Measuring Low Frequency Tilts

A borehole tiltmeter with a sensitivity of a few nanoradians is described. It is composed of two orthogonal horizontal pendulums with free periods of 1 s. The pendulums are insensitive to barometric pressure fluctuations, and the measured temperature coefficient is less than 30 nrad/°C. The range of...

Descripción completa

Detalles Bibliográficos
Autores principales: Kohl, M. L., Levine, J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1993
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4909177/
https://www.ncbi.nlm.nih.gov/pubmed/28053466
http://dx.doi.org/10.6028/jres.098.014
Descripción
Sumario:A borehole tiltmeter with a sensitivity of a few nanoradians is described. It is composed of two orthogonal horizontal pendulums with free periods of 1 s. The pendulums are insensitive to barometric pressure fluctuations, and the measured temperature coefficient is less than 30 nrad/°C. The range of the pendulums is about ±5 μ rad, and their response is linear within 1% and stable over several years. The performance of the tiltmeter in the field was evaluated using tidal data obtained from a closely spaced array of boreholes in Southern California. The long-term stability of the tiltmeter is generally better than 1 μ rad/yr. The data also indicate that instruments in boreholes at least 24 m deepare independent of surface effects. Several different capsules designed to couple the instrument to the surrounding material have been tested. In addition, an experimental method for estimating the magnitudes of local perturbation in the regional tilt field is described.