Preparation and Calibration of Carrier-Free (209)Po Solution Standards
Carrier-free (209)Po solution standards have been prepared and calibrated. The standards, which will be disseminated by the National Institute of Standards and Technology as Standard Reference Material SRM 4326, consist of (5.1597 ±0.0024) g of a solution of polonium in nominal 2 mol · L(−1) hydroch...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
1995
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4887221/ https://www.ncbi.nlm.nih.gov/pubmed/29151724 http://dx.doi.org/10.6028/jres.100.002 |
_version_ | 1782434705856004096 |
---|---|
author | Collé, R. Lin, Zhichao Schima, F. J. Hodge, P. A. Thomas, J. W. L. Hutchinson, J. M. R. Coursey, B. M. |
author_facet | Collé, R. Lin, Zhichao Schima, F. J. Hodge, P. A. Thomas, J. W. L. Hutchinson, J. M. R. Coursey, B. M. |
author_sort | Collé, R. |
collection | PubMed |
description | Carrier-free (209)Po solution standards have been prepared and calibrated. The standards, which will be disseminated by the National Institute of Standards and Technology as Standard Reference Material SRM 4326, consist of (5.1597 ±0.0024) g of a solution of polonium in nominal 2 mol · L(−1) hydrochloric acid (having a solution density of (1.031±0.004) g · mL(−1) at 22 °C) that is contained in 5 mL flame-sealed borosilicate glass ampoules, and are certified to contain a (209)Po alpha-particle emission rate concentration of (85.42±0.29) s(−1) · g(−1) (corresponding to a (209)Po activity concentration of (85.83 ±0.30) Bq · g(−1)) as of the reference time of 1200 EST 15 March 1994. The calibration was based on 4πα liquid scintillation (LS) measurements with two different LS counting systems and under wide variations in measurement and sample conditions. Confirmatory measurements by 2πα gas-flow proportional counting were also performed. The only known radionuclidic impurity, based on α- and photon-emission spectrometry, is a trace quantity of (208)Po. The (208)Po to (209)Po impurity ratio as of the reference time was 0.00124 ±0.00020. All of the above cited uncertainty intervals correspond to a combined standard uncertainty multiplied by a coverage factor of k = 2. Although (209)Po is nearly a pure α emitter with only a weak electron capture branch to (209)Bi, LS measurements of the (209)Po a decay are confounded by an a transition to a 2.3 keV (J(π)= 1/2(−)) level in (205)Pb which was previously unknown to be a delayed isomeric state. |
format | Online Article Text |
id | pubmed-4887221 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 1995 |
publisher | [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology |
record_format | MEDLINE/PubMed |
spelling | pubmed-48872212017-11-17 Preparation and Calibration of Carrier-Free (209)Po Solution Standards Collé, R. Lin, Zhichao Schima, F. J. Hodge, P. A. Thomas, J. W. L. Hutchinson, J. M. R. Coursey, B. M. J Res Natl Inst Stand Technol Article Carrier-free (209)Po solution standards have been prepared and calibrated. The standards, which will be disseminated by the National Institute of Standards and Technology as Standard Reference Material SRM 4326, consist of (5.1597 ±0.0024) g of a solution of polonium in nominal 2 mol · L(−1) hydrochloric acid (having a solution density of (1.031±0.004) g · mL(−1) at 22 °C) that is contained in 5 mL flame-sealed borosilicate glass ampoules, and are certified to contain a (209)Po alpha-particle emission rate concentration of (85.42±0.29) s(−1) · g(−1) (corresponding to a (209)Po activity concentration of (85.83 ±0.30) Bq · g(−1)) as of the reference time of 1200 EST 15 March 1994. The calibration was based on 4πα liquid scintillation (LS) measurements with two different LS counting systems and under wide variations in measurement and sample conditions. Confirmatory measurements by 2πα gas-flow proportional counting were also performed. The only known radionuclidic impurity, based on α- and photon-emission spectrometry, is a trace quantity of (208)Po. The (208)Po to (209)Po impurity ratio as of the reference time was 0.00124 ±0.00020. All of the above cited uncertainty intervals correspond to a combined standard uncertainty multiplied by a coverage factor of k = 2. Although (209)Po is nearly a pure α emitter with only a weak electron capture branch to (209)Bi, LS measurements of the (209)Po a decay are confounded by an a transition to a 2.3 keV (J(π)= 1/2(−)) level in (205)Pb which was previously unknown to be a delayed isomeric state. [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1995 /pmc/articles/PMC4887221/ /pubmed/29151724 http://dx.doi.org/10.6028/jres.100.002 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Institute of Standards and Technology is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright. |
spellingShingle | Article Collé, R. Lin, Zhichao Schima, F. J. Hodge, P. A. Thomas, J. W. L. Hutchinson, J. M. R. Coursey, B. M. Preparation and Calibration of Carrier-Free (209)Po Solution Standards |
title | Preparation and Calibration of Carrier-Free (209)Po Solution Standards |
title_full | Preparation and Calibration of Carrier-Free (209)Po Solution Standards |
title_fullStr | Preparation and Calibration of Carrier-Free (209)Po Solution Standards |
title_full_unstemmed | Preparation and Calibration of Carrier-Free (209)Po Solution Standards |
title_short | Preparation and Calibration of Carrier-Free (209)Po Solution Standards |
title_sort | preparation and calibration of carrier-free (209)po solution standards |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4887221/ https://www.ncbi.nlm.nih.gov/pubmed/29151724 http://dx.doi.org/10.6028/jres.100.002 |
work_keys_str_mv | AT coller preparationandcalibrationofcarrierfree209posolutionstandards AT linzhichao preparationandcalibrationofcarrierfree209posolutionstandards AT schimafj preparationandcalibrationofcarrierfree209posolutionstandards AT hodgepa preparationandcalibrationofcarrierfree209posolutionstandards AT thomasjwl preparationandcalibrationofcarrierfree209posolutionstandards AT hutchinsonjmr preparationandcalibrationofcarrierfree209posolutionstandards AT courseybm preparationandcalibrationofcarrierfree209posolutionstandards |