Cargando…
Creep and Creep Recovery Response of Load Cells Tested According to U.S. and International Evaluation Procedures
The 60 min creep data from National Type Evaluation Procedure (NTEP) tests performed at the National Institute of Standards and Technology (NIST) on 65 load cells have been analyzed in order to compare their creep and creep recovery responses, and to compare the 60 min creep with creep over shorter...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
1997
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4894594/ https://www.ncbi.nlm.nih.gov/pubmed/27805151 http://dx.doi.org/10.6028/jres.102.024 |
_version_ | 1782435693664927744 |
---|---|
author | Bartel, Thomas W. Yaniv, Simone L. |
author_facet | Bartel, Thomas W. Yaniv, Simone L. |
author_sort | Bartel, Thomas W. |
collection | PubMed |
description | The 60 min creep data from National Type Evaluation Procedure (NTEP) tests performed at the National Institute of Standards and Technology (NIST) on 65 load cells have been analyzed in order to compare their creep and creep recovery responses, and to compare the 60 min creep with creep over shorter time periods. To facilitate this comparison the data were fitted to a multiple-term exponential equation, which adequately describes the creep and creep recovery responses of load cells. The use of such a curve fit reduces the effect of the random error in the indicator readings on the calculated values of the load cell creep. Examination of the fitted curves show that the creep recovery responses, after inversion by a change in sign, are generally similar in shape to the creep response, but smaller in magnitude. The average ratio of the absolute value of the maximum creep recovery to the maximum creep is 0.86; however, no reliable correlation between creep and creep recovery can be drawn from the data. The fitted curves were also used to compare the 60 min creep of the NTEP analysis with the 30 min creep and other parameters calculated according to the Organization Internationale de Métrologie Légale (OIML) R 60 analysis. The average ratio of the 30 min creep value to the 60 min value is 0.84. The OIML class C creep tolerance is less than 0.5 of the NTEP tolerance for classes III and III L. |
format | Online Article Text |
id | pubmed-4894594 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 1997 |
publisher | [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology |
record_format | MEDLINE/PubMed |
spelling | pubmed-48945942016-10-28 Creep and Creep Recovery Response of Load Cells Tested According to U.S. and International Evaluation Procedures Bartel, Thomas W. Yaniv, Simone L. J Res Natl Inst Stand Technol Article The 60 min creep data from National Type Evaluation Procedure (NTEP) tests performed at the National Institute of Standards and Technology (NIST) on 65 load cells have been analyzed in order to compare their creep and creep recovery responses, and to compare the 60 min creep with creep over shorter time periods. To facilitate this comparison the data were fitted to a multiple-term exponential equation, which adequately describes the creep and creep recovery responses of load cells. The use of such a curve fit reduces the effect of the random error in the indicator readings on the calculated values of the load cell creep. Examination of the fitted curves show that the creep recovery responses, after inversion by a change in sign, are generally similar in shape to the creep response, but smaller in magnitude. The average ratio of the absolute value of the maximum creep recovery to the maximum creep is 0.86; however, no reliable correlation between creep and creep recovery can be drawn from the data. The fitted curves were also used to compare the 60 min creep of the NTEP analysis with the 30 min creep and other parameters calculated according to the Organization Internationale de Métrologie Légale (OIML) R 60 analysis. The average ratio of the 30 min creep value to the 60 min value is 0.84. The OIML class C creep tolerance is less than 0.5 of the NTEP tolerance for classes III and III L. [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1997 /pmc/articles/PMC4894594/ /pubmed/27805151 http://dx.doi.org/10.6028/jres.102.024 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Institute of Standards and Technology is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright. |
spellingShingle | Article Bartel, Thomas W. Yaniv, Simone L. Creep and Creep Recovery Response of Load Cells Tested According to U.S. and International Evaluation Procedures |
title | Creep and Creep Recovery Response of Load Cells Tested According to U.S. and International Evaluation Procedures |
title_full | Creep and Creep Recovery Response of Load Cells Tested According to U.S. and International Evaluation Procedures |
title_fullStr | Creep and Creep Recovery Response of Load Cells Tested According to U.S. and International Evaluation Procedures |
title_full_unstemmed | Creep and Creep Recovery Response of Load Cells Tested According to U.S. and International Evaluation Procedures |
title_short | Creep and Creep Recovery Response of Load Cells Tested According to U.S. and International Evaluation Procedures |
title_sort | creep and creep recovery response of load cells tested according to u.s. and international evaluation procedures |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4894594/ https://www.ncbi.nlm.nih.gov/pubmed/27805151 http://dx.doi.org/10.6028/jres.102.024 |
work_keys_str_mv | AT bartelthomasw creepandcreeprecoveryresponseofloadcellstestedaccordingtousandinternationalevaluationprocedures AT yanivsimonel creepandcreeprecoveryresponseofloadcellstestedaccordingtousandinternationalevaluationprocedures |