The Characterization of a Piston Displacement-Type Flowmeter Calibration Facility and the Calibration and Use of Pulsed Output Type Flowmeters

Critical measurement performance of fluid flowmeters requires proper and quantified verification data. These data should be generated using calibration and traceability techniques established for these verification purposes. In these calibration techniques, the calibration facility should be well-ch...

Descripción completa

Detalles Bibliográficos
Autor principal: Mattingly, G. E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1992
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4914226/
https://www.ncbi.nlm.nih.gov/pubmed/28053444
http://dx.doi.org/10.6028/jres.097.023
Descripción
Sumario:Critical measurement performance of fluid flowmeters requires proper and quantified verification data. These data should be generated using calibration and traceability techniques established for these verification purposes. In these calibration techniques, the calibration facility should be well-characterized and its components and performance properly traced to pertinent higher standards. The use of this calibrator to calibrate flowmeters should be appropriately established and the manner in which the calibrated flowmeter is used should be specified in accord with the conditions of the calibration. These three steps: 1) characterizing the calibration facility itself, 2) using the characterized facility to calibrate a flowmeter, and 3) using the calibrated flowmeter to make a measurement are described and the pertinent equations are given for an encoded-stroke, piston displacement-type calibrator and a pulsed output flowmeter. It is concluded that, given these equations and proper instrumentation of this type of calibrator, very high levels of performance can be attained and, in turn, these can be used to achieve high fluid flow rate measurement accuracy with pulsed output flowmeters.