Haemoglobin J-Baltimore can be detected by HbA(1c) electropherogram but with underestimated HbA(1c) value

Glycated haemoglobin (HbA(1c)) is considered the gold standard for assessing diabetes compensation and treatment. In addition, fortuitous detection of haemoglobin variants during HbA(1c) measurement is not rare. Recently, two publications reported different conclusions on accuracy of HbA(1c) value u...

Descripción completa

Detalles Bibliográficos
Autores principales: Brunel, Valéry, Lahary, Agnčs, Chagraoui, Abdeslam, Thuillez, Christian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4910267/
https://www.ncbi.nlm.nih.gov/pubmed/27346969
http://dx.doi.org/10.11613/BM.2016.026
Descripción
Sumario:Glycated haemoglobin (HbA(1c)) is considered the gold standard for assessing diabetes compensation and treatment. In addition, fortuitous detection of haemoglobin variants during HbA(1c) measurement is not rare. Recently, two publications reported different conclusions on accuracy of HbA(1c) value using capillary electrophoresis method in presence of haemoglobin J-Baltimore (HbJ).
Here we describe the fortuitous detection of unknown HbJ using capillary electrophoresis for measurement of HbA(1c). A patient followed for gestational diabetes in our laboratory presented unknown haemoglobin on Capillarys 2 Flex Piercing analyser which was identified as HbJ. HbJ is not associated with haematological abnormalities. High Performance Liquid Chromatography methods are known to possibly underestimate HbA(1c) value in the presence of this variant. This variant and its glycated form are clearly distinguished on electropherogram but HbJ was responsible for underestimating the true area of HbA(1c).
Capillary electrophoresis is a good method for detecting HbJ but does not seem suitable for evaluation of HbA(1C) value in patients in presence of HbJ variant.