Comparative evaluation of applanation and indentation tonometers in a community ophthalmology setting in Southern India

PURPOSE: Measurement of intraocular pressure (IOP) is one of the basic investigations in a general ophthalmic workup. In this study, we attempt to determine the agreement in the measurement of IOP obtained by Perkin’s applanation tonometer, noncontact tonometer and Schiotz indentation tonometer in p...

Descripción completa

Detalles Bibliográficos
Autores principales: Nagarajan, Swathi, Velayutham, Veerabahu, Ezhumalai, G.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4907972/
https://www.ncbi.nlm.nih.gov/pubmed/27330381
http://dx.doi.org/10.1016/j.sjopt.2015.11.002
_version_ 1782437603372433408
author Nagarajan, Swathi
Velayutham, Veerabahu
Ezhumalai, G.
author_facet Nagarajan, Swathi
Velayutham, Veerabahu
Ezhumalai, G.
author_sort Nagarajan, Swathi
collection PubMed
description PURPOSE: Measurement of intraocular pressure (IOP) is one of the basic investigations in a general ophthalmic workup. In this study, we attempt to determine the agreement in the measurement of IOP obtained by Perkin’s applanation tonometer, noncontact tonometer and Schiotz indentation tonometer in patients attending general ophthalmology OPD in a tertiary care centre in South India and its use in a community ophthalmology setting. METHODS: A cross-sectional analytical study in which IOP was measured in patients using the three tonometers. Central corneal thickness (CCT) was measured using Ultrasonic pachymetry. Bland Altman analysis was done to evaluate the agreement between instruments. RESULTS: 800 eyes of 400 patients were included in the study. By Bland Altman method, Schiotz indentation tonometer was found to have better correlation to IOP obtained by Perkin’s applanation tonometer. Schiotz indentation tonometer was found to be most accurate when CCT was in the range of 501–550 μm and noncontact tonometer was found to be least accurate when CCT was greater than 600 microns. On comparing correlation at different age groups, both the methods had better correlation at <40 years age group. CONCLUSION: Both the tonometers showed a significant correlation with the gold standard technique (Perkin’s applanation tonometer) over a range of IOP and CCT with the Schiotz tonometer better than the NCT. This study proves that Schiotz tonometer can be recommended as a reliable screening tool in community outreach ophthalmology services. The twin advantages of portability and availability make the Schiotz tonometer a popular choice among ophthalmology trainees and optometrists in a developing country like India.
format Online
Article
Text
id pubmed-4907972
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-49079722016-06-21 Comparative evaluation of applanation and indentation tonometers in a community ophthalmology setting in Southern India Nagarajan, Swathi Velayutham, Veerabahu Ezhumalai, G. Saudi J Ophthalmol Original Article PURPOSE: Measurement of intraocular pressure (IOP) is one of the basic investigations in a general ophthalmic workup. In this study, we attempt to determine the agreement in the measurement of IOP obtained by Perkin’s applanation tonometer, noncontact tonometer and Schiotz indentation tonometer in patients attending general ophthalmology OPD in a tertiary care centre in South India and its use in a community ophthalmology setting. METHODS: A cross-sectional analytical study in which IOP was measured in patients using the three tonometers. Central corneal thickness (CCT) was measured using Ultrasonic pachymetry. Bland Altman analysis was done to evaluate the agreement between instruments. RESULTS: 800 eyes of 400 patients were included in the study. By Bland Altman method, Schiotz indentation tonometer was found to have better correlation to IOP obtained by Perkin’s applanation tonometer. Schiotz indentation tonometer was found to be most accurate when CCT was in the range of 501–550 μm and noncontact tonometer was found to be least accurate when CCT was greater than 600 microns. On comparing correlation at different age groups, both the methods had better correlation at <40 years age group. CONCLUSION: Both the tonometers showed a significant correlation with the gold standard technique (Perkin’s applanation tonometer) over a range of IOP and CCT with the Schiotz tonometer better than the NCT. This study proves that Schiotz tonometer can be recommended as a reliable screening tool in community outreach ophthalmology services. The twin advantages of portability and availability make the Schiotz tonometer a popular choice among ophthalmology trainees and optometrists in a developing country like India. Elsevier 2016 2015-11-19 /pmc/articles/PMC4907972/ /pubmed/27330381 http://dx.doi.org/10.1016/j.sjopt.2015.11.002 Text en © 2015 Production and hosting by Elsevier B.V. on behalf of Saudi Ophthalmological Society, King Saud University. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Article
Nagarajan, Swathi
Velayutham, Veerabahu
Ezhumalai, G.
Comparative evaluation of applanation and indentation tonometers in a community ophthalmology setting in Southern India
title Comparative evaluation of applanation and indentation tonometers in a community ophthalmology setting in Southern India
title_full Comparative evaluation of applanation and indentation tonometers in a community ophthalmology setting in Southern India
title_fullStr Comparative evaluation of applanation and indentation tonometers in a community ophthalmology setting in Southern India
title_full_unstemmed Comparative evaluation of applanation and indentation tonometers in a community ophthalmology setting in Southern India
title_short Comparative evaluation of applanation and indentation tonometers in a community ophthalmology setting in Southern India
title_sort comparative evaluation of applanation and indentation tonometers in a community ophthalmology setting in southern india
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4907972/
https://www.ncbi.nlm.nih.gov/pubmed/27330381
http://dx.doi.org/10.1016/j.sjopt.2015.11.002
work_keys_str_mv AT nagarajanswathi comparativeevaluationofapplanationandindentationtonometersinacommunityophthalmologysettinginsouthernindia
AT velayuthamveerabahu comparativeevaluationofapplanationandindentationtonometersinacommunityophthalmologysettinginsouthernindia
AT ezhumalaig comparativeevaluationofapplanationandindentationtonometersinacommunityophthalmologysettinginsouthernindia