Improving Image Quality of On-Board Cone-Beam CT in Radiation Therapy Using Image Information Provided by Planning Multi-Detector CT: A Phantom Study

PURPOSE: The aim of this study was to improve the image quality of cone-beam computed tomography (CBCT) mounted on the gantry of a linear accelerator used in radiation therapy based on the image information provided by planning multi-detector CT (MDCT). METHODS: MDCT-based shading correction for CBC...

Descripción completa

Detalles Bibliográficos
Autores principales: Yang, Ching-Ching, Chen, Fong-Lin, Lo, Yeh-Chi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4900643/
https://www.ncbi.nlm.nih.gov/pubmed/27280593
http://dx.doi.org/10.1371/journal.pone.0157072
_version_ 1782436677085560832
author Yang, Ching-Ching
Chen, Fong-Lin
Lo, Yeh-Chi
author_facet Yang, Ching-Ching
Chen, Fong-Lin
Lo, Yeh-Chi
author_sort Yang, Ching-Ching
collection PubMed
description PURPOSE: The aim of this study was to improve the image quality of cone-beam computed tomography (CBCT) mounted on the gantry of a linear accelerator used in radiation therapy based on the image information provided by planning multi-detector CT (MDCT). METHODS: MDCT-based shading correction for CBCT and virtual monochromatic CT (VMCT) synthesized using the dual-energy method were performed. In VMCT, the high-energy data were obtained from CBCT, while the low-energy data were obtained from MDCT. An electron density phantom was used to investigate the efficacy of shading correction and VMCT on improving the target detectability, Hounsfield unit (HU) accuracy and variation, which were quantified by calculating the contrast-to-noise ratio (CNR), the percent difference (%Diff) and the standard deviation of the CT numbers for tissue equivalent background material, respectively. Treatment plan studies for a chest phantom were conducted to investigate the effects of image quality improvement on dose planning. RESULTS: For the electron density phantom, the mean value of CNR was 17.84, 26.78 and 34.31 in CBCT, shading-corrected CBCT and VMCT, respectively. The mean value of %Diff was 152.67%, 11.93% and 7.66% in CBCT, shading-corrected CBCT and VMCT, respectively. The standard deviation within a uniform background of CBCT, shading-corrected CBCT and VMCT was 85, 23 and 15 HU, respectively. With regards to the chest phantom, the monitor unit (MU) difference between the treatment plan calculated using MDCT and those based on CBCT, shading corrected CBCT and VMCT was 6.32%, 1.05% and 0.94%, respectively. CONCLUSIONS: Enhancement of image quality in on-board CBCT can contribute to daily patient setup and adaptive dose delivery, thus enabling higher confidence in patient treatment accuracy in radiation therapy. Based on our results, VMCT has the highest image quality, followed by the shading corrected CBCT and the original CBCT. The research results presented in this study should be able to provide a route to reach a high level of image quality for CBCT imaging in radiation oncology.
format Online
Article
Text
id pubmed-4900643
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-49006432016-06-24 Improving Image Quality of On-Board Cone-Beam CT in Radiation Therapy Using Image Information Provided by Planning Multi-Detector CT: A Phantom Study Yang, Ching-Ching Chen, Fong-Lin Lo, Yeh-Chi PLoS One Research Article PURPOSE: The aim of this study was to improve the image quality of cone-beam computed tomography (CBCT) mounted on the gantry of a linear accelerator used in radiation therapy based on the image information provided by planning multi-detector CT (MDCT). METHODS: MDCT-based shading correction for CBCT and virtual monochromatic CT (VMCT) synthesized using the dual-energy method were performed. In VMCT, the high-energy data were obtained from CBCT, while the low-energy data were obtained from MDCT. An electron density phantom was used to investigate the efficacy of shading correction and VMCT on improving the target detectability, Hounsfield unit (HU) accuracy and variation, which were quantified by calculating the contrast-to-noise ratio (CNR), the percent difference (%Diff) and the standard deviation of the CT numbers for tissue equivalent background material, respectively. Treatment plan studies for a chest phantom were conducted to investigate the effects of image quality improvement on dose planning. RESULTS: For the electron density phantom, the mean value of CNR was 17.84, 26.78 and 34.31 in CBCT, shading-corrected CBCT and VMCT, respectively. The mean value of %Diff was 152.67%, 11.93% and 7.66% in CBCT, shading-corrected CBCT and VMCT, respectively. The standard deviation within a uniform background of CBCT, shading-corrected CBCT and VMCT was 85, 23 and 15 HU, respectively. With regards to the chest phantom, the monitor unit (MU) difference between the treatment plan calculated using MDCT and those based on CBCT, shading corrected CBCT and VMCT was 6.32%, 1.05% and 0.94%, respectively. CONCLUSIONS: Enhancement of image quality in on-board CBCT can contribute to daily patient setup and adaptive dose delivery, thus enabling higher confidence in patient treatment accuracy in radiation therapy. Based on our results, VMCT has the highest image quality, followed by the shading corrected CBCT and the original CBCT. The research results presented in this study should be able to provide a route to reach a high level of image quality for CBCT imaging in radiation oncology. Public Library of Science 2016-06-09 /pmc/articles/PMC4900643/ /pubmed/27280593 http://dx.doi.org/10.1371/journal.pone.0157072 Text en © 2016 Yang et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Yang, Ching-Ching
Chen, Fong-Lin
Lo, Yeh-Chi
Improving Image Quality of On-Board Cone-Beam CT in Radiation Therapy Using Image Information Provided by Planning Multi-Detector CT: A Phantom Study
title Improving Image Quality of On-Board Cone-Beam CT in Radiation Therapy Using Image Information Provided by Planning Multi-Detector CT: A Phantom Study
title_full Improving Image Quality of On-Board Cone-Beam CT in Radiation Therapy Using Image Information Provided by Planning Multi-Detector CT: A Phantom Study
title_fullStr Improving Image Quality of On-Board Cone-Beam CT in Radiation Therapy Using Image Information Provided by Planning Multi-Detector CT: A Phantom Study
title_full_unstemmed Improving Image Quality of On-Board Cone-Beam CT in Radiation Therapy Using Image Information Provided by Planning Multi-Detector CT: A Phantom Study
title_short Improving Image Quality of On-Board Cone-Beam CT in Radiation Therapy Using Image Information Provided by Planning Multi-Detector CT: A Phantom Study
title_sort improving image quality of on-board cone-beam ct in radiation therapy using image information provided by planning multi-detector ct: a phantom study
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4900643/
https://www.ncbi.nlm.nih.gov/pubmed/27280593
http://dx.doi.org/10.1371/journal.pone.0157072
work_keys_str_mv AT yangchingching improvingimagequalityofonboardconebeamctinradiationtherapyusingimageinformationprovidedbyplanningmultidetectorctaphantomstudy
AT chenfonglin improvingimagequalityofonboardconebeamctinradiationtherapyusingimageinformationprovidedbyplanningmultidetectorctaphantomstudy
AT loyehchi improvingimagequalityofonboardconebeamctinradiationtherapyusingimageinformationprovidedbyplanningmultidetectorctaphantomstudy