EFFECT OF RADIATION DOSE LEVEL ON ACCURACY AND PRECISION OF MANUAL SIZE MEASUREMENTS IN CHEST TOMOSYNTHESIS EVALUATED USING SIMULATED PULMONARY NODULES

The aim of the present study was to investigate the dependency of the accuracy and precision of nodule diameter measurements on the radiation dose level in chest tomosynthesis. Artificial ellipsoid-shaped nodules with known dimensions were inserted in clinical chest tomosynthesis images. Noise was a...

Descripción completa

Detalles Bibliográficos
Autores principales: Söderman, Christina, Johnsson, Åse Allansdotter, Vikgren, Jenny, Norrlund, Rauni Rossi, Molnar, David, Svalkvist, Angelica, Månsson, Lars Gunnar, Båth, Magnus
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4911967/
https://www.ncbi.nlm.nih.gov/pubmed/26994093
http://dx.doi.org/10.1093/rpd/ncw041
_version_ 1782438205054779392
author Söderman, Christina
Johnsson, Åse Allansdotter
Vikgren, Jenny
Norrlund, Rauni Rossi
Molnar, David
Svalkvist, Angelica
Månsson, Lars Gunnar
Båth, Magnus
author_facet Söderman, Christina
Johnsson, Åse Allansdotter
Vikgren, Jenny
Norrlund, Rauni Rossi
Molnar, David
Svalkvist, Angelica
Månsson, Lars Gunnar
Båth, Magnus
author_sort Söderman, Christina
collection PubMed
description The aim of the present study was to investigate the dependency of the accuracy and precision of nodule diameter measurements on the radiation dose level in chest tomosynthesis. Artificial ellipsoid-shaped nodules with known dimensions were inserted in clinical chest tomosynthesis images. Noise was added to the images in order to simulate radiation dose levels corresponding to effective doses for a standard-sized patient of 0.06 and 0.04 mSv. These levels were compared with the original dose level, corresponding to an effective dose of 0.12 mSv for a standard-sized patient. Four thoracic radiologists measured the longest diameter of the nodules. The study was restricted to nodules located in high-dose areas of the tomosynthesis projection radiographs. A significant decrease of the measurement accuracy and intraobserver variability was seen for the lowest dose level for a subset of the observers. No significant effect of dose level on the interobserver variability was found. The number of non-measurable small nodules (≤5 mm) was higher for the two lowest dose levels compared with the original dose level. In conclusion, for pulmonary nodules at positions in the lung corresponding to locations in high-dose areas of the projection radiographs, using a radiation dose level resulting in an effective dose of 0.06 mSv to a standard-sized patient may be possible in chest tomosynthesis without affecting the accuracy and precision of nodule diameter measurements to any large extent. However, an increasing number of non-measurable small nodules (≤5 mm) with decreasing radiation dose may raise some concerns regarding an applied general dose reduction for chest tomosynthesis examinations in the clinical praxis.
format Online
Article
Text
id pubmed-4911967
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-49119672016-06-20 EFFECT OF RADIATION DOSE LEVEL ON ACCURACY AND PRECISION OF MANUAL SIZE MEASUREMENTS IN CHEST TOMOSYNTHESIS EVALUATED USING SIMULATED PULMONARY NODULES Söderman, Christina Johnsson, Åse Allansdotter Vikgren, Jenny Norrlund, Rauni Rossi Molnar, David Svalkvist, Angelica Månsson, Lars Gunnar Båth, Magnus Radiat Prot Dosimetry Paper The aim of the present study was to investigate the dependency of the accuracy and precision of nodule diameter measurements on the radiation dose level in chest tomosynthesis. Artificial ellipsoid-shaped nodules with known dimensions were inserted in clinical chest tomosynthesis images. Noise was added to the images in order to simulate radiation dose levels corresponding to effective doses for a standard-sized patient of 0.06 and 0.04 mSv. These levels were compared with the original dose level, corresponding to an effective dose of 0.12 mSv for a standard-sized patient. Four thoracic radiologists measured the longest diameter of the nodules. The study was restricted to nodules located in high-dose areas of the tomosynthesis projection radiographs. A significant decrease of the measurement accuracy and intraobserver variability was seen for the lowest dose level for a subset of the observers. No significant effect of dose level on the interobserver variability was found. The number of non-measurable small nodules (≤5 mm) was higher for the two lowest dose levels compared with the original dose level. In conclusion, for pulmonary nodules at positions in the lung corresponding to locations in high-dose areas of the projection radiographs, using a radiation dose level resulting in an effective dose of 0.06 mSv to a standard-sized patient may be possible in chest tomosynthesis without affecting the accuracy and precision of nodule diameter measurements to any large extent. However, an increasing number of non-measurable small nodules (≤5 mm) with decreasing radiation dose may raise some concerns regarding an applied general dose reduction for chest tomosynthesis examinations in the clinical praxis. Oxford University Press 2016-06 2016-06-07 /pmc/articles/PMC4911967/ /pubmed/26994093 http://dx.doi.org/10.1093/rpd/ncw041 Text en © The Author 2016. Published by Oxford University Press. http://creativecommons.org/licenses/by/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Paper
Söderman, Christina
Johnsson, Åse Allansdotter
Vikgren, Jenny
Norrlund, Rauni Rossi
Molnar, David
Svalkvist, Angelica
Månsson, Lars Gunnar
Båth, Magnus
EFFECT OF RADIATION DOSE LEVEL ON ACCURACY AND PRECISION OF MANUAL SIZE MEASUREMENTS IN CHEST TOMOSYNTHESIS EVALUATED USING SIMULATED PULMONARY NODULES
title EFFECT OF RADIATION DOSE LEVEL ON ACCURACY AND PRECISION OF MANUAL SIZE MEASUREMENTS IN CHEST TOMOSYNTHESIS EVALUATED USING SIMULATED PULMONARY NODULES
title_full EFFECT OF RADIATION DOSE LEVEL ON ACCURACY AND PRECISION OF MANUAL SIZE MEASUREMENTS IN CHEST TOMOSYNTHESIS EVALUATED USING SIMULATED PULMONARY NODULES
title_fullStr EFFECT OF RADIATION DOSE LEVEL ON ACCURACY AND PRECISION OF MANUAL SIZE MEASUREMENTS IN CHEST TOMOSYNTHESIS EVALUATED USING SIMULATED PULMONARY NODULES
title_full_unstemmed EFFECT OF RADIATION DOSE LEVEL ON ACCURACY AND PRECISION OF MANUAL SIZE MEASUREMENTS IN CHEST TOMOSYNTHESIS EVALUATED USING SIMULATED PULMONARY NODULES
title_short EFFECT OF RADIATION DOSE LEVEL ON ACCURACY AND PRECISION OF MANUAL SIZE MEASUREMENTS IN CHEST TOMOSYNTHESIS EVALUATED USING SIMULATED PULMONARY NODULES
title_sort effect of radiation dose level on accuracy and precision of manual size measurements in chest tomosynthesis evaluated using simulated pulmonary nodules
topic Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4911967/
https://www.ncbi.nlm.nih.gov/pubmed/26994093
http://dx.doi.org/10.1093/rpd/ncw041
work_keys_str_mv AT sodermanchristina effectofradiationdoselevelonaccuracyandprecisionofmanualsizemeasurementsinchesttomosynthesisevaluatedusingsimulatedpulmonarynodules
AT johnssonaseallansdotter effectofradiationdoselevelonaccuracyandprecisionofmanualsizemeasurementsinchesttomosynthesisevaluatedusingsimulatedpulmonarynodules
AT vikgrenjenny effectofradiationdoselevelonaccuracyandprecisionofmanualsizemeasurementsinchesttomosynthesisevaluatedusingsimulatedpulmonarynodules
AT norrlundraunirossi effectofradiationdoselevelonaccuracyandprecisionofmanualsizemeasurementsinchesttomosynthesisevaluatedusingsimulatedpulmonarynodules
AT molnardavid effectofradiationdoselevelonaccuracyandprecisionofmanualsizemeasurementsinchesttomosynthesisevaluatedusingsimulatedpulmonarynodules
AT svalkvistangelica effectofradiationdoselevelonaccuracyandprecisionofmanualsizemeasurementsinchesttomosynthesisevaluatedusingsimulatedpulmonarynodules
AT manssonlarsgunnar effectofradiationdoselevelonaccuracyandprecisionofmanualsizemeasurementsinchesttomosynthesisevaluatedusingsimulatedpulmonarynodules
AT bathmagnus effectofradiationdoselevelonaccuracyandprecisionofmanualsizemeasurementsinchesttomosynthesisevaluatedusingsimulatedpulmonarynodules