Home monitoring report from a single lead Lumax DX implantable cardioverter defibrillator: New observations in a new system

A 56-year-old man underwent a single lead Lumax 640 DX implantable cardioverter defibrillator implantation for primary prevention of sudden cardiac death. A DX system consists of a single lead, which provides atrial as well as ventricular electrograms, and enhances atrial arrhythmia detection. Three...

Descripción completa

Detalles Bibliográficos
Autores principales: Konstantino, Yuval, Kleiman, Alex, Amit, Guy
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4913149/
https://www.ncbi.nlm.nih.gov/pubmed/27354868
http://dx.doi.org/10.1016/j.joa.2015.12.004
Descripción
Sumario:A 56-year-old man underwent a single lead Lumax 640 DX implantable cardioverter defibrillator implantation for primary prevention of sudden cardiac death. A DX system consists of a single lead, which provides atrial as well as ventricular electrograms, and enhances atrial arrhythmia detection. Three months after the implantation, high-frequency episodes were detected on the far field and the atrial channels, but not on the bipolar right ventricular channel; these were classified as atrial tachycardia. In the present report, we discussed the unusual pattern of the artifacts that was related to an electromagnetic interference detected by the novel DX system.