Cargando…

An Oil-Bath-Based 293 K to 473 K Blackbody Source

A high temperature oil-bath-based-black-body source has been designed and constructed in the Radiometric Physics Division at the National Institute of Standards and Technology, Gaithersburg, MD. The goal of this work was to design a large aperture blackbody source with highly uniform radiance across...

Descripción completa

Detalles Bibliográficos
Autor principal: Fowler, Joel B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1996
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4907614/
https://www.ncbi.nlm.nih.gov/pubmed/27805082
http://dx.doi.org/10.6028/jres.101.062
_version_ 1782437562186465280
author Fowler, Joel B.
author_facet Fowler, Joel B.
author_sort Fowler, Joel B.
collection PubMed
description A high temperature oil-bath-based-black-body source has been designed and constructed in the Radiometric Physics Division at the National Institute of Standards and Technology, Gaithersburg, MD. The goal of this work was to design a large aperture blackbody source with highly uniform radiance across the aperture, good temporal stability, and good reproducibility. This blackbody source operates in the 293 K to 473 K range with blackbody temperature combined standard uncertainties of 7.2 mK to 30.9 mK. The calculated emissivity of this source is 0.9997 with a standard uncertainty of 0.0003. With a 50 mm limiting aperture at the cavity entrance, the emissivity increases to 0.99996.
format Online
Article
Text
id pubmed-4907614
institution National Center for Biotechnology Information
language English
publishDate 1996
publisher [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
record_format MEDLINE/PubMed
spelling pubmed-49076142016-10-28 An Oil-Bath-Based 293 K to 473 K Blackbody Source Fowler, Joel B. J Res Natl Inst Stand Technol Article A high temperature oil-bath-based-black-body source has been designed and constructed in the Radiometric Physics Division at the National Institute of Standards and Technology, Gaithersburg, MD. The goal of this work was to design a large aperture blackbody source with highly uniform radiance across the aperture, good temporal stability, and good reproducibility. This blackbody source operates in the 293 K to 473 K range with blackbody temperature combined standard uncertainties of 7.2 mK to 30.9 mK. The calculated emissivity of this source is 0.9997 with a standard uncertainty of 0.0003. With a 50 mm limiting aperture at the cavity entrance, the emissivity increases to 0.99996. [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1996 /pmc/articles/PMC4907614/ /pubmed/27805082 http://dx.doi.org/10.6028/jres.101.062 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Institute of Standards and Technology is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright.
spellingShingle Article
Fowler, Joel B.
An Oil-Bath-Based 293 K to 473 K Blackbody Source
title An Oil-Bath-Based 293 K to 473 K Blackbody Source
title_full An Oil-Bath-Based 293 K to 473 K Blackbody Source
title_fullStr An Oil-Bath-Based 293 K to 473 K Blackbody Source
title_full_unstemmed An Oil-Bath-Based 293 K to 473 K Blackbody Source
title_short An Oil-Bath-Based 293 K to 473 K Blackbody Source
title_sort oil-bath-based 293 k to 473 k blackbody source
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4907614/
https://www.ncbi.nlm.nih.gov/pubmed/27805082
http://dx.doi.org/10.6028/jres.101.062
work_keys_str_mv AT fowlerjoelb anoilbathbased293kto473kblackbodysource
AT fowlerjoelb oilbathbased293kto473kblackbodysource