODS: The spatial resolution, sensitivity, count rate curves, and Image Quality (IQ) parameters following the National Electrical Manufactures Association (NEMA) NU2-2012 standard were evaluated. For resolution measurements, an (18)F point source inside a glass capillary tube was used. Sensitivity measurements were based on a 70-cm-long polyethylene tube, filled with 4.5 MBq of FDG. Scatter fraction and count rates were measured using a 70-cm-long polyethylene cylinder with a diameter of 20 cm and a line source (1.04 GBq of FDG) inserted axially into the cylinder 4.5 cm off-centered. A NEMA IQ phantom containing six spheres (10- to 37-mm diameter) was used for the evaluation of the image quality. First, a single-bed scan was acquired (NEMA standard), followed by a two-bed scan (4 min each) of the IQ phantom with the image plane containing the spheres centered in the overlap region of the two bed positions. In addition, a scan of the same region in CTM mode was performed with a table speed of 0.6 mm/s. Furthermore, two patient scans were performed in CTM and SS mode. Image contrasts and patient images were compared between SS and CTM acquisitions. RESULTS: Full Width Half Maximum (FWHM) of the spatial resolution ranged from 4.3 to 7.8 mm (radial distance 1 to 20 cm). The measured sensitivity was 9.6 kcps/MBq, both at the center of the FOV and 10 cm off-center. The measured noise equivalent count rate (NECR) peak was 185 kcps at 29.0 kBq/ml. The scatter fraction was 33.5 %. Image contrast recovery values (sphere-to-background of 8:1) were between 42 % (10-mm sphere) to 79 % (37-mm sphere). The background variability was between 2.1 and 5.3 % (SS) and between 2.4 and 6.9 % (CTM). No significant difference in image quality was observed between SS and CTM mode. CONCLUSIONS: The spatial resolution, sensitivity, scatter fraction, and count rates were in concordance with the published values for the predecessor system, the Biograph mCT. Contrast recovery values as well as image quality obtained in SS and CTM acquisition modes were similar.
format Online
Article
Text
id pubmed-4883615
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-48836152016-06-21 Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard Rausch, Ivo Cal-González, Jacobo Dapra, David Gallowitsch, Hans Jürgen Lind, Peter Beyer, Thomas Minear, Gregory EJNMMI Phys Original Research BACKGROUND: The purpose of the study is to evaluate the physical performance of a Biograph mCT Flow 64-4R PET/CT system (Siemens Healthcare, Germany) and to compare clinical image quality in step-and-shoot (SS) and continuous table motion (CTM) acquisitions. METHODS: The spatial resolution, sensitivity, count rate curves, and Image Quality (IQ) parameters following the National Electrical Manufactures Association (NEMA) NU2-2012 standard were evaluated. For resolution measurements, an (18)F point source inside a glass capillary tube was used. Sensitivity measurements were based on a 70-cm-long polyethylene tube, filled with 4.5 MBq of FDG. Scatter fraction and count rates were measured using a 70-cm-long polyethylene cylinder with a diameter of 20 cm and a line source (1.04 GBq of FDG) inserted axially into the cylinder 4.5 cm off-centered. A NEMA IQ phantom containing six spheres (10- to 37-mm diameter) was used for the evaluation of the image quality. First, a single-bed scan was acquired (NEMA standard), followed by a two-bed scan (4 min each) of the IQ phantom with the image plane containing the spheres centered in the overlap region of the two bed positions. In addition, a scan of the same region in CTM mode was performed with a table speed of 0.6 mm/s. Furthermore, two patient scans were performed in CTM and SS mode. Image contrasts and patient images were compared between SS and CTM acquisitions. RESULTS: Full Width Half Maximum (FWHM) of the spatial resolution ranged from 4.3 to 7.8 mm (radial distance 1 to 20 cm). The measured sensitivity was 9.6 kcps/MBq, both at the center of the FOV and 10 cm off-center. The measured noise equivalent count rate (NECR) peak was 185 kcps at 29.0 kBq/ml. The scatter fraction was 33.5 %. Image contrast recovery values (sphere-to-background of 8:1) were between 42 % (10-mm sphere) to 79 % (37-mm sphere). The background variability was between 2.1 and 5.3 % (SS) and between 2.4 and 6.9 % (CTM). No significant difference in image quality was observed between SS and CTM mode. CONCLUSIONS: The spatial resolution, sensitivity, scatter fraction, and count rates were in concordance with the published values for the predecessor system, the Biograph mCT. Contrast recovery values as well as image quality obtained in SS and CTM acquisition modes were similar. Springer International Publishing 2015-10-26 /pmc/articles/PMC4883615/ /pubmed/26501827 http://dx.doi.org/10.1186/s40658-015-0132-1 Text en © Rausch et al. 2015 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Research
Rausch, Ivo
Cal-González, Jacobo
Dapra, David
Gallowitsch, Hans Jürgen
Lind, Peter
Beyer, Thomas
Minear, Gregory
Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard
title Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard
title_full Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard
title_fullStr Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard
title_full_unstemmed Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard
title_short Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard
title_sort performance evaluation of the biograph mct flow pet/ct system according to the nema nu2-2012 standard
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4883615/
https://www.ncbi.nlm.nih.gov/pubmed/26501827
http://dx.doi.org/10.1186/s40658-015-0132-1
work_keys_str_mv AT rauschivo performanceevaluationofthebiographmctflowpetctsystemaccordingtothenemanu22012standard
AT calgonzalezjacobo performanceevaluationofthebiographmctflowpetctsystemaccordingtothenemanu22012standard
AT dapradavid performanceevaluationofthebiographmctflowpetctsystemaccordingtothenemanu22012standard
AT gallowitschhansjurgen performanceevaluationofthebiographmctflowpetctsystemaccordingtothenemanu22012standard
AT lindpeter performanceevaluationofthebiographmctflowpetctsystemaccordingtothenemanu22012standard
AT beyerthomas performanceevaluationofthebiographmctflowpetctsystemaccordingtothenemanu22012standard
AT mineargregory performanceevaluationofthebiographmctflowpetctsystemaccordingtothenemanu22012standard